Sparse proportional re-scanning with hadron beams

Published:September 07, 2019DOI:https://doi.org/10.1016/j.ejmp.2019.07.022

      Highlights

      • The novel sparse proportional re-scanning method reduces the treatment time.
      • The advantage is quantified using Eclipse treatment plans of 54 proton fields.
      • It is proven that the time “reduction factor” depends only on the number N of re-scannings.
      • The reduction factor does not depend on the shape and volume of the target, the numbers of layers or on the spot grid.
      • The same reduction factors apply also to carbon ion treatments.

      Abstract

      Spot Scanning is a well-established technique to deliver the dose with hadron therapy systems. For many years re-scanning (called also re-painting) has been used to achieve uniform dose distribution in particular for moving organs, although it leads to an increase of the treatment time. Reducing this time is a major focus of present research. In this paper, after reviewing the current re-scanning techniques, sparse proportional re-scanning is defined and applied to 29 proton patient cases for a total of 54 fields. In this technique, only the highest weighted spot in the whole target is visited a number of times that is equal to the number N of re-scans. The number of visits of the beam spot to all remaining spots is scaled down proportionally to their weight. Sparse proportional re-scanning is advantageous especially in volumetric re-scanning. In order to quantify the potential advantages of this technique in terms of treatment time, a reduction factor of the number of scanned spots has been introduced, evaluated and analysed for 54 proton fields. The conclusion is that the reduction factor is a function of N (having values equal to 2.8 ± 0.3 and 3.6 ± 0.4 for N = 5 and N = 12 respectively) and does not depend either on the shape and volume of the target or on the distance between the scanned layers and the spot grid. The same values are approximately valid also for carbon ion treatments.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Physica Medica: European Journal of Medical Physics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Haberer T.
        • Becher W.
        • Schardt D.
        • Kraft G.
        Magnetic scanning system for heavy ion therapy.
        Nucl Inst Methods Phys Res A. 1993; https://doi.org/10.1016/0168-9002(93)91335-K
        • Pedroni E.
        • Bacher R.
        • Blattmann H.
        • Bohrinaer T.
        • Coray A.
        • Lomax A.
        • et al.
        The 200-Mev proton therapy project at the Paul Scherrer Institute: conceptual design and practical realization.
        Med Phys. 1995; https://doi.org/10.1118/1.597522
        • Ulmer W.
        • Schaffner B.
        Foundation of an analytical proton beamlet model for inclusion in a general proton dose calculation system.
        Radiat Phys Chem. 2011; https://doi.org/10.1016/j.radphyschem.2010.10.006
        • Mizushima K.
        • Furukawa T.
        • Iwata Y.
        • Hara Y.
        • Saotome N.
        • Saraya Y.
        • et al.
        Performance of the HIMAC beam control system using multiple-energy synchrotron operation.
        Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms. 2017; https://doi.org/10.1016/j.nimb.2017.03.051
        • Amaldi U.
        • Braccini S.
        • Puggioni P.
        High frequency linacs for hadrontherapy.
        Rev Accel Sci Technol. 2009; https://doi.org/10.1142/s179362680900020x
        • Lomax A.J.
        • Böhringer T.
        • Bolsi A.
        • Coray D.
        • Emert F.
        • Goitein G.
        • et al.
        Treatment planning and verification of proton therapy using spot scanning: Initial experiences.
        Med Phys. 2004; https://doi.org/10.1118/1.1779371
        • Kardar L.
        • Li Y.
        • Li X.
        • Li H.
        • Cao W.
        • Chang J.Y.
        • et al.
        Evaluation and mitigation of the interplay effects of intensity modulated proton therapy for lung cancer in a clinical setting.
        Pract Radiat Oncol. 2014; https://doi.org/10.1016/j.prro.2014.06.010
        • Albertini F.
        • Hug E.B.
        • Lomax A.J.
        Is it necessary to plan with safety margins for actively scanned proton therapy?.
        Phys Med Biol. 2011; https://doi.org/10.1088/0031-9155/56/14/011
        • Schätti A.
        • Zakova M.
        • Meer D.
        • Lomax A.J.
        Experimental verification of motion mitigation of discrete proton spot scanning by re-scanning.
        Phys Med Biol. 2013; https://doi.org/10.1088/0031-9155/58/23/8555
        • Bernatowicz K.
        • Lomax A.J.
        • Knopf A.
        Comparative study of layered and volumetric rescanning for different scanning speeds of proton beam in liver patients.
        Phys Med Biol. 2013; https://doi.org/10.1088/0031-9155/58/22/7905
        • Knopf A.C.
        • Lomax A.J.
        In the context of radiosurgery – pros and cons of rescanning as a solution for treating moving targets with scanned particle beams.
        Phys Medica. 2014; https://doi.org/10.1016/j.ejmp.2014.03.010
        • Bert C.
        • Grözinger S.O.
        • Rietzel E.
        Quantification of interplay effects of scanned particle beams and moving targets.
        Phys Med Biol. 2008; https://doi.org/10.1088/0031-9155/53/9/003
        • Bert C.
        • Durante M.
        Motion in radiotherapy: particle therapy.
        Phys Med Biol. 2011; https://doi.org/10.1088/0031-9155/56/16/R01
        • Minohara S.
        • Kanai T.
        • Endo M.
        • Noda K.
        • Kanazawa M.
        Respiratory gated irradiation system for heavy-ion radiotherapy.
        Int J Radiat Oncol Biol Phys. 2000; https://doi.org/10.1016/S0360-3016(00)00524-1
        • Bert C.
        • Gemmel A.
        • Saito N.
        • Rietzel E.
        Gated irradiation with scanned particle beams.
        Int J Radiat Oncol Biol Phys. 2009; https://doi.org/10.1016/j.ijrobp.2008.11.014
        • Zenklusen S.M.
        • Pedroni E.
        • Meer D.
        A study on repainting strategies for treating moderately moving targets with proton pencil beam scanning at the new gantry 2 at PSI.
        Phys Med Biol. 2010; https://doi.org/10.1088/0031-9155/55/17/014
        • Engwall E.
        • Glimelius L.
        • Hynning E.
        Effectiveness of different rescanning techniques for scanned proton radiotherapy in lung cancer patients.
        Phys Med Biol. 2018; https://doi.org/10.1088/1361-6560/aabb7b
        • Parodi K.
        • Saito N.
        • Chaudhri N.
        • Richter C.
        • Durante M.
        • Enghardt W.
        • et al.
        4D in-beam positron emission tomography for verification of motion-compensated ion beam therapy.
        Med Phys. 2009; https://doi.org/10.1118/1.3196236
        • Zhang Y.
        • Knopf A.
        • Tanner C.
        • Lomax A.J.
        Online image guided tumour tracking with scanned proton beams: a comprehensive simulation study.
        Phys Med Biol. 2014; https://doi.org/10.1088/0031-9155/59/24/7793
        • Charlie Ma C.-M.
        • Lomax A.J.
        Proton and carbon ion therapy.
        CRC Press Taylor & Francis Group, 2013
        • Zhang Y.
        • Huth I.
        • Wegner M.
        • Weber D.C.
        • Lomax A.J.
        An evaluation of rescanning technique for liver tumour treatments using a commercial PBS proton therapy system.
        Radiother Oncol. 2016; https://doi.org/10.1016/j.radonc.2016.09.011
      1. https://www.varian.com/oncology/products/software/treatment-planning/eclipse-proton.

        • Knopf A.C.
        • Hong T.S.
        • Lomax A.
        Scanned proton radiotherapy for mobile targets – the effectiveness of re-scanning in the context of different treatment planning approaches and for different motion characteristics.
        Phys Med Biol. 2011; https://doi.org/10.1088/0031-9155/56/22/016
        • Kubiak T.
        Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation.
        Br J Radiol. 2016; https://doi.org/10.1259/bjr.20150275
        • Trbojevic D.
        • Parker B.
        • Keil E.
        • Sessler A.M.
        Carbon/proton therapy: a novel gantry design.
        Phys Rev Spec Top - Accel Beams. 2007; https://doi.org/10.1103/PhysRevSTAB.10.053503
        • Brouwer L.
        • Caspi S.
        • Hafalia R.
        • Hodgkinson A.
        • Prestemon S.
        • Robin D.
        • et al.
        Design of an achromatic superconducting magnet for a proton therapy gantry.
        IEEE Trans Appl Supercond. 2017; https://doi.org/10.1109/TASC.2016.2628305
        • Gerbershagen A.
        • Meer D.
        • Schippers J.M.
        • Seidel M.
        A novel beam optics concept in a particle therapy gantry utilizing the advantages of superconducting magnets.
        Z Med Phys. 2016; https://doi.org/10.1016/j.zemedi.2016.03.006
      2. Garonna A, Alharbi N, Amaldi U, Benedetto E, Ljubicic V, Riboni PL. Delivery and gantry concepts from TERA. Presentation at Workshop on Ideas and technologies for a next generation facility for medical research and therapy with ions, Archamps (ESI) 2018, https://indico.cern.ch/event/682210/contributions/2997883/attachments/1671379/2711300/20180620_PIMMS2workshopAG_v7_-_filt.pdf n.d.

        • Amaldi U.
        Oblique raster scanning: an ion dose delivery procedure with variable energy layers.
        Phys Med Biol. 2019; 64: 115003https://doi.org/10.1088/1361-6560/ab0920