Integration of the M6 Cyberknife in the Moderato Monte Carlo platform and prediction of beam parameters using machine learning

Published:January 30, 2020DOI:https://doi.org/10.1016/j.ejmp.2020.01.018

      Highlights

      • A Monte Carlo model for the M6 Cyberknife is integrated in the Moderato platform.
      • The model is validated using patient plans re-calculations and film measurements.
      • A machine learning algorithm is trained to predict electron beam parameters.
      • The method is validated on other M6 devices through Monte Carlo simulations.

      Abstract

      Purpose

      This work describes the integration of the M6 Cyberknife in the Moderato Monte Carlo platform, and introduces a machine learning method to accelerate the modelling of a linac.

      Methods

      The MLC-equipped M6 Cyberknife was modelled and integrated in Moderato, our in-house platform offering independent verification of radiotherapy dose distributions. The model was validated by comparing TPS dose distributions with Moderato and by film measurements. Using this model, a machine learning algorithm was trained to find electron beam parameters for other M6 devices, by simulating dose curves with varying spot size and energy. The algorithm was optimized using cross-validation and tested with measurements from other institutions equipped with a M6 Cyberknife.

      Results

      Optimal agreement in the Monte Carlo model was reached for a monoenergetic electron beam of 6.75 MeV with Gaussian spatial distribution of 2.4 mm FWHM. Clinical plan dose distributions from Moderato agreed within 2% with the TPS, and film measurements confirmed the accuracy of the model. Cross-validation of the prediction algorithm produced mean absolute errors of 0.1 MeV and 0.3 mm for beam energy and spot size respectively. Prediction-based simulated dose curves for other centres agreed within 3% with measurements, except for one device where differences up to 6% were detected.

      Conclusions

      The M6 Cyberknife was integrated in Moderato and validated through dose re-calculations and film measurements. The prediction algorithm was successfully applied to obtain electron beam parameters for other M6 devices. This method would prove useful to speed up modelling of new machines in Monte Carlo systems.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Physica Medica: European Journal of Medical Physics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Lambin P.
        • Roelofs E.
        • Reymen B.
        • Velazquez E.R.
        • Buijsen J.
        • Zegers C.M.L.
        • et al.
        “Rapid Learning health care in oncology” - an approach towards decision support systems enabling customised radiotherapy.
        Radiother Oncol. 2013; 109: 159-164
        • Jaffray D.A.
        • Lindsay P.E.
        • Brock K.K.
        • Deasy J.O.
        • Tomé W.A.
        Accurate accumulation of dose for improved understanding of radiation effects in normal tissue.
        Int J Radiat Oncol Biol Phys. 2010; 76: 135-139
        • Reynaert N.
        • Demol B.
        • Charoy M.
        • Bouchoucha S.
        • Crop F.
        • Wagner A.
        • et al.
        Clinical implementation of a Monte Carlo based treatment plan QA platform for validation of Cyberknife and Tomotherapy treatments.
        Phys Med. 2016; 32
        • Wagner A.
        • Crop F.
        • Mirabel X.
        • Tailly C.
        • Reynaert N.
        Use of an in-house Monte Carlo platform to assess the clinical impact of algorithm-related dose differences on DVH constraints.
        Phys Med. 2017;
        • Kawrakow I.
        Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version.
        Med Phys. 2000; 27: 485-498
        • Rogers D.W.O.
        • Faddegon B.A.
        • Ding G.X.
        • Ma C.M.
        • We J.
        • Mackie T.R.
        BEAM: a Monte Carlo code to simulate radiotherapy treatment units.
        Med Phys. 1995; 22: 503-524
        • Asmerom G.
        • Bourne D.
        • Chappelow J.
        • Goggin L.M.
        • Heitz R.
        • Jordan P.
        • et al.
        The design and physical characterization of a multileaf collimator for robotic radiosurgery.
        Biomed Phys Eng Express. 2016; 2
        • Jelen U.
        • Söhn M.
        • Alber M.
        A finite size pencil beam for IMRT dose optimization.
        Phys Med Biol. 2005; 50: 1747-1766
        • Ma C.M.
        • Li J.S.
        • Pawlicki T.
        • Jiang S.B.
        • Deng J.
        • Lee M.C.
        • et al.
        A Monte Carlo dose calculation tool for radiotherapy treatment planning.
        Phys Med Biol. 2002; 47: 1671-1689
        • Ma C.
        • Li J.
        • Deng J.
        Implementation of Monte Carlo Dose calculation for CyberKnife treatment planning.
        J Phys Conf Ser. 2008; 102
      1. Pena J, González-castaño DM, Gómez F, Sánchez-doblado F. Automatic determination of primary electron beam parameters in Monte Carlo simulation n.d.

        • Conneely E.
        • Alexander A.
        • Stroian G.
        • Seuntjens J.
        • Foley M.J.
        An investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application.
        J Appl Clin Med Phys. 2013; 14: 3-14
        • Rogers D.W.O.
        • Walters B.
        • Kawrakow I.
        BEAMnrc users manual.
        Source. 2011; 509: 1-260
      2. Walters B, Kawrakow I, Rogers DWO. DOSXYZnrc Users Manual. NRCC Rep PIRS-0794 2016(April):1–125.

        • Wagner A.
        • Crop F.
        • Lacornerie T.
        • Vandevelde F.
        • Reynaert N.
        Use of a liquid ionization chamber for stereotactic radiotherapy dosimetry.
        Phys Med Biol. 2013; 58
        • Demol B.
        • Viard R.
        • Reynaert N.
        Monte Carlo calculation based on hydrogen composition of the tissue for MV photon radiotherapy.
        J Appl Clin Med Phys. 2015; 16: 117-130
        • Vanderstraeten B.
        • Chin P.W.
        • Fix M.
        • Leal A.
        • Mora G.
        • Reynaert N.
        • et al.
        Conversion of CT numbers into tissue parameters for Monte Carlo dose calculations: a multi-centre study.
        Phys Med Biol. 2007; 52: 539
        • Micke A.
        • Lewis D.F.
        • Yu X.
        Multichannel film dosimetry with nonuniformity correction.
        Med Phys. 2011; 38: 2523-2534
        • Pedregosa F.
        • Weiss R.
        • Brucher M.
        Scikit-learn: machine learning in python.
        J Mach Learn Res. 2011; 12: 2825-2830
        • Mackeprang P.
        • Vuong D.
        • Volken W.
        • Henzen D.
        • Schmidhalter D.
        • Malthaner M.
        • et al.
        Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy.
        Phys Med Biol. 2018; 63
        • Heidorn S.
        • Kilby W.
        • Fürweger C.
        Novel Monte Carlo dose calculation algorithm for robotic radiosurgery with multi leaf collimator: dosimetric evaluation.
        Phys Med. October 2018; 2019: 25-32
        • Mackeprang P.
        • Vuong D.
        • Volken W.
        • Henzen D.
        • Schmidhalter D.
        • Malthaner M.
        • et al.
        Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments.
        Radiat Oncol. 2019; 14: 1-11