Abstract
The beam quality correction factor and the perturbation factor pQ, commonly considered in dosimetry with ionization chambers, were calculated for the
NE2571 and the Standard Imaging A19 and A12S chambers, using the Monte Carlo simulation
code PENELOPE. For the NE2571 chamber, the values of obtained are in very good agreement with those found in previous works by Wulff et al.
and Muir and Rogers with the code EGSnrc and also with the experimental results summarized
in the NCS code of practice. For pQ, a difference of ∼0.4% has been found between our results and those obtained with
EGSnrc for 60Co and this difference increases slightly with values. These factors have been calculated also for the A19 and A12S chambers of
Standard Imaging. The values of show reasonable agreement with those recently calculated by Muir and Rogers and the
measurements of McEwen.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Physica Medica: European Journal of Medical PhysicsAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams.Med Phys. 1999; 26: 1847-1870
- Absorbed dose determination in external beam radiotherapy.(IAEA technical reports series 398) IAEA, Vienna2000
- Electron beam quality correction factors for plane-parallel ionization chambers in high energy photon beams: Monte Carlo calculations using the PENELOPE system.Phys Med Biol. 2004; 49: 4427-4444
- PENELOPE: a code system for Monte Carlo simulation of electron and photon transport.Nuclear Energy Agency, Paris2008
- Monte Carlo based perturbation and beam quality corrections factors for thimble ionization chambers in high energy photon beams.Phys Med Biol. 2008; 53: 2823-2836
- Monte Carlo calculations of kQ, the beam quality conversion factor.Med Phys. 2010; 37: 5939-5950
- Code of practice for the absorbed dose determination in high energy photon and electron beams.(NCS Report 18) NCS, Delft2008
- Measurement of ionization chamber absorbed dose kQ factors in megavoltage photon beams.Med Phys. 2010; 37: 2179-2193
- The EGSnrc code system: Monte Carlo simulation of electron and photon transport.(NRC technical report no. PIRS-701, v4-2-2-5. See:) National Research Council of Canada, Ottawa2007
- Geant4 – a simulation tool kit.Nucl Instrum Methods Phys Res A. 2003; 506: 250-303
- Pelowitz D.B. MCNPX users manual version 2.5.0, Report LA-UR-02-2607. Los Alamos National Laboratory, Los Alamos2005
- penEasy: a modular main program and voxelized geometry package for PENELOPE.2009
- clonEasy: a package of linux scripts for the parallelization of Monte Carlo simulations.Comp Phys Commun. 2005; 175: 440-450
- The design of a thimble chamber for the Farmer dosemeter.Phys Med Biol. 1972; 17: 169-174
- On the effective point of measurement in megavoltage photon beams.Med Phys. 2006; 33: 1829-1839
- Chamber-quality factors in 60Co for three plane parallel chambers for the dosimetry of electrons, protons and heavy charged particles: PENELOPE Monte Carlo simulations.Phys Med Biol. 2008; 53: 5917-5926
- Improved calculations of stopping-power ratios and their correlation with the quality of therapeutic photon beams.in: Proceedings symposium on measurement assurance in dosimetry. IAEA, Vienna1994
- Investigation of systematic uncertainties in Monte Carlo-calculated beam quality correction factors.Phys Med Biol. 2010; 55: 4481-4493
- Monte Carlo simulation of a typical 60Co therapy source.Med Phys. 1999; 26: 1847-1870
- Monte Carlo calculation of nine megavoltage photon beam spectra using BEAM code.Med Phys. 2002; 29: 391-402
- An empirical relationship for determining photon beam quality in TG-21 from a ratio of percent depth doses.Med Phys. 1998; 25: 1202-1205
- Chamber quality factors for the NACP-02 chamber in 60Co beams: comparison of EGSnrc and PENELOPE Monte Carlo simulations.in: International symposium on standards, applications and quality assurance in medical radiation dosimetry. IAEA, Vienna2010
- Measured and Monte Carlo calculated kQ factors: Accuracy and comparison.Med Phys. 2011; 38: 4600-4609
- Which accelerator photon beams are clinic-like for reference dosimetry purposes?.Med Phys. 2003; 30: 1546-1555
Article info
Publication history
Published online: January 25, 2012
Accepted:
January 5,
2012
Received in revised form:
December 31,
2011
Received:
February 17,
2011
Identification
Copyright
© 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Inc. All rights reserved.