Review Paper| Volume 32, ISSUE 4, P541-556, April 2016

Reference radiochromic film dosimetry: Review of technical aspects

  • Slobodan Devic
    Correspondence
    Corresponding author. Jewish General Hospital, 3755 chemin de la Côte-Sainte-Catherine, Montréal, Québec H3T 1E2, Canada. Tel.: +1 514 340 8222 x 2595; fax: 514 340 8642.
    Affiliations
    Medical Physics Unit, Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montréal, Québec, Canada
    Search for articles by this author
  • Nada Tomic
    Affiliations
    Medical Physics Unit, Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montréal, Québec, Canada
    Search for articles by this author
  • David Lewis
    Affiliations
    RCF Consulting, LLC, 54 Benedict Road, Monroe, Connecticut 06468
    Search for articles by this author
Published:March 22, 2016DOI:https://doi.org/10.1016/j.ejmp.2016.02.008

      Highlights

      • Review of technical aspects for radiochromic film dosimetry system.
      • Recipe on how to design a reference radiochromic film dosimetry system.
      • Summary of parameters that define radiochromic film dosimetry protocol.

      Abstract

      For decades, film was used as a powerful two-dimensional (2D) dosimetry tool for radiotherapy treatment verification and quality assurance. Unlike the old silver-halide based radiographic films, radiochromic films change its color upon irradiation without the need for chemical development. Radiation dose deposited within a sensitive layer of the radiochromic film initiates polymerization of the active component, the degree of which depends on the amount of energy deposited. Response of the film to radiation is commonly expressed in terms of optical density change, which can be easily measured by any photometric device. However, a number of factors may have an impact on the signal detected by the measuring device. This review summarizes technical aspects associated with the establishment of reference radiochromic film dosimetry and its subsequent use for either clinical or research applications.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Physica Medica: European Journal of Medical Physics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Saylor M.C.
        • Tamargo T.T.
        • McLaughlin W.L.
        • Khan H.M.
        • Lewis D.F.
        • Schenfele R.D.
        A thin film recording medium for use in food irradiation.
        Radiat Phys Chem. 1988; 31: 529-536
        • Chu R.D.H.
        • VanDyke G.
        • Lewis D.F.
        • O'Hara K.P.J.
        • Buckland B.R.
        • Dinelle F.
        GafChromic dosimetry media: a new high dose rate thin film routine dosimeter and dose mapping tool.
        Radiat Phys Chem. 1990; 35: 767-773
        • Butson M.J.
        • Cheung T.
        • Yu P.K.N.
        Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp x-ray range.
        Appl Radiat Isot. 2006; 64: 60-62
        • Rink A.
        • Vitkin I.A.
        • Jaffray D.A.
        Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter.
        Med Phys. 2007; 34: 458-463
        • Poon E.
        • Reniers B.
        • Devic S.
        • Vuong T.
        • Verhaegen F.
        Dosimetric characterization of a novel intracavitary mold applicator for Ir-192 high dose rate endorectal brachytherapy treatment.
        Med Phys. 2006; 33: 4515-4526
        • Chiu-Tsao S.
        • Hanley J.
        • Napoli J.
        • Davis S.
        • Pike T.
        • DeWerd L.
        Determination of TG43 parameters for Cs-131 model CS-1R2 seed using radiochromic EBT film dosimetry.
        Med Phys. 2007; 34: 2434-2435
        • Le Y.
        • Armour E.
        • Wong J.
        Evaluation of heterogeneity effect in intra-operative HDR (IOHDR) brachytherapy dose calculation using Monte Carlo simulation and GAFCHROMIC EBT film measurement.
        Med Phys. 2007; 34: 2450
        • Evans M.
        • Devic S.
        • Podgorsak E.B.
        High dose-rate brachytherapy source position quality assurance using radiochromic film.
        Med Dosim. 2007; 32: 13-15
        • Bufacchi A.
        • Carosi A.
        • Adorante N.
        • Delle Canne S.
        • Malatesta T.
        • Capparella R.
        • et al.
        In vivo EBT radiochromic film dosimetry of electron beam for Total Skin Electron Therapy (TSET).
        Phys Med. 2007; 23: 67-72
        • Lightfoot D.
        Total skin electron beam commissioning with EBT film.
        Med Phys. 2006; 33: 2146
        • Su F.C.
        • Liu Y.
        • Stathakis S.
        • Shi C.
        • Esquivel C.
        • Papanikolaou N.
        Dosimetry characteristics of GAFCHROMIC (R) EBT film responding to therapeutic electron beams.
        Appl Radiat Isot. 2007; 65: 1187-1192
        • Gerbi B.J.
        • Han E.Y.
        The response of radiochromic EBT film in high-energy electron beams.
        Med Phys. 2006; 33: 2144
        • Butson M.J.
        • Cheung T.
        • Yu P.K.N.
        Megavoltage x-ray skin dose variation with an angle using grid carbon fibre couch tops.
        Phys Med Biol. 2007; 52: N485-92
        • Devic S.
        • Seuntjens J.
        • Abdel-Rahman W.
        • Evans M.
        • Olivares M.
        • Podgorsak E.B.
        • et al.
        Accurate skin dose measurements using radiochromic film in clinical applications.
        Med Phys. 2006; 33: 1116-1124
        • Su F.C.
        • Shi C.Y.
        • Papanikolaou N.
        Clinical application of GAFCHROMIC (R) EBT film for in vivo dose measurements of total body irradiation radiotherapy.
        Appl Radiat Isot. 2008; 66: 389-394
        • Nioutsikou E.
        • Seppenwoolde Y.
        • Symonds-Tayler J.R.
        • Heijmen B.
        • Evans P.
        • Webb S.
        Dosimetric investigation of lung tumor motion compensation with a robotic respiratory tracking system: an experimental study.
        Med Phys. 2008; 35: 1232-1240
        • Wilcox E.E.
        • Daskalov G.
        Use of EBT film for dose measurement in heterogeneous phantoms containing lung and bone equivalent materials for 6 MV photon fields in the range 0.5 to 4 cm diameter produced by Cyberknife.
        Radiother Oncol. 2007; 84: S53
        • Polednik M.
        • Abo Madyan Y.
        • Schneider F.
        • Wolff D.
        • Bannach B.
        • Lambrecht U.
        • et al.
        Evaluation of calculation algorithms implemented in different commercial planning systems on an anthropomorphic breast phantom using film dosimetry.
        Strahlenther Onkol. 2007; 183: 667-672
        • Wilcox E.E.
        • Daskalov G.M.
        Evaluation of GAFCHROMIC (R) EBT film for CyberKnife (R) dosimetry.
        Med Phys. 2007; 34: 1967-1974
        • Wilcox E.E.
        • Daskalov G.M.
        Accuracy of dose measurements and calculations within and beyond heterogeneous tissues for 6 MV photon fields smaller than 4 cm produced by Cyberknife.
        Med Phys. 2008; 35: 2259-2266
        • Sturtewagen E.
        • Fuss M.
        • Paelinck L.
        • De Wagter C.
        • Georg D.
        Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT film.
        Z Med Phys. 2008; 18: 27-36
        • Ciangaru G.
        • Yang J.N.
        • Oliver P.J.
        • Bues M.
        • Zhu M.
        • Nakagawa F.
        • et al.
        Verification procedure for isocentric alignment of proton beams.
        J Appl Clin Med Phys. 2007; 8: 65-75
        • Tomic N.
        • Gosselin M.
        • Wan J.F.
        • Saragovi U.
        • Podgorsak E.B.
        • Evans M.
        • et al.
        Verification of cell irradiation dose deposition using radiochromic film.
        Phys Med Biol. 2007; 52: 3121-3131
        • Lynch B.
        • Ranade M.
        • Li J.
        • Dempsey J.
        Characteristics of a new very high sensitivity radiochromic film.
        Med Phys. 2004; 31: 1873
        • Sankar A.
        • Ayyangar K.M.
        • Nehru R.M.
        • Kurup P.G.
        • Murali V.
        • Enke C.A.
        • et al.
        Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.
        Med Dosim. 2006; 31: 273-282
        • Tomic N.
        • Quintero C.
        • Whiting B.R.
        • Aldelaijan S.
        • Bekerat H.
        • Liang L.
        • et al.
        Characterization of the calibration curves and energy dependence GafchromicTM XR-QA2 model based radiochromic film dosimetry system.
        Med Phys. 2014; 41: 062105
        • Bekerat H.
        • Devic S.
        • DeBlois F.
        • Singh K.
        • Sarfehnia A.
        • Seuntjens J.
        • et al.
        Improving the Energy Response of External Beam Therapy (EBT) GafChromicTM dosimetry films at low energies (≤100 keV).
        Med Phys. 2014; 41: 022101
        • Sarfehnia A.
        • Stewart K.
        • Seuntjens J.
        An absorbed dose to water standard for HDR 192Ir brachytherapy sources based on water calorimetry: numerical and experimental proof-of-principle.
        Med Phys. 2007; 34: 4957-4961
        • Lillo F.
        • Mettivier G.
        • Sarno A.
        • Tromba G.
        • Tomic N.
        • Devic S.
        • et al.
        Energy dependent calibration of XR-QA2 radiochromic film with monochromatic and polychromatic x-rays beams.
        Med Phys. 2016; 43: 583-588
        • Alva H.
        • Mercado-Uribe H.
        • Rodriguez-Villafuertre M.
        • Brandan M.E.
        The use of a reflective scanner to study radiochromic film response.
        Phys Med Biol. 2002; 47: 2925-2933
        • Papaconstadopoulos P.
        • Hegyi G.
        • Seuntjens J.
        • Devic S.
        A protocol for EBT3 radiochromic film dosimetry using reflection scanning.
        Med Phys. 2014; 41: 122101
        • Devic S.
        • Seuntjens J.
        • Hegyi G.
        • Podgorsak E.B.
        • Soares C.G.
        • Kirov A.S.
        • et al.
        Dosimetric properties of improved GafChromic films for seven different digitizers.
        Med Phys. 2004; 31: 2392-2401
        • Baker J.
        • Reinstein L.
        • Gluckman G.
        Finding the optimal digitizer for use with GafChromic EBT film.
        Med Phys. 2005; 32: 1983
        • Devic S.
        Radiochromic film dosimetry: past, present, and future.
        Phys Med. 2011; 27: 122-134
        • Butson M.J.
        • Yu P.K.
        • Cheung T.
        • Inwood D.
        Polarization effects on a high-sensitivity radiochromic film.
        Phys Med Biol. 2003; 48: N207-11
        • Aldelaijan S.
        • Devic S.
        • Mohammed H.
        • Tomic N.
        • Liang L.
        • DeBlois F.
        • et al.
        Evaluation of EBT-2 Model GAFCHROMICTM film performance in water.
        Med Phys. 2010; 37: 3687-3693
        • Paelinck L.
        • De Neve W.
        • De Wagter C.
        Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry.
        Phys Med Biol. 2007; 52: 231-242
        • Ferreira B.C.
        • Lopes M.C.
        • Capela M.
        Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry.
        Phys Med Biol. 2009; 54: 1073-1085
        • Menegotti L.
        • Delana A.
        • Martignano A.
        Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure.
        Med Phys. 2008; 35: 3078-3085
        • Chung H.
        • Lynch B.
        • Samant S.
        High-precision GAFCHROMIC EBT film-based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non-uniformity correction.
        J Appl Clin Med Phys. 2010; 11: 3112
        • Matney J.E.
        • Parker B.C.
        • Neck D.W.
        • Henkelmann G.
        • Rosen I.I.
        Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry.
        J Appl Clin Med Phys. 2010; 11: 3165
        • Desroches J.
        • Bouchard H.
        • Lacroix F.
        Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry.
        Med Phys. 2010; 37: 1565-1570
        • Alnawaf H.
        • Yu P.K.
        • Butson M.
        Comparison of Epson scanner quality for radiochromic film evaluation.
        J Appl Clin Med Phys. 2012; 13: 3957
        • Andreo P.
        • Burns D.T.
        • Hohlfeld K.
        • Huq M.S.
        • Kanai T.
        • Laitano F.
        • et al.
        Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water.
        Technical Report Series. vol. 398. IAEA, Vienna2000
        • Almond P.R.
        • Biggs P.J.
        • Coursey B.M.
        • Hanson W.H.
        • Huq M.S.
        • Nath R.
        • et al.
        AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams.
        Med Phys. 1999; 26: 1847-1870
        • Ma C.M.
        • Coffey C.W.
        • DeWerd L.A.
        • Liu C.
        • Nath R.
        • Seltzer S.M.
        • et al.
        AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology.
        Med Phys. 2001; 28: 868-893
        • Micke A.
        • Lewis D.F.
        • Yu X.
        Multichannel film dosimetry with nonuniformity correction.
        Med Phys. 2011; 38: 2523-2534
        • Devic S.
        • Seuntjens J.
        • Sham E.
        • Podgorsak E.B.
        • Kirov A.S.
        • Schmidtlein R.C.
        • et al.
        Precise radiochromic film dosimetry using a flat-bed document scanner.
        Med Phys. 2005; 32: 2245-2253
        • Devic S.
        • Wang Y.
        • Tomic N.
        • Podgorsak E.B.
        Sensitivity of linear CCD array based film scanners used for film dosimetry.
        Med Phys. 2006; 33: 3993-3996
        • van Battum L.J.
        • Huizenga H.
        • Verdaasdonk R.M.
        • Heukelom S.
        How flatbed scannersupset accurate film dosimetry.
        Phys Med Biol. 2016; 61: 625-649
        • Devic S.
        • Aldelaijan S.
        • Mohammed H.
        • Tomic N.
        • Liang L.
        • DeBlois F.
        • et al.
        Absorption Spectra Time Evolution of EBT-2 Model GAFCHROMICTM Film.
        Med Phys. 2010; 37: 2207-2214
        • Hartmann B.
        • Martisiková M.
        • Jäkel O.
        Homogeneity of Gafchromic EBT2 film.
        Med Phys. 2010; 37: 1753-1756
      1. GAFCHROMICTM EBT2 self-developing film for radiotherapy dosimetry, ISP White Paper, March, 2010.

        • McCaw T.J.
        • Micka J.A.
        • Dewerd L.A.
        Characterizing the marker-dye correction for Gafchromic(®) EBT2 film: a comparison of three analysis methods.
        Med Phys. 2011; 38: 5771-5777
        • Mizuno H.
        • Takahashi Y.
        • Tanaka A.
        • Hirayama T.
        • Yamaguchi T.
        • Katou H.
        • et al.
        Homogeneity of GAFCHROMIC EBT2 film among different lot numbers.
        J Appl Clin Med Phys. 2012; 13: 3763
        • Rink A.
        • Lewis D.F.
        • Varma S.
        • Vitkin I.A.
        • Jaffray D.A.
        Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium.
        Med Phys. 2008; 35: 4545-4555
        • Park S.
        • Kang S.K.
        • Cheong K.H.
        • Hwang T.
        • Kim H.
        • Han T.
        • et al.
        Variations in dose distribution and optical properties of Gafchromic(TM) EBT2 film according to scanning mode.
        Med Phys. 2012; 39: 2524-2535
        • Girard F.
        • Bouchard H.
        • Lacroix F.
        Reference dosimetry using radiochromic film.
        J Appl Clin Med Phys. 2012; 13: 339-353
        • Tomic N.
        • Devic S.
        • DeBlois F.
        • Seuntjens J.
        Reference radiochromic film dosimetry in kilo-voltage photon beams during CBCT image acquisition.
        Med Phys. 2010; 37: 1083-1092
        • Aldelaijan S.
        • Mohammed H.
        • Tomic N.
        • Liang L.H.
        • Deblois F.
        • Sarfehnia A.
        • et al.
        Radiochromic film dosimetry of HDR (192)Ir source radiation fields.
        Med Phys. 2011; 38: 6074-6083
        • Aldelaijan S.
        • Alzorkany F.
        • Moftah B.
        • Buzurovic I.
        • Seuntjens J.
        • Tomic N.
        • et al.
        Use of a control film piece in radiochromic film dosimetry.
        Phys Med. 2016; https://doi.org/10.1016/j.ejmp.2015.12.004
        • Devic S.
        • Tomic N.
        • Aldelaijan S.
        • DeBlois F.
        • Seuntjens J.
        • Chan M.F.
        • et al.
        Linearization of dose–response curve of the radiochromic film dosimetry system.
        Med Phys. 2012; 39: 4850-4857
        • Bevington P.R.
        • Robinson D.K.
        Data reduction and error analysis for the physicsal sciences.
        WCB/McGraw-Hill, Boston1992
        • Mayer R.R.
        • Ma F.
        • Chen Y.
        • Miller R.I.
        • Belard A.
        • McDonough J.
        • et al.
        Enhanced dosimetry procedures and assessment for EBT2 radiochromic film.
        Med Phys. 2012; 39: 2147-2155
        • Méndez I.
        • Peterlin P.
        • Hudej R.
        • Strojnik A.
        • Casar B.
        On multichannel film dosimetry with channel-independent perturbations.
        Med Phys. 2014; 41: 011705
        • Saur S.
        • Frengen J.
        GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.
        Med Phys. 2008; 35: 3094-3101
        • Martisíková M.
        • Ackermann B.
        • Jäkel O.
        Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams.
        Phys Med Biol. 2008; 53: 7013-7027
        • Garcia L.I.
        • Azorin J.F.
        Improving the calibration of radiochromic films by the use of uncertainties in optical density and dose.
        Med Phys. 2013; 40: 071726
        • Devic S.
        • Tomic N.
        • Soares C.G.
        • Podgorsak E.B.
        Optimizing the dynamic range extension of a radiochromic film dosimetry system.
        Med Phys. 2009; 36: 429-437