External beam radiation therapy with kilovoltage x-rays

Published:November 19, 2020DOI:https://doi.org/10.1016/j.ejmp.2020.11.001

      Highlights

      • Kilovoltage x-ray beam therapy is cost effective but challenging.
      • Novel x-ray beam sources are discussed.
      • Novel techniques to deliver dose at depth are summarized.
      • Kilovoltage x-ray therapy with radiosensitizers shows promise clinically.

      Abstract

      Kilovoltage (kV) x-rays are most commonly used for diagnostic imaging due to their sensitivity to tissue composition. In radiation therapy (RT), due to their fast attenuation, kV x-rays are typically only used for superficial irradiation of skin cancer and for intra-operative RT (IORT). Recently, however, a number of kV RT techniques have emerged. In this review article, we provide a brief overview of the use of kV x-rays for RT.
      Various kV x-ray source technologies suitable for RT, such as conventional x-ray tubes as well as novel x-ray sources, are first described. This x-ray source section is then followed by a section on their implementation in terms of clinical, veterinary and preclinical applications. Specifically, IORT, superficial RT and dose enhancement with iodine and gold nanoparticles, as well as microbeam RT and FLASH RT are discussed in this context. Then, a number of kV x-ray RT applications in modeling and proof-of-principle stages, such as breast external beam RT with rotational sources, kilovoltage arc therapy and the BriXS Compton pulsed x-ray sources, are reviewed. Finally, some clinical and economic considerations for the development of kV RT techniques are discussed.

      Keywords

      1. Introduction

      1.1 History of kilovoltage x-ray beam radiotherapy

      Wilhelm Roentgen discovered x-rays in 1895. Within months of the discovery, x-rays were being applied to medical problems, primarily for imaging. The first radiation treatment of cancer using x-rays was recorded in 1896 when Victor Despeignes treated a patient with stomach cancer. Despite this first application of x-rays for stomach cancer, the primary therapeutic use of x-rays in these early days was only for superficial lesions. Early x-ray tube technology was only capable of generating x-rays in the kilovoltage (kV) energy range and the treatment of deep-seated lesions was limited by radiation-induced skin damage. In 1909, Alban Kohler, in an attempt to reduce the skin toxicity of kV x-rays and enable their use for the treatment of deep-seated lesions, devised the method of grid therapy [
      • Laissue J.A.
      • Blattmann H.
      • Slatkin D.N.
      Alban Köhler (1874-1947): Erfinder der Gittertherapie.
      ]. By passing the x-rays through a metal grid, the dose to skin was spatially fractionated thereby reducing the magnitude of skin toxicity. The basic technology of the x-ray tube remained relatively unchanged in the century following the discovery of x-rays.
      Today, 20–150 kVp x-rays are primarily used for diagnostic imaging. Higher-energy 100–500 kVp x-rays are used for superficial therapy, and in some cases, intraoperative radiation therapy. In this review article, we will summarize applications of kV radiotherapy with x-rays of energies between 50 and 320 kVp, typically referred to as superficial and orthovoltage x-rays. While the most common clinical applications of kV x-rays to superficial RT and intraoperative RT are briefly mentioned, this review focuses on ‘exotic’ applications of kV x-rays designed to treat deep-seated tumors either by combining kilovoltage RT with either contrast material injection or by designing non-standard x-ray sources. Various applications of kV RT, mostly in the treatment planning stage but also in pre-clinical or clinical trials are summarized here.

      1.2 Why use kV radiotherapy?

      Medical applications of x-ray beams in RT have been strictly segmented according to the energy employed by the radiation machines. High-energy (megavoltage, MV) systems are used for cancer therapy because their beams spare the skin and readily penetrate tissue. On the other hand, low-energy kV devices are mainly used for imaging, e.g. in computed tomography (CT), mammography and chest/bone x-ray imaging. These kV x-ray beams are more sensitive to tissue composition and less penetrating and their use for treatment is conventionally limited to skin cancer. While it is challenging to treat deep-seated tumors with kV x-rays, benefits of radiation treatment with kV x-rays exist.
      First, the generation of kV x-rays (see section 2) is much less complex than generation of MV x-rays. Conventional MV linear accelerators (linacs) used for clinical MV RT are sophisticated in their design and contain many components including an accelerating waveguide with a high manufacturing cost. The power and space requirements for operating linacs are greater than for operating kV x-ray devices. In addition, linacs require housing with 2-meter thick concrete walls to protect personnel; kV x-ray units only require ~ 2-millimeter lead shielding. In light of the considerable cost reduction possible with kV RT, it could become a suitable cancer treatment modality in the limited-resource settings of developing countries.
      Despite advancing healthcare technology, world-wide annual cancer mortality exceeds deaths from HIV-AIDS, tuberculosis and malaria combined. In part, this is the result of a significant and disturbing inequality in global delivery of cancer treatment; 70% of new cases are in the developing world where cases go undiagnosed and untreated because of cost [
      • Kerr D.J.
      • Midgley R.
      Can We Treat Cancer for a Dollar a Day? Guidelines for Low-Income Countries.
      ].
      X-ray treatment is effective as a single or combined modality for local–regional tumor control. It is also one of the most precisely planned and delivered therapeutic options. MV linacs and the infrastructure needed for treatment cost ~ US$5M and carry significant operating expenses in terms of personnel, supporting devices and services. The unit’s housing, requiring concrete bunkers, costs ∼$1.5 M to build. There has been little challenge to the present strategy restricting MV linear accelerators to treat most cancers, while consigning kV x-rays to imaging. Though kV x-ray systems have not been optimal for therapy within the body, they offer the potential for simultaneous high-quality imaging during treatment and are undoubtedly much less costly than MV systems.

      2. X-ray beam generation

      2.1 Conventional x-ray tubes

      The technology currently used to produce kV x-rays has not fundamentally changed from the original vacuum tube design used in Roentgen’s experiments. An x-ray tube consists of an evacuated glass tube containing an anode and a cathode. High voltage is applied across the vacuum tube and thermionic emission is used to liberate electrons from the cathode. The electrons are accelerated across the tube and strike the anode. The anode is made of a high atomic number (Z) metal and the rapid deceleration of the electrons incident on the anode produces photons by bremsstrahlung radiation. Anode design varies among x-ray tubes based on their application. X-ray tubes used for imaging employ a rotating anode which allows for smaller electron focal spots with high instantaneous current without exceeding the heat load specifications of the anode. X-ray tubes for therapy use larger electron focal spots along with lower currents over longer periods and use a large stationary anode [
      • Stathakis S.
      The Physics of Radiation Therapy.
      ].
      X-ray tube potential and added filtration are the main factors affecting the beam quality typically quoted in either mean energy or the half-value layer (HVL). For radiation therapy of deep-seated tumors, more filtration with higher tube potential, resulting in higher beam energy associated with deeper tissue penetration, is typically desirable. However, higher beam energy results in more heat-loading of the anode that can lead to anode melting. In addition, increased beam filtration decreases x-ray output, which in turn increases treatment time. A compromise between x-ray beam energy and beam filtration has to be identified for each radiotherapy application.

      2.2 Novel technology

      Apart from conventional x-ray tubes, a number of other x-ray sources that could be considered for the delivery of radiotherapy have been developed and will be discussed here.

      2.2.1 Scanning electron sources

      Since electrons are light charged particles, they can be easily steered electromagnetically with magnetic fields on the order of only 10–20 Gauss. There are two existing examples of diagnostic x-ray sources generated by scanned electron beam: the electron-beam computed tomography (EBCT) system [
      • Rand R.E.
      Scanning electron beam computed tomography scanner with ion aided focusing.
      ] and the scanning beam digital x-ray (SBDX) source. A modified design of the SBDX system has been simulated for the delivery of kilovoltage x-ray beam arc therapy (KVAT) discussed in section 3.4.2.
      The SBDX source consists of a large (23 × 23) cm2 tungsten anode and an electron beam scanning the anode in an array of 100 × 100 spots. The resulting x-ray beams from the transmission target are then shaped into 1.5-mm diameter beamlets directed towards the (5.5 × 5.5) cm2 detector by an ~ 2-cm-thick brass collimator [

      Van Lysel MS, Solomon EG, Wilfley BP, Dutta A, Speidel MA. Performance assessment of the scanning beam digital X-ray (SBDX) system. Proceedings of the SPIE - The International Society for Optical Engineering. 1997;3032:161-70.

      ]. The source was designed for the use in interventional and diagnostic radiography with lower radiation exposure to an interventionalist [
      • Speidel M.A.
      • Wilfley B.P.
      • Star-Lack J.M.
      • Heanue J.A.
      • Van Lysel M.S.
      Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography: SBDX technology for interventional cardiac angiography.
      ].

      2.2.2 Miniature x-ray tubes

      Small-size x-ray sources can be conveniently used for interstitial treatment of tumors in intraoperative treatments or for brachytherapy. In 1996, a novel miniature x-ray tube, the primary component of a Photon Radiosurgery System (PRS), was presented [
      • Dinsmore M.
      • Harte K.J.
      • Sliski A.P.
      • Smith D.O.
      • Nomikos P.M.
      • Dalterio M.J.
      • Boom A.J.
      • Leonard W.F.
      • Oettinger P.E.
      • Yanch J.C.
      A new miniature x-ray source for interstitial radiosurgery: Device description.
      ]. The electron gun was a conventional miniature gun with a barium oxide thermoionic cathode. Three anodes were used to extract, focus and accelerate the electron beam that impinged on a gold target to produce x-rays. The portable low-power 3-mm diameter 100-mm long x-ray source was capable of operating at 40 kV tube voltage and 40 µA tube current generating a beam with mean energy of 16 keV.
      The more commonly used miniature Xoft Axxent x-ray tube consists of a disposable microminiature x‐ray tube with a < 1-mm anode integrated into a cooled, flexible, disposable sheath of 5.3 mm in diameter. The Axxent x-ray source operates at tube voltages of 20 to 50 kV and tube currents of 1–300 µA producing air kerma strength of up to 1600 Gy.cm2.h−1 and it is designed for a lifetime of 2.5 h [
      • Rivard M.J.
      • Davis S.D.
      • DeWerd L.A.
      • Rusch T.W.
      • Axelrod S.
      Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent X-Ray Source: An electronic brachytherapy sourcea): Dosimetry parameters for the Xoft Axxent X-Ray Source.
      ]. Another commercial system which uses a small form (7 cm × 11 cm × 17 cm) x-ray tube is the INTRABEAM (Carl Zeiss Surgical, Oberkochen, Germany) intraoperative radiotherapy system. This system is capable of using various applicators which are inserted directly inside the surgical cavity which is then irradiated with 50 kV photons [
      • Sethi Anil
      • Emami Bahman
      • Small Jr., William
      • Thomas Tarita
      Intraoperative Radiotherapy With INTRABEAM: Technical and Dosimetric Considerations.
      ].

      2.2.3 Carbon nanotubes

      In most commercial x-ray sources, the properties of the electrons striking an anode drive many of the properties of the resulting x-ray beam, such as its spatial and temporal distribution. Conventional x-ray sources take advantage of thermionic emission (TE) for the generation of the electron beam, which poses challenges in terms of temporal (long beam ramp-up and cool-down times, up to ~ 1 s) and spatial (large ~ 1 mm beam size) x-ray beam characteristics. However, in field emission (FE), electrons tunnel through the narrowed potential barrier caused by an application of a high electric field. As a result, a very fast x-ray emission with response times on the order of 50 µs can be achieved [
      • Lei W.
      • Li C.
      • Cole M.T.
      • Qu K.e.
      • Ding S.
      • Zhang Y.
      • Warner J.H.
      • Zhang X.
      • Wang B.
      • Milne W.I.
      A graphene-based large area surface-conduction electron emission display.
      ]. In addition, due to the high-spatial resolution fabrication of carbon nanotubes (CNTs), FE-based CNT x-ray sources of 5 µm in size have been manufactured [
      • Heo S.H.
      • Ihsan A.
      • Cho S.O.
      Transmission-type microfocus x-ray tube using carbon nanotube field emitters.
      ].
      CNTs x-ray sources have been designed with imaging tasks in mind. They are typically capable of producing microfocus x-ray beams with anode voltages up to 70 kV, albeit with typically low ~ mA currents [
      • Parmee R.J.
      • Collins C.M.
      • Milne W.I.
      • Cole M.T.
      X-ray generation using carbon nanotubes.
      ]. Thanks to their small focal spot on the order of µm, CNTs have been studied for the use in spatially-fractionated microbeam radiotherapy (MRT) discussed in section 3.3.2.
      A miniaturized CNT has been developed by KAIST University in Korea for the use in intra-cavitary radiotherapy [
      • Kim H.J.
      • Ha J.M.
      • Heo S.H.
      • Cho S.O.
      Small-Sized Flat-Tip CNT Emitters for Miniaturized X-Ray Tubes.
      ]. The 7-mm diameter 47-mm long x-ray tube operates at tube voltages of up to 60 kV producing dose rates of 100 Gy/min in permanent operation. The advantage of a CNT-based miniature x-ray tube compared to conventional miniature x-ray tubes described in section 2.2.2 is their longer lifetime and cold emission radiation not requiring additional cooling.

      2.2.4 Synchrotron sources

      When electrons are accelerated on a curved path, they lose some of their energy by emission of x-rays termed synchrotron radiation. Synchrotron radiation is typically extracted in bending magnets of a ~ GeV electron beam circulating in a storage ring. Similar to x-ray tube beams, synchrotron radiation has a broad energy spectrum, but it is distinctive in a number of aspects compared to x-ray sources featuring an electron beam striking an anode. Synchrotron x-ray beams are of high flux, high brilliance, pulsed with short ~ picosecond-long pulses and polarized.
      Three synchrotron beamlines were used in radiation therapy experiments discussed in this article: the ID17 beamline at the European Synchrotron Research Facility (ESRF), the x17B beamline at Brookhaven National Laboratory (BNL) and the Imaging and Medical Beamline (IMBL) at the Australian Synchrotron. All beamlines are similar in their beam characteristics and only the ID17 ESRF beamline, with the most RT publications, will be briefly described. The high-brilliance low-divergence ID17 x-ray beam of 1–3 mm in height and ~ 40 mm in width can deliver dose rates of 8,000 to 16,000 Gy/ s [

      Serduc R, Bräuer-Krisch E, Siegbahn EA, Bouchet A, Pouyatos B, Carron R, et al. High-precision radiosurgical dose delivery by interlaced microbeam arrays of high-flux low-energy synchrotron x-rays. PLoS ONE. 2010;5:1-12.

      ,
      • Bräuer-Krisch E.
      • Adam J.-F.
      • Alagoz E.
      • Bartzsch S.
      • Crosbie J.
      • DeWagter C.
      • Dipuglia A.
      • Donzelli M.
      • Doran S.
      • Fournier P.
      • Kalef-Ezra J.
      • Kock A.
      • Lerch M.
      • McErlean C.
      • Oelfke U.
      • Olko P.
      • Petasecca M.
      • Povoli M.
      • Rosenfeld A.
      • Siegbahn E.A.
      • Sporea D.
      • Stugu B.
      Medical physics aspects of the synchrotron radiation therapies: Microbeam radiation therapy (MRT) and synchrotron stereotactic radiotherapy (SSRT).
      ]. The x-ray beam contains 30–600 keV x-rays with a mean energy of ~ 105 keV [
      • Laissue J.A.
      • Bartzsch S.
      • Blattmann H.
      • Bräuer-Krisch E.
      • Bravin A.
      • Dalléry D.
      • Djonov V.
      • Hanson A.L.
      • Hopewell J.W.
      • Kaser-Hotz B.
      • Keyriläinen J.
      • Laissue P.P.
      • Miura M.
      • Serduc R.
      • Siegbahn A.E.
      • Slatkin D.N.
      Response of the rat spinal cord to X-ray microbeams.
      ].

      2.2.5 Compton sources

      Compton x-ray sources work on the principle of inverse Compton scattering (ICS). In 1998, Huang and Ruth proposed a compact laser-electron storage ring (LESR) for electron beam cooling or x-ray generation [
      • Huang Z.
      • Ruth R.D.
      Laser-Electron Storage Ring.
      ]. When a laser beam interacted with an electron bunch, a high-brilliance x-ray beam is generated. In the LESR source operated in transient mode, a 2-J power laser with 3-mm long pulses is scattered off 100 MeV electron bunches with 1.3 × 1010 electrons and a 200 keV photon beam is generated with a flux of 2.6 × 1020 photons/s. A commercial example of such a source, the Compact Light Source (CLS) manufactured by Lyncean Technologies Inc, was operated at 13.5 keV with 3% energy spread to image a moth placed at 10 m from the x-ray source using phase-contrast x-ray imaging [
      • Bech M.
      • Bunk O.
      • David C.
      • Ruth R.
      • Rifkin J.
      • Loewen R.
      • Feidenhans'l R.
      • Pfeiffer F.
      Hard X-ray phase-contrast imaging with the Compact Light Source based on inverse Compton X-rays.
      ]. The BriXS (Bright and compact X-ray Source), used for an RT study presented in section 3.4.3, is a twin Compton x-ray source. The operation of the BriXS source is based on superconductive cavities technology for the electron beam with energy recirculation and on a laser system in Fabry-Pérot cavity at a repetition rate of 100 MHz, producing 20–180 keV x-ray beam [

      Drebot I, Bacci A, Bosotti A, Broggi F, Canella F, Cardarelli P, et al. BriXs ultra high flux inverse compton source based on modified push-pull energy recovery linacs. Instruments. 2019;3:49.

      ].

      3. Applications of kV radiotherapy

      3.1 Clinical applications

      3.1.1 Intraoperative radiotherapy

      Intraoperative radiotherapy (IORT) is a method by which a large dose of radiation is delivered directly to the tumor site. This technique is possible due to direct access to the tumor site during a surgical procedure. Since there is no overlying tissue during IORT, and therefore no need for skin-sparing, kV x-rays are preferred over higher energy photons due to their reduced penetration, which reduces dose to normal tissues beyond the tumor site. IORT was first introduced in the 1960 s and consisted primarily of x-ray tube systems installed in surgery suites [

      Pilar A, Gupta M, Laskar SG, Laskar S. Intraoperative radiotherapy: review of techniques and results. Ecancermedicalscience. 2017;11.

      ]. While not a common method of radiotherapy, due to the surgical requirements, the ability to deliver a large dose directly to the tumor site, as well as the potential for physically spacing surrounding normal tissue from the irradiated area, results in an enhanced therapeutic ratio. However, the increased attenuation of the kV x-rays in bone due to a higher photoelectric cross-section is a disadvantage. Alternatives to kV x-ray IORT include MeV electron therapy, which has the advantage of reduced dose past the tumor site in comparison to kV x-rays. In recent years, advances in technology have resulted in mobile kV x-ray IORT platforms which can provide advanced beam shaping and require less room shielding.
      One novel technology which is well suited to IORT is the miniature x-ray tube. The small size of miniature x-ray tubes allows them to be placed directly within a patient and directly at the site requiring irradiation. The Xoft Axxent miniature x-ray tube has been extensively used for intraoperative treatment of early stage breast cancer [
      • Dickler A.
      • Ivanov O.
      • Francescatti D.
      Intraoperative radiation therapy in the treatment of early-stage breast cancer utilizing xoft axxent electronic brachytherapy.
      ]. CNT-based miniature x-ray tubes in combination with endoscopic interventional technology have also been considered to treat the oral cavity and larynx [
      • Boese A.
      • Johnson F.
      • Ebert T.
      • Mahmoud-Pashazadeh A.
      • Arens C.
      • Friebe M.
      Trans-oral miniature X-ray radiation delivery system with endoscopic optical feedback.
      ,

      Mahmoud-Pashazadeh A, Illanes A, Joseph FJ, van Oepen A, Boese A, Friebe M. Miniature CNT-based X-ray tube: assessment for use in intraoperative radiation therapy: Miniature CNT-based X-ray tube. Current Directions in Biomedical Engineering. 2017;3:643-6.

      ]. IORT technology is an expansive field and dedicated reviews on IORT exist and should be consulted by the reader for further information [

      Pilar A, Gupta M, Laskar SG, Laskar S. Intraoperative radiotherapy: review of techniques and results. Ecancermedicalscience. 2017;11.

      ].

      3.1.2 Superficial radiotherapy

      Kilovoltage x-rays are also used for treatment of superficial cancers [
      • McGregor S.
      • Minni J.
      • Herold D.
      Superficial radiation therapy for the treatment of nonmelanoma skin cancers.
      ]. The energy spectrum of x-rays for this application ranges from 50 to 150 kVp. Since the cancer is superficial, skin-sparing is not generally a restrictive concern in these treatments. Due to the capability of many linacs to generate MeV electrons, it is common for superficial cancers to be treated with MeV electrons rather than kV x-rays which would require additional technology. Further reasons for the use of MeV electrons over kV x-rays are the finite range of electrons, higher dose rates and ease of modelling in modern treatment planning systems.

      3.1.3 Dose enhancement with iodine

      Due to the increased probability of photo-electric interactions, kV x-ray beams have the potential to increase radiation dose to a target loaded with x-ray contrast agents of high atomic number (Z), such as iodine. The increased photo-electric effect occurring in the high-Z material results in the emission of short-range Meitner electrons, more commonly referred to as Auger-electrons first discovered by Lise Meitner in 1922 [
      • Meitner L.
      Über die entstehung der β-strahl-spektren radioaktiver substanzen.
      ]. Such electrons deposit their energy locally in the vicinity of the high-Z material and damage cells loaded with the contrast agent, causing the so-called “radiosensitization.”
      In 1999, two publications investigating the treatment of brain tumors with a modified CT scanner were published [
      • Mesa A.V.
      • Norman A.
      • Solberg T.D.
      • Demarco J.J.
      • Smathers J.B.
      Dose distributions using kilovoltage x-rays and dose enhancement from iodine contrast agents.
      ,
      • Rose J.H.
      • Norman A.
      • Ingram M.
      • Aoki C.
      • Solberg T.
      • Mesa A.
      First radiotherapy of human metastatic brain tumors delivered by a computerized tomography scanner (CTRx).
      ]. Termed CTRx, the CT scanner with a modified pencil-beam collimator was able to function as both a diagnostic and therapeutic system. Mesa et al. used Monte Carlo dose calculations to investigate the dosimetric properties of the CTRx system when used to treat brain tumors injected with iodine contrast for tumor dose enhancement. The energy spectrum used in the simulations was taken from the 140 kVp setting of a GE CT scanner. Skull dose reduction was explored both by means of increasing the concentration of iodine (up to 20 mg/ml) in the tumor as well as the use of three non-coplanar (20°, 0° and −20°) treatment arcs by tilting the gantry. The CTRx treatments were compared to simulated treatments with 10 MV photons. The primary factor in achieving desired dose distributions within the tumor, as well as skull dose reduction, was found to be the concentration of iodine. Higher concentrations resulted in more tumor dose and less skull dose. The highest skull dose was 80% of the maximum dose delivered with 0 mg/ml of iodine and 20% of the maximum dose with 15 mg/ml of iodine contrast. It should be noted that in clinical practice only concentrations of up to 5 mg/ml are routinely achieved using standard intravenous injection techniques. Iodine concentrations above 15 mg/ml were found to result in less homogenous dose distributions within the tumor. The 10 MV treatment was more uniform than the CTRx treatment and also had a smaller skull dose.
      Rose et al. later published the results of a phase I clinical trial in which 8 patients with metastatic brain lesions smaller than 2.5 cm in diameter were treated with a modified CT scanner [
      • Rose J.H.
      • Norman A.
      • Ingram M.
      • Aoki C.
      • Solberg T.
      • Mesa A.
      First radiotherapy of human metastatic brain tumors delivered by a computerized tomography scanner (CTRx).
      ]. The CT scanner was set to 140 kVp for treatment. The patients were injected with iodine contrast and received 3 to 5 weekly fractions of 5 Gy. Each patient had at least one other metastatic brain lesion which did not receive CTRx and served as a control. The patients also received conventional whole brain radiotherapy either before, during or after their CTRx treatment. The authors reported that no patients demonstrated any side effects from the CTRx. Of the eight patients, the irradiated tumors shrank in six cases, remained constant in one case and grew in one case. The control tumors shrank in four cases, grew in two cases and remained constant in one case. The control tumor data for one patient was not available. The authors conclude that CTRx treatment is safe and that an advanced version of the CTRx system could provide diagnostic and therapeutic treatment for select patients.
      Between 1997 and 1999, Weil et al. explored the safety of kV radiotherapy with dose enhancement using mostly directly injected iodine contrast for palliation of advanced cancers of different histologies in a Phase I/II human trial (unpublished data). The study was conducted in a university setting under IRB oversight. Eleven patients with recurrent disease in a variety of sites ranging from extremities to brain and liver were treated via CT-guided intralesional injection of 0.5–4.5 ml of iopromide with 37% w/v iodine using a 120 kVp beam of a radiotherapy simulator (Fig. 1a). The spatial distribution and concentration of iodine contrast in the tumors was derived from the Hounsfield numbers of the CT image in near real time. Thereby, the dose enhancement and dosimetry of the kV beam was optimized for each treatment. Several significant responses were noted with no skin reaction or sequelae attributable to RT. For example, a patient with multiply recurrent non-Hodgkin’s lymphoma on the forehead completely responded to a single 1.5-Gy treatment fraction delivered with contrast-enhanced RT (CERT) and underwent successful reconstructive surgery to cover a skull defect as the tumor had previously penetrated the bone (Fig. 1b-c).
      Figure thumbnail gr1
      Fig. 1Iodine contrast in a metastatic ovarian adenocarcinoma to an axillary lymph node after an intratumoral injection of iopromide (a). Patient with a recurrent Non-Hodgkin's lymphoma penetrating skull before (b) and 9 months after (c) single-fraction 1.5-Gy treatment delivered with a 120 kVp beam after an intratumoral iopromide injection.

      3.2 Veterinary applications

      Radiation therapy options for companion animals include the use of kV x-rays. In 1997, Norman et al. performed a study in which canines diagnosed with brain tumors were treated with the CTRx system mentioned above [
      • Norman A.
      • Ingram M.
      • Skillen R.G.
      • Freshwater D.B.
      • Iwamoto K.S.
      • Solberg T.
      X-ray phototherapy for canine brain masses.
      ]. The study involved 47 dogs from private households. The modified CT scanner was equipped with two separate collimators, the standard imaging collimator and a custom collimator for treatment. The dogs were treated with weekly fractions of 5.6 Gy using 140 kVp x-rays on the CTRx after anesthetization and intravenous injection of iodinated contrast (280 mg/ml). The iodine-based contrast was used both for improved localization of the tumor and a dose enhancement effect. The tumor was located using the diagnostic imaging function of the CTRx after which the lesion was positioned at isocenter. The tumors were treated with the therapy collimator set to a rectangular aperture of 1 cm by 3 cm. The scanner delivered the radiation using a scan speed of 4 s per rotation with a total of 60 rotations per fraction. The gantry was tilted at −20, 0 and 20 degrees for 20 rotations each to deliver non-coplanar beams. The median survival for all treated canines was 133 days. However, the authors note that many owners decided to euthanize their animals early during treatment which decreased the median survival. The median survival of animals that completed treatment (defined as 6 or more fractions) was 231 days. Furthermore, the median survival of the canines who completed treatment was similar to previously published data of canines treated with cobalt therapy and orthovoltage therapy for brain tumors. In comparison, historical studies of the survival of canines with brain tumors suggest a median survival of 6 – 13 days. Autopsies of canines who received CTRx treatment indicated no damage to healthy brain tissue, bone or skin. The authors concluded that CTRx treatment of brain tumors in canines was a safe and effective method.
      Radiotherapy treatment options for companion animals are somewhat limited as linear accelerators are too costly for most veterinary hospitals to own. kV x-ray systems may be more suitable for companion animal radiation therapy due to their lower cost and, in addition, companion animals are smaller than humans and therefore involve much shallower treatment depths, which makes them ideal for lower energy kV x-ray treatment.
      In 2018, Seo et al. published an article investigating the treatment of cancer in companion animals (canines) with 300 and 450 kV x-rays [
      • Seo J.
      • Son J.
      • Cho Y.
      • Park N.
      • Kim D.W.
      • Kim J.
      • Yoon M.
      Kilovoltage radiotherapy for companion animals: dosimetric comparison of 300 kV, 450 kV, and 6 MV X-ray beams.
      ]. The authors used numerical methods to calculate dose distributions using 300 kV, 450 kV and 6 MV photons for brain, nasal cavity, forefoot and abdomen cancers in companion animals. Dose distributions were evaluated by means of DVH, homogeneity index and conformity index. DVH analysis indicated that all three x-ray beam energies yielded comparable dose distributions for brain and nasal tumors. The metrics of homogeneity and conformity index were optimal for the 6 MV x-rays for the nasal, forefoot and abdominal tumors. The authors noted that in general dose to at-risk organs was less for the 6 MV x-rays but was not significantly higher for the kV x-rays.

      3.3 Preclinical applications

      3.3.1 Dose enhancement with gold nanoparticles

      Due to their biocompatibility, gold nanoparticles (AuNP) have been most commonly investigated for RT dose enhancement. Hainfeld et al. in 2004 irradiated tumors of mice with subcutaneous EMT-6 mammary carcinomas after an injection of AuNP with 250 kVp x-rays [
      • Hainfeld J.F.
      • Slatkin D.N.
      • Smilowitz H.M.
      The use of gold nanoparticles to enhance radiotherapy in mice.
      ]. One-year survival of 86% was significantly higher than the 20% survival after radiation alone and 0% with AuNP alone.
      Many following computational, in vitro and in vivo studies demonstrated the NP radiosensitization effect with kV x-ray beams and an overview of the topic can be found in [
      • Haume K.
      • Rosa S.
      • Grellet S.
      • Śmiałek M.A.
      • Butterworth K.T.
      • Solov’yov A.V.
      • Prise K.M.
      • Golding J.
      • Mason N.J.
      Gold nanoparticles for cancer radiotherapy: a review.
      ]. As expected, computational Monte Carlo dose calculation studies clearly show that NP dose enhancement in kV x-ray beams is in the order of 30–40% and it is negligible in the more penetrating clinically relevant MV photon beams. However, Jain et al. demonstrated a 41% and 29% dose enhancement in a 160 kVp and 6 MV x-ray beam for breast cancer MDA-MB-231 cells [
      • Jain S.
      • Coulter J.A.
      • Hounsell A.R.
      • Butterworth K.T.
      • McMahon S.J.
      • Hyland W.B.
      • Muir M.F.
      • Dickson G.R.
      • Prise K.M.
      • Currell F.J.
      • O’Sullivan J.M.
      • Hirst D.G.
      Cell-Specific Radiosensitization by Gold Nanoparticles at Megavoltage Radiation Energies.
      ]. The same study presented no dose enhancement in normal L132, prostate cancer DU145 cells at kV or MV energies, suggesting that radiosensitization even with kV x-ray beams is cell specific. Regardless, kV x-ray beams are most suitable for dose enhancement in combination with NP therapy. Thanks to their shallow tissue penetration, kV x-ray beams in the arrangement of rotational therapy described in section 3.4 might be best suited for optimal results of NP radiosensitization.

      3.3.2 Microbeam radiation therapy (MRT)

      In MRT, radiation is not delivered in the conventional broad-beam pattern, but it is delivered with an array of parallel 100–300 µm wide beams separated by 100–300 µm. Compared to MV beams, kV x-ray beams are better suited for MRT delivery. Due to the short range of secondary electrons, the spatial fractionation of the beam can be preserved in tissue. MRT has been mostly delivered with high-brilliance synchrotron beams at dose rates of > 100 Gy/s and has shown to improve normal tissue sparing while maintaining tumor control [
      • Laissue J.A.
      • Geiser G.
      • Spanne P.O.
      • Dilmanian F.A.
      • Gebbers J.-O.
      • Geiser M.
      • Wu X.-Y.
      • Makar M.S.
      • Micca P.L.
      • Nawrocky M.M.
      • Joel D.D.
      • Slatkin D.N.
      Neuropathology of ablation of rat gliosarcomas and contiguous brain tissues using a microplanar beam of synchrotron-wiggler-generated X rays.
      ]. For example, Dilmanian et al. irradiated rat intracranial 9L gliosarcoma with various arrangements of microbeam size and spacing on the BNL synchrotron beamline with a mean x-ray beam energy of 70 keV. The median survival of a subset of rats was 170 days compared to the previously reported 34-day survival of similar models irradiated with a broad beam [
      • Dilmanian F.A.
      • Button T.M.
      • Duc G.L.
      • Zhong N.
      • Peña L.A.
      • Smith J.A.L.
      • et al.
      Response of rat intracranial 9L gliosarcoma to microbeam radiation therapy.
      ]. In 2003, subcutaneous mouse models of ENT-6 murine mammary carcinoma were irradiated with microbeams and a broad beam on the same beamline using x-ray beams with mean energy of 100 or 118 keV [
      • Dilmanian F.A.
      • Morris G.M.
      • Zhong N.
      • Bacarian T.
      • Hainfeld J.F.
      • Kalef-Ezra J.
      • Brewington L.J.
      • Tammam J.
      • Rosen E.M.
      Murine EMT-6 Carcinoma: High Therapeutic Efficacy of Microbeam Radiation Therapy.
      ]. Tumor ablation rates were higher and normal-tissue toxicity, acute (moist desquamation and epilation) and delayed (failure of hair regrowth), was lower for MRT compared to the broad beam irradiation.
      Hadsell et al. investigated whether high dose-rate MRT can be delivered using a 50 kV anode voltage and 0.5 mA beam current CNT x-ray source [
      • Hadsell M.
      • Cao G.
      • Zhang J.
      • Burk L.
      • Schreiber T.
      • Schreiber E.
      • Chang S.
      • Lu J.
      • Zhou O.
      Pilot study for compact microbeam radiation therapy using a carbon nanotube field emission micro-CT scanner: Pilot study for MRT using CNT micro-CT scanner.
      ]. Based on their experimental pilot study they concluded that the required MRT pattern could be achieved with a CNT x-ray source, however, their setup only allowed for a dose rate of 1.54 mGy/s. The authors predicted that a CNT x-ray source with a 160 kV anode voltage and 1 A beam current would be required for a 100 Gy/s dose rate delivery of MRT. In their follow-up work, the group at University of North Carolina has built a compact prototype MRT system with 160 kV anode voltage and 30 mA beam current and delivered 2–3-beam MRT to mouse models of glioblastoma multiforme (GBM) with a skin dose rate of 1.2 Gy/min [
      • Zhang L.
      • Yuan H.
      • Burk L.M.
      • Inscoe C.R.
      • Hadsell M.J.
      • Chtcheprov P.
      • Lee Y.Z.
      • Lu J.
      • Chang S.
      • Zhou O.
      Image-guided microbeam irradiation to brain tumour bearing mice using a carbon nanotube x-ray source array.
      ]. The evidence of DNA double-strand breaks in the path of the microbeams was demonstrated via γ-H2AX staining. Yuan et al. irradiated mouse GBM models reaching 48 Gy and 72 Gy dose levels with MRT (Fig. 2) versus 10 Gy with broad beam and found similar survival when employing 72-Gy irradiation or 10 Gy broad beam irradiation [

      Yuan H, Zhang L, Frank JE, Inscoe CR, Burk LM, Hadsell M, et al. Treating brain tumor with microbeam radiation generated by a compact carbon-nanotube-based irradiator: initial radiation efficacy study. Radiation research. 2015;184:322-33.

      ]. A review of the group’s MRT work with CNTs can be found in [
      • Zhang L.
      • Yuan H.
      • Inscoe C.
      • Chtcheprov P.
      • Hadsell M.
      • Lee Y.
      • Lu J.
      • Chang S.
      • Zhou O.
      Nanotube x-ray for cancer therapy: a compact microbeam radiation therapy system for brain tumor treatment.
      ].
      Figure thumbnail gr2
      Fig. 2Microbeam profiles using Gafchromic films and γ-H2AX staining on irradiated brain. The beam width (FWHM) was 280 µm at the entrance plane on the top of the mouse brain (panel A) and 380 µm at the exit plane at the bottom of the head (panel B). Normal brain tissue section (panel C) and tumor-bearing mouse brain (panel D) were stained with anti-γ-H2AX antibody at 1 h after microbeam irradiation. Red fluorescence signal indicates the positive expression of γ-H2AX signal. Staining of γ-H2AX demonstrated the microbeam radiation track. The average FWHM of the γ-H2AX staining was 343 µm. (Reprinted with permission from Yuan et al., 2015, © 2020 Radiation Research Society.) (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
      In 2011, Le Duc et al. presented a study combining the delivery of gadolinium-based nanoparticles (GBN) with MRT performed at the ID17 ESRF beamline for the treatment of a rats bearing intracerebral 9L gliosarcoma [
      • Le Duc G.
      • Miladi I.
      • Alric C.
      • Mowat P.
      • Bräuer-Krisch E.
      • Bouchet A.
      • Khalil E.
      • Billotey C.
      • Janier M.
      • Lux F.
      • Epicier T.
      • Perriat P.
      • Roux S.
      • Tillement O.
      Toward an Image-Guided Microbeam Radiation Therapy Using Gadolinium-Based Nanoparticles.
      ]. When MRT irradiation using cross-fired 25-µm microbeams separated by 200 µm gaps was delivered 20 min post GBN injection, median survival increased to 90 days compared to 19 days and 47 days for no treatment and MRT treatment without GBN, respectively. Interestingly, the median survival decreased to 34 days when MRT was delivered only 5 min after the injection of GBN. The authors concluded that the ratio of GBN concentration in the tumor to GBN concentration in normal tissue is a critical radiosensitization parameter rather than GBN concentration in the tumor itself.

      3.3.3 FLASH therapy

      Possibly the most recent application of kV x-rays in RT is the delivery of > 40 Gy/s ultrahigh dose-rate (FLASH) RT. After initial studies in the late 1960 s [

      Todd P, Winchell H, Feola J, Jones G. Inactivation by pulsed high-intensity x-rays of human cells cultured in vitro. RADIATION RESEARCH: RADIATION RESEARCH SOC 2021 SPRING RD, STE 600, OAK BROOK, IL 60521; 1967. p. 644-&.

      ,
      • TOWN C.D.
      Effect of High Dose Rates on Survival of Mammalian Cells.
      ], ultrahigh dose-rate RT was rediscovered in 2014 by Favaudon et al. [

      Favaudon V, Caplier L, Monceau V, Pouzoulet F, Sayarath M, Fouillade C, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Science translational medicine. 2014;6:245ra93-ra93.

      ]. In their study, the authors irradiated thoraxes of mice with a 4.5 MeV electron beam with a dose rate of 40 Gy/s and reported a significant decrease in radiation-induced lung fibrosis when compared to conventional irradiations at a dose rate of 0.03 Gy/s performed with a 137Cs gamma ray source. Most clinical RT is performed with x-ray beams and there was interest to investigate whether the FLASH effect can be also triggered with an x-ray beam. Due to engineering challenges of accelerator-based megavoltage x-ray beam generation at high dose rates, a synchrotron source was chosen for this purpose.
      In 2018, mouse whole brain irradiations (WBI) were performed with the ID17 ESRF beamline using a thin 50-µm wide synchrotron x-ray beam scanned across the mouse brain at an in-slice dose rate of 12,000 Gy/s and a mean brain dose rate of 37 Gy/s [
      • Montay-Gruel Pierre
      • Bouchet Audrey
      • Jaccard Maud
      • Patin David
      • Serduc Raphael
      • Aim Warren
      • Petersson Kristoffer
      • Petit Benoit
      • Bailat Claude
      • Bourhis Jean
      • Bräuer-Krisch Elke
      • Vozenin Marie-Catherine
      X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice.
      ]. The conventional low-dose rate WBI was delivered with the 225 kVp beam on the image-guided small animal radiation research platform (SARRP). After a 10-Gy brain dose, unlike after the conventional SARRP irradiation, no induction in memory deficit was observed after FLASH irradiation and the case for normal tissue sparing of x-ray FLASH was therefore strengthened.
      More recently, Bazalova-Carter and Esplen have proposed that it might be possible to use conventional x-ray tubes to deliver ultrahigh dose-rate RT [
      • Bazalova‐Carter Magdalena
      • Esplen Nolan
      On the capabilities of conventional x‐ray tubes to deliver ultra‐high (FLASH) dose rates.
      ]. Dose rates of > 40 Gy/s can be achieved for an unfiltered x-ray beam at shallow < 1 mm depth in a sample placed at the x-ray tube surface. The FLASH effect, however, has not yet been investigated with this continuous x-ray source. A comprehensive review of FLASH therapy can be found in [
      • Esplen Nolan
      • Mendonca Marc
      • Bazalova-Carter Magdalena
      Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review.
      ].

      3.4 Proof-of-principle and computational models

      3.4.1 Kilovoltage external beam radiotherapy (kV-EBRT)

      Dedicated breast CT (bCT) platforms were first developed in 2008 in order to improve the diagnosis of breast cancer [
      • Lindfors Karen K.
      • Boone John M.
      • Nelson Thomas R.
      • Yang Kai
      • Kwan Alexander L.C.
      • Miller DeWitt F.
      Dedicated Breast CT: Initial Clinical Experience.
      ]. In 2012, Prionas et al. published work investigating rotational kilovoltage external beam radiation therapy (kV-EBRT) of breast cancer with 320 kVp x-rays using a bCT system [
      • Prionas Nicolas D.
      • McKenney Sarah E.
      • Stern Robin L.
      • Boone John M.
      Kilovoltage Rotational External Beam Radiotherapy on a Breast Computed Tomography Platform: A Feasibility Study.
      ]. kV-EBRT was simulated using Monte Carlo methods for both cylindrical phantoms and breast CT data. The beam was simulated and 2D and 3D dose distributions were generated with a range of collimation settings. The smallest diameter was 1 cm and the largest was 14 cm in the XY plane. Z collimation was 0.1 cm for 2D profiles and 2 cm for 3D dose distribution. Smaller XY collimation settings demonstrated centrally peaked dose deposition with the 1 cm setting depositing as little as 7% dose at the periphery of the cylindrical phantom. The 14 cm collimation setting resulting in a centrally cupped dose distribution.
      Experimental validation was performed using a CT scanner operating at a tube voltage of 120 kV. Dose measurements of the radial dose distributions in a 14-cm diameter polyethylene cylinder were made using an ionization chamber. These measurements were compared to Monte Carlo simulated radial dose distributions with a matched energy spectrum. The mean percentage of difference between the simulated and measured dose distributions was 10.5% +- 7.7% for the 1-cm collimated beam and 2.8% +- 2.3% for the 14-cm collimated beam.
      The authors also investigated whole breast irradiation using phantoms, dose “painting” and the treatment of patients with Monte Carlo simulations. Of particular note are the simulations of a breast cancer patient imaged with iodine contrast (Fig. 3). Using a 2 cm collimated beam, the calculated dose distribution shows that the lesion was contained within the 90% isodose line and that the skin received 10% of the maximum dose.
      Figure thumbnail gr3
      Fig. 3Contrast-enhanced breast computed tomography image containing enhancing invasive mammary carcinoma used to simulate rotational kilovoltage external beam radiotherapy with 2-cm x–y collimation centered at lesion (a). Resulting dose distribution (b) and fusion image (c). (Reprinted with permission from Prionas et al., 2012.)
      A number of follow-up studies on breast kV-EBRT were published by a group at the University of Naples. In 2019, Buonano et al. published a similar study to the one by Prionas et al., in which they investigated breast phantom dose distributions achieved by 150, 300, and 320 kVp x-ray beams generated by an x-ray tube by means of MC simulations and experiments [
      • Buonanno F.
      • Sarno A.
      • De Lucia P.A.
      • Di Lillo F.
      • Masi M.
      • Di Franco F.
      • Mettivier G.
      • Russo P.
      Rotational radiotherapy of breast cancer with polyenergetic kilovoltage X-ray beams: An experimental and Monte Carlo phantom study.
      ]. Rotational kV-EBRT permitted a periphery-to-center dose ratio from 13% to 9% in homogeneous breast phantoms of 14-cm in diameter for beams of ~ 1 cm in size. Earlier the group investigated breast EBRT with a synchrotron beam termed synchrotron radiation rotational radiotherapy (SR3T) by means of MC simulations and experiments with the same 14-cm diameter phantom [
      • Di Lillo Francesca
      • Mettivier Giovanni
      • Sarno Antonio
      • Castriconi Roberta
      • Russo Paolo
      Towards breast cancer rotational radiotherapy with synchrotron radiation.
      ,

      Di Lillo F, Mettivier G, Castriconi R, Sarno A, Stevenson AW, Hall CJ, et al. Synchrotron radiation external beam rotational radiotherapy of breast cancer: proof of principle. Journal of synchrotron radiation. 2018;25:857-68.

      ]. They performed Geant4 simulations and measurements with Gafchromic films, ionization chambers and TLD of phantom dose delivered with monoenergetic 60–175 keV synchrotron x-ray beams at the Australian Synchrotron IMBL. They calculated and measured a tumor-to-skin absorbed dose ratio ranging from about 7:1 (at 60 keV photon energy) to about 10:1 (at 175 keV) for a 1.5-cm wide beam, which was deemed sufficient for skin sparing during radiotherapy.

      3.4.2 Kilovoltage x-ray beam arc therapy

      In the EBCT and SBDX x-ray sources electron beam energies of 80–140 keV are used, resulting in an x-ray beam with shallow dose deposition and not suitable for radiotherapy. To overcome this shortcoming, a design of a stationary-anode source was modeled for the delivery of kilovoltage arc therapy (KVAT) to deep-seated tumors employing non-coplanar kV beams (Fig. 4). First, a source design study was performed to optimize the delivery of conformal radiation to a 4-cm diameter, 10-cm deep lesion by varying beam energy, x-ray beam filtration, target thickness and collimator design [
      • Bazalova-Carter Magdalena
      • Weil Michael D.
      • Breitkreutz Dylan Yamabe
      • Wilfley Brian P.
      • Graves Edward E.
      Feasibility of external beam radiation therapy to deep-seated targets with kilovoltage x-rays.
      ]. Several follow-up studies published by Breitkreutz et al. are described briefly below.
      Figure thumbnail gr4
      Fig. 4A schematic drawing of the proposed KVAT x-ray source (a). The x-ray source was designed to treat a 4-cm diameter at 10-cm depth in a 40-cm diameter cylinder using arc therapy (b). (Reprinted with permission from Bazalova-Carter et al., 2017.)
      The first publication by Breitkreutz et al. involved Monte Carlo simulations of a novel geometry, scanning electron-beam x-ray tube design which produces a non-coplanar linear array of x-ray beamlets with an energy of 200 kV. The system was designed with a gantry to allow 360° rotation around the patient. The scanning nature of the x-ray tube allows for adjustment of beamlet dwell times to produce true intensity modulated therapy. Proof-of-principle simulations were first used with a transmission anode design. These simulations for cylindrical water phantoms indicated that a skin sparing effect could be produced while delivering a dose rate up to 345 cGy/min for a 1-cm diameter spherical target, with dose rates decreasing at greater target size and depth [
      • Breitkreutz Dylan Y.
      • Weil Michael D.
      • Zavgorodni Sergei
      • Bazalova-Carter Magdalena
      Monte Carlo simulations of a kilovoltage external beam radiotherapy system on phantoms and breast patients.
      ].
      Further work employed Monte Carlo simulations to investigate the treatment of lung, breast and prostate cancer using inversely optimized treatment planning [
      • Breitkreutz Dylan Y.
      • Renaud Marc‐André
      • Seuntjens Jan
      • Weil Michael D.
      • Zavgorodni Sergei
      • Bazalova‐Carter Magdalena
      Inverse optimization of low‐cost kilovoltage x‐ray arc therapy plans.
      ]. These simulated kV treatments were compared to MV photon treatments for reference. Dose constraints taken from RTOG reports were used to evaluate the breast, lung and prostate KVAT plans and were met in all cases. However, the KVAT prostate plan likely was untreatable due to lengthy delivery times and suboptimal dosimetry for healthy tissues. This was due primarily to the size of the patient, the less penetrating radiation and avoidance of bony anatomy. One limitation of this study was the use of idealized, spherical lesions as tumor targets.
      Lastly, Breitkreutz et al. simulated KVAT of lung tumors with realistic lung lesions based on CT scan data from three lung cancer patients [

      Breitkreutz DY, Renaud M-A, Weil MD, Zavgorodni S, Han J, Baxter H, et al. Monte Carlo calculated kilovoltage x-ray arc therapy plans for three lung cancer patients. Biomedical Physics & Engineering Express. 2019;5:065022.

      ]. In this study, the KVAT model was modified to a reflection anode. The location of the lung tumors varied in each patient. It was found that for a patient with a tumor directly next to the chest wall that the rib dose may exceed dose constraints. All other dose constraints were met for each patient.
      In summary, the KVAT x-ray source has been developed to address kV shortcomings by rapidly distributing multiple kV beams via electronic manipulation rather than with conventional mechanical control. It focuses dose deposition while dispersing entrance dose and overrides attenuation constraints of lower energy x-rays and thereby enables less expensive kV x-ray technology for radiotherapy treatment.
      The KVAT system is the main component of the linear converging radiotherapy system (LCRS) under development by PrecisionRT (Las Vegas, NV, USA). The LCRS will be capable of simultaneous tomosynthesis imaging of real-time tumor motion and radiation therapy delivery [

      Stalbaum T, Boyd D, Weil M, Chen H, Plies M, Song S, et al. 200 kV x-ray source for radiotherapy and imaging: preliminary results and discussion. Medical Imaging 2020: Physics of Medical Imaging: International Society for Optics and Photonics; 2020. p. 113123I.

      ]. Thanks to the addition of a detector, the source can create CT scans and image in real-time during therapy without a separate CT unit. Simulations have demonstrated the design can plan and complete effective therapy within one hour. Most of these models demonstrated equivalent performance to MV volumetric modulated arc therapy (VMAT) at 10% of the cost per system and likely reducing cost per patient to US$100 or less.
      It will be possible to house the LCRS in a conventional 3.2-mm thick, lead-lined room found in most clinics or in a mobile trailer. The treatment planning software currently under development will be capable of optimizing a radiosurgical treatment prescription. The patient will be immobilized and aligned per conventional techniques. Thanks to machine learning (ML), treatment planning and delivery will be performed in a matter of minutes, slightly more time than it takes to do a diagnostic scan [

      Zankowski C, Adelsheim C, Pyyry J, Kuusela E. Radiotherapy treatment planning using artificial intelligence (ai) engines. Google Patents; 2019.

      ].

      3.4.3 BriXS source for rotational breast radiotherapy

      In 2019, Sarno et al. published work investigating the use of a BriXS twin Compton pulsed x-ray source as an alternative source for the treatment of breast cancer [

      Sarno A, Mettivier G, Russo P, Drebot I, Petrillo V, Bacci A, et al. Kilovoltage rotational radiotherapy of breast cancer with the BriXS source. Journal of Instrumentation. 2020;15:C05012.

      ]. The BriXS source has a physical footprint of 40 m by 20 m and produces an x-ray spectrum with a mean energy of 88 keV with a high photon flux suitable for radiation therapy. The Geant4 Monte Carlo software was used to simulate rotational radiotherapy of breast cancer using the BriXS source. The breast was modelled as a 14 cm diameter cylinder with a 1 cm circular lesion at the center and off-axis. The modeled beam was rotated 360° around the cylinder to simulate a prone breast geometry. The main metric used to evaluate the simulated dose distributions was skin-to-tumor dose ratio. The skin-to-tumor dose ratio was 7% for the central lesion and 36% for the off-axis lesion.

      3.4.4 Kilovoltage intensity modulated radiotherapy

      In 2019, a group at Wayne State University presented a MC computer model and a proof-of-principle experiment for a kilovoltage intensity modulated radiotherapy (IMRT) for the treatment of GBM by means of CERT with iodine [

      Loughery B, Halford R, Dess K, Rakowski J, Cox M, Koh J, et al. A proof of concept kilovoltage intensity modulated radiotherapy platform for the treatment of glioblastoma multiforme. Biomedical Physics & Engineering Express. 2019;5:055013.

      ]. They used Geant4 to simulate dose deposition in a 15-cm diameter spherical water phantom with a 1.4-cm thick spherical shell mimicking the skull and a target of 1.8-cm in diameter filled with 10 mg/mL of iodine. A CERT IMRT treatment with 17 heavily tungsten-filtered 120 kVp coplanar beams, the dose to the target was by a factor of 3–4 higher than dose to the skull. The proof-of-principle experiment was performed with an in-house multi-rod collimator using a set of deliveries and phantoms with radiochromic films. A good agreement between a 3D conformal radiotherapy plan delivery and its MC dose calculation was found. The main limitation of the proof-of-principle system was found in the delivery time, that was from 3 to 20 h long for the IMRT plans.

      4. Benefits and drawbacks of kV radiotherapy

      While there are benefits of kV radiotherapy listed in section 1.2, there are also practical drawbacks to the use of kV x-rays for the treatment of deep-seated tumors. The benefits and drawbacks are briefly summarized in qualitative and quantitative terms in Table 1.
      Table 1Summary of benefits and drawbacks of kV RT compared to MV RT.
      kV RT (200 kV)MV RT (6 MV)
      Skin dosehigh(~100% of Dmax)low(40% of Dmax)
      Dose fall offfast(20% of Dmax at 10 cm depth)slow(70% of Dmax at 10 cm depth)
      Range of secondary e-short(1–2 mm)long(~1–2 cm)
      Treatment timelong(~20 min)short(~2 min)
      Shieldingthin(~2 mm lead)thick(~2 m concrete)
      Machine costlow(US$200,000)high(US$3,000,000)
      The first consideration is the limitation of kV x-ray beam penetration. Kilovoltage x-ray beams are characterized by no build-up and therefore maximum dose deposition Dmax occurs in a patient’s skin. This is problematic as any dose delivered to a target below the surface will necessitate a larger dose to the tissue above it, which can be problematic unless beams from multiple directions intersecting at the target are used. Secondary electrons produced by kV x-rays have less energy and a shorter range than those created by MV x-rays resulting in a smaller percentage of the maximum dose being delivered at larger depths. This fact compounds the problem of delivering dose to a lesion below the surface. On the other hand, the shorter range of secondary electrons is responsible for the sharper beam penumbra in kV x-ray beams compared to MV x-rays. When compared to MV x-rays the faster fall-off of kV x-ray beam dose results in lower exit dose. The lack of penetration of kV x-rays is the primary reason why novel kV-based technologies use more complex source geometry and delivery techniques to reduce surface dose.
      Treatment times of kV x-ray sources for deep-seated lesions can also be problematic. Due to the physics of the bremsstrahlung process, a lower percentage of the kV electrons incident on the bremsstrahlung target is converted into photons compared to MV electrons resulting in the generation of a large amount of heat. Ultimately, there is a practical limit to the electron current at which an x-ray tube may be safely operated, which limits its achievable dose rate. This causes problems for patient treatment as delivery times will be longer than systems using MV x-rays. Longer treatment times will result in less patient comfort and problems associated with patient motion during treatment, if not addressed by real-time imaging.

      5. Other considerations

      Kilovoltage x-rays have been used for the treatment of cancer alongside MV x-rays for decades. For treatments of deep-seated tumors, a number of challenges for kV therapy compared to MV therapy exist, such as the increased treatment time, dose calculation complexity, potential differences in radiobiology between kV and MV beams. Some of these topics for external beam kV RT are discussed here.

      5.1 Clinical consideration

      Consideration of normal tissue and tumor control probability can be derived from the hypo-fractionated protocols conventionally employed in the clinic [
      • Bentzen Søren M.
      • Constine Louis S.
      • Deasy Joseph O.
      • Eisbruch Avi
      • Jackson Andrew
      • Marks Lawrence B.
      • Ten Haken Randall K.
      • Yorke Ellen D.
      Quantitative Analyses of Normal Tissue Effects in the Clinic (QUANTEC): An Introduction to the Scientific Issues.
      ]. For example, the presence of a higher cross-section of cortical bone to kV x-rays might limit use of certain beam angles when planned per accepted tolerances. Even though it is mostly accepted that cell killing potential for kV and MV x-rays is similar [
      • Icrp I.
      Publication 61: Annual limits on intake of radionuclides for workers based on the 1990 recommendations.
      ], some literature suggests differences in the radiobiology of kV and MV beams exist [
      • Amols H.I.
      • Lagueux B.
      • Cagna D.
      Radiobiological Effectiveness (RBE) of Megavoltage X-Ray and Electron Beams in Radiotherapy.
      ]. There is no compelling support in the literature for significant differences in the radiobiology of the respective beam energies. Therefore, the dose effects are considered similar, except for the significant enhancement possible in the presence of high-Z contrast agents described in sections 3.1.3 and 3.3.1.

      5.2 Economic considerations

      There are three macro trends in healthcare that could affect use of a kV treatment system. First, cancer therapy is a growing concern as the populations in developing countries age. Longer life spans and changes with economic advances are associated with increased incidence of cancer, e.g., lung cancer is rising dramatically in Asia due to high tobacco consumption. At the same time, since standard treatment is costly and complex, the additional population burdens on advanced care could make it more inaccessible in the majority of developing economies where finances and infrastructure are limited. Second, the consolidation and squeezing of hospitals and third-party payers (insurers, businesses and government) caused by the 2020 pandemic has decreased cases and income for technical and professional oncology billing. If this trend continues, it will put pressure on a portion of the market and some facilities may have to forgo high-end construction and equipment. Finally, the successful introduction of immuno-oncologic and targeted drug therapies is making significant inroads in cancer care, and is also of import in delivering palliative care. Given their improved prospects to slow a tumor's growth, a patient might live with a protracted malignancy, which requires additional loco-regional control. Thus, if a new cancer therapy induces chronic or stable disease, such disease status can potentially lead to more treatments with kV radiotherapy for adjuvant or palliative care.
      Of note, some predictions can be made in terms of the number of patients treated by converging kV teletherapy, e.g. the LCRS source, in the USA thanks to data collected by the National Cancer Institute (NCI) [
      • Noone A.
      • Howlader N.
      • Krapcho M.
      • Miller D.
      • Brest A.
      • Yu M.
      • et al.
      SEER cancer statistics review, 1975–2015.
      ]. Previous in silico studies have demonstrated that lung and breast are suitable sites for the effective application of converging kV teletherapy due to their reduced attenuation of the beams [
      • Breitkreutz Dylan Y.
      • Renaud Marc‐André
      • Seuntjens Jan
      • Weil Michael D.
      • Zavgorodni Sergei
      • Bazalova‐Carter Magdalena
      Inverse optimization of low‐cost kilovoltage x‐ray arc therapy plans.
      ]. In addition, though the increased dose to bone might limit dose escalation for most treatment sites, it can be exploited for the focused kV treatments of bone metastases. Under the assumption that approximately 50% of cancer patients receive conventional radiotherapy for local control, and an estimated 50–90% likelihood of referrals, it can be predicted that in the USA alone, about 460,000 patients with bone, breast and lung cancer might be treated annually (Table 2).
      Table 2Projections on annual kV teletherapy cases for the USA.
      SiteTotal Cases
      National Cancer Institute, SEER Cancer Statistics Review [61].
      Treatable fraction
      Treatable fraction based on advanced stage, technical feasibility and referral patterns.
      Projected # patients
      Bone, secondary208,0000.90187,000
      Breast266,0000.50133,000
      Lung, primary234,0000.60140,000
      Total708,000460,000
      1 National Cancer Institute, SEER Cancer Statistics Review
      • Noone A.
      • Howlader N.
      • Krapcho M.
      • Miller D.
      • Brest A.
      • Yu M.
      • et al.
      SEER cancer statistics review, 1975–2015.
      .
      2 Treatable fraction based on advanced stage, technical feasibility and referral patterns.

      6. Conclusions

      In this review article, we have provided a brief overview of a number of radiotherapy techniques employing kilovoltage x-rays for treatments of cancer. We summarized the potential benefits and drawbacks of the currently investigated treatment modalities and discussed the importance of developing low-cost radiotherapy.

      Acknowledgements

      The authors would like to acknowledge the support of the Canada Research Chairs program and the Natural Sciences and Engineering Research Council ( NSERC ).

      References

        • Laissue J.A.
        • Blattmann H.
        • Slatkin D.N.
        Alban Köhler (1874-1947): Erfinder der Gittertherapie.
        Zeitschrift für Medizinische Physik. 2012; 22: 90-99https://doi.org/10.1016/j.zemedi.2011.07.002
        • Kerr D.J.
        • Midgley R.
        Can We Treat Cancer for a Dollar a Day? Guidelines for Low-Income Countries.
        N Engl J Med. 2010; 363: 801-803https://doi.org/10.1056/NEJMp1002812
        • Stathakis S.
        The Physics of Radiation Therapy.
        Med. Phys. 2010; 37: 1374-1375https://doi.org/10.1118/1.3319185
        • Rand R.E.
        Scanning electron beam computed tomography scanner with ion aided focusing.
        Google Patents. 1985;
      1. Van Lysel MS, Solomon EG, Wilfley BP, Dutta A, Speidel MA. Performance assessment of the scanning beam digital X-ray (SBDX) system. Proceedings of the SPIE - The International Society for Optical Engineering. 1997;3032:161-70.

        • Speidel M.A.
        • Wilfley B.P.
        • Star-Lack J.M.
        • Heanue J.A.
        • Van Lysel M.S.
        Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography: SBDX technology for interventional cardiac angiography.
        Med. Phys. 2006; 33: 2714-2727https://doi.org/10.1118/1.2208736
        • Dinsmore M.
        • Harte K.J.
        • Sliski A.P.
        • Smith D.O.
        • Nomikos P.M.
        • Dalterio M.J.
        • Boom A.J.
        • Leonard W.F.
        • Oettinger P.E.
        • Yanch J.C.
        A new miniature x-ray source for interstitial radiosurgery: Device description.
        Med. Phys. 1996; 23: 45-52https://doi.org/10.1118/1.597790
        • Rivard M.J.
        • Davis S.D.
        • DeWerd L.A.
        • Rusch T.W.
        • Axelrod S.
        Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent X-Ray Source: An electronic brachytherapy sourcea): Dosimetry parameters for the Xoft Axxent X-Ray Source.
        Med. Phys. 2006; 33: 4020-4032https://doi.org/10.1118/1.2357021
        • Lei W.
        • Li C.
        • Cole M.T.
        • Qu K.e.
        • Ding S.
        • Zhang Y.
        • Warner J.H.
        • Zhang X.
        • Wang B.
        • Milne W.I.
        A graphene-based large area surface-conduction electron emission display.
        Carbon. 2013; 56: 255-263https://doi.org/10.1016/j.carbon.2013.01.004
        • Heo S.H.
        • Ihsan A.
        • Cho S.O.
        Transmission-type microfocus x-ray tube using carbon nanotube field emitters.
        Appl. Phys. Lett. 2007; 90: 183109https://doi.org/10.1063/1.2735549
        • Parmee R.J.
        • Collins C.M.
        • Milne W.I.
        • Cole M.T.
        X-ray generation using carbon nanotubes.
        Nano Convergence. 2015; 2https://doi.org/10.1186/s40580-014-0034-2
        • Kim H.J.
        • Ha J.M.
        • Heo S.H.
        • Cho S.O.
        Small-Sized Flat-Tip CNT Emitters for Miniaturized X-Ray Tubes.
        J Nanomater. 2012; 2012: 1-6https://doi.org/10.1155/2012/854602
      2. Serduc R, Bräuer-Krisch E, Siegbahn EA, Bouchet A, Pouyatos B, Carron R, et al. High-precision radiosurgical dose delivery by interlaced microbeam arrays of high-flux low-energy synchrotron x-rays. PLoS ONE. 2010;5:1-12.

        • Bräuer-Krisch E.
        • Adam J.-F.
        • Alagoz E.
        • Bartzsch S.
        • Crosbie J.
        • DeWagter C.
        • Dipuglia A.
        • Donzelli M.
        • Doran S.
        • Fournier P.
        • Kalef-Ezra J.
        • Kock A.
        • Lerch M.
        • McErlean C.
        • Oelfke U.
        • Olko P.
        • Petasecca M.
        • Povoli M.
        • Rosenfeld A.
        • Siegbahn E.A.
        • Sporea D.
        • Stugu B.
        Medical physics aspects of the synchrotron radiation therapies: Microbeam radiation therapy (MRT) and synchrotron stereotactic radiotherapy (SSRT).
        Physica Med. 2015; 31: 568-583https://doi.org/10.1016/j.ejmp.2015.04.016
        • Laissue J.A.
        • Bartzsch S.
        • Blattmann H.
        • Bräuer-Krisch E.
        • Bravin A.
        • Dalléry D.
        • Djonov V.
        • Hanson A.L.
        • Hopewell J.W.
        • Kaser-Hotz B.
        • Keyriläinen J.
        • Laissue P.P.
        • Miura M.
        • Serduc R.
        • Siegbahn A.E.
        • Slatkin D.N.
        Response of the rat spinal cord to X-ray microbeams.
        Radiother Oncol. 2013; 106: 106-111https://doi.org/10.1016/j.radonc.2012.12.007
        • Huang Z.
        • Ruth R.D.
        Laser-Electron Storage Ring.
        Phys Rev Lett. 1998; 80: 976-979https://doi.org/10.1103/PhysRevLett.80.976
        • Bech M.
        • Bunk O.
        • David C.
        • Ruth R.
        • Rifkin J.
        • Loewen R.
        • Feidenhans'l R.
        • Pfeiffer F.
        Hard X-ray phase-contrast imaging with the Compact Light Source based on inverse Compton X-rays.
        J Synchrotron Rad. 2009; 16: 43-47https://doi.org/10.1107/S090904950803464X
      3. Drebot I, Bacci A, Bosotti A, Broggi F, Canella F, Cardarelli P, et al. BriXs ultra high flux inverse compton source based on modified push-pull energy recovery linacs. Instruments. 2019;3:49.

      4. Pilar A, Gupta M, Laskar SG, Laskar S. Intraoperative radiotherapy: review of techniques and results. Ecancermedicalscience. 2017;11.

        • Dickler A.
        • Ivanov O.
        • Francescatti D.
        Intraoperative radiation therapy in the treatment of early-stage breast cancer utilizing xoft axxent electronic brachytherapy.
        World J Surg Onc. 2009; 7https://doi.org/10.1186/1477-7819-7-24
        • Boese A.
        • Johnson F.
        • Ebert T.
        • Mahmoud-Pashazadeh A.
        • Arens C.
        • Friebe M.
        Trans-oral miniature X-ray radiation delivery system with endoscopic optical feedback.
        Int J CARS. 2017; 12: 1995-2002https://doi.org/10.1007/s11548-017-1601-x
      5. Mahmoud-Pashazadeh A, Illanes A, Joseph FJ, van Oepen A, Boese A, Friebe M. Miniature CNT-based X-ray tube: assessment for use in intraoperative radiation therapy: Miniature CNT-based X-ray tube. Current Directions in Biomedical Engineering. 2017;3:643-6.

        • McGregor S.
        • Minni J.
        • Herold D.
        Superficial radiation therapy for the treatment of nonmelanoma skin cancers.
        J Clinic Aesth Dermatol. 2015; 8: 12
        • Meitner L.
        Über die entstehung der β-strahl-spektren radioaktiver substanzen.
        Zeitschrift für Physik. 1922; 9: 131-144
        • Mesa A.V.
        • Norman A.
        • Solberg T.D.
        • Demarco J.J.
        • Smathers J.B.
        Dose distributions using kilovoltage x-rays and dose enhancement from iodine contrast agents.
        Phys Med Biol. 1999; 44: 1955-1968https://doi.org/10.1088/0031-9155/44/8/308
        • Rose J.H.
        • Norman A.
        • Ingram M.
        • Aoki C.
        • Solberg T.
        • Mesa A.
        First radiotherapy of human metastatic brain tumors delivered by a computerized tomography scanner (CTRx).
        Int J Rad Oncol Biol Phys. 1999; 45: 1127-1132https://doi.org/10.1016/S0360-3016(99)00347-8
        • Norman A.
        • Ingram M.
        • Skillen R.G.
        • Freshwater D.B.
        • Iwamoto K.S.
        • Solberg T.
        X-ray phototherapy for canine brain masses.
        Radiat. Oncol. Investig. 1997; 5: 8-14https://doi.org/10.1002/(SICI)1520-6823(1997)5:1<8::AID-ROI2>3.0.CO;2-1
        • Seo J.
        • Son J.
        • Cho Y.
        • Park N.
        • Kim D.W.
        • Kim J.
        • Yoon M.
        Kilovoltage radiotherapy for companion animals: dosimetric comparison of 300 kV, 450 kV, and 6 MV X-ray beams.
        J Vet Sci. 2018; 19: 550https://doi.org/10.4142/jvs.2018.19.4.550
        • Hainfeld J.F.
        • Slatkin D.N.
        • Smilowitz H.M.
        The use of gold nanoparticles to enhance radiotherapy in mice.
        Phys Med Biol. 2004; 49: N309-N315https://doi.org/10.1088/0031-9155/49/18/N03
        • Haume K.
        • Rosa S.
        • Grellet S.
        • Śmiałek M.A.
        • Butterworth K.T.
        • Solov’yov A.V.
        • Prise K.M.
        • Golding J.
        • Mason N.J.
        Gold nanoparticles for cancer radiotherapy: a review.
        Cancer Nano. 2016; 7https://doi.org/10.1186/s12645-016-0021-x
        • Jain S.
        • Coulter J.A.
        • Hounsell A.R.
        • Butterworth K.T.
        • McMahon S.J.
        • Hyland W.B.
        • Muir M.F.
        • Dickson G.R.
        • Prise K.M.
        • Currell F.J.
        • O’Sullivan J.M.
        • Hirst D.G.
        Cell-Specific Radiosensitization by Gold Nanoparticles at Megavoltage Radiation Energies.
        Int J Rad Oncol Biol Phys. 2011; 79: 531-539https://doi.org/10.1016/j.ijrobp.2010.08.044
        • Laissue J.A.
        • Geiser G.
        • Spanne P.O.
        • Dilmanian F.A.
        • Gebbers J.-O.
        • Geiser M.
        • Wu X.-Y.
        • Makar M.S.
        • Micca P.L.
        • Nawrocky M.M.
        • Joel D.D.
        • Slatkin D.N.
        Neuropathology of ablation of rat gliosarcomas and contiguous brain tissues using a microplanar beam of synchrotron-wiggler-generated X rays.
        Int. J. Cancer. 1998; 78: 654-660https://doi.org/10.1002/(SICI)1097-0215(19981123)78:5<654::AID-IJC21>3.0.CO;2-L
        • Dilmanian F.A.
        • Button T.M.
        • Duc G.L.
        • Zhong N.
        • Peña L.A.
        • Smith J.A.L.
        • et al.
        Response of rat intracranial 9L gliosarcoma to microbeam radiation therapy.
        Neuro Oncol. 2002; 4: 26-38
        • Dilmanian F.A.
        • Morris G.M.
        • Zhong N.
        • Bacarian T.
        • Hainfeld J.F.
        • Kalef-Ezra J.
        • Brewington L.J.
        • Tammam J.
        • Rosen E.M.
        Murine EMT-6 Carcinoma: High Therapeutic Efficacy of Microbeam Radiation Therapy.
        Radiat Res. 2003; 159: 632-641https://doi.org/10.1667/0033-7587(2003)159[0632:MECHTE]2.0.CO;2
        • Hadsell M.
        • Cao G.
        • Zhang J.
        • Burk L.
        • Schreiber T.
        • Schreiber E.
        • Chang S.
        • Lu J.
        • Zhou O.
        Pilot study for compact microbeam radiation therapy using a carbon nanotube field emission micro-CT scanner: Pilot study for MRT using CNT micro-CT scanner.
        Med Phys. 2014; 41: 061710https://doi.org/10.1118/1.4873683
        • Zhang L.
        • Yuan H.
        • Burk L.M.
        • Inscoe C.R.
        • Hadsell M.J.
        • Chtcheprov P.
        • Lee Y.Z.
        • Lu J.
        • Chang S.
        • Zhou O.
        Image-guided microbeam irradiation to brain tumour bearing mice using a carbon nanotube x-ray source array.
        Phys Med Biol. 2014; 59: 1283-1303https://doi.org/10.1088/0031-9155/59/5/1283
      6. Yuan H, Zhang L, Frank JE, Inscoe CR, Burk LM, Hadsell M, et al. Treating brain tumor with microbeam radiation generated by a compact carbon-nanotube-based irradiator: initial radiation efficacy study. Radiation research. 2015;184:322-33.

        • Zhang L.
        • Yuan H.
        • Inscoe C.
        • Chtcheprov P.
        • Hadsell M.
        • Lee Y.
        • Lu J.
        • Chang S.
        • Zhou O.
        Nanotube x-ray for cancer therapy: a compact microbeam radiation therapy system for brain tumor treatment.
        Expert Rev Anticancer Ther. 2014; 14: 1411-1418https://doi.org/10.1586/14737140.2014.978293
        • Le Duc G.
        • Miladi I.
        • Alric C.
        • Mowat P.
        • Bräuer-Krisch E.
        • Bouchet A.
        • Khalil E.
        • Billotey C.
        • Janier M.
        • Lux F.
        • Epicier T.
        • Perriat P.
        • Roux S.
        • Tillement O.
        Toward an Image-Guided Microbeam Radiation Therapy Using Gadolinium-Based Nanoparticles.
        ACS Nano. 2011; 5: 9566-9574https://doi.org/10.1021/nn202797h
      7. Todd P, Winchell H, Feola J, Jones G. Inactivation by pulsed high-intensity x-rays of human cells cultured in vitro. RADIATION RESEARCH: RADIATION RESEARCH SOC 2021 SPRING RD, STE 600, OAK BROOK, IL 60521; 1967. p. 644-&.

        • TOWN C.D.
        Effect of High Dose Rates on Survival of Mammalian Cells.
        Nature. 1967; 215: 847-848https://doi.org/10.1038/215847a0
      8. Favaudon V, Caplier L, Monceau V, Pouzoulet F, Sayarath M, Fouillade C, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Science translational medicine. 2014;6:245ra93-ra93.

        • Montay-Gruel Pierre
        • Bouchet Audrey
        • Jaccard Maud
        • Patin David
        • Serduc Raphael
        • Aim Warren
        • Petersson Kristoffer
        • Petit Benoit
        • Bailat Claude
        • Bourhis Jean
        • Bräuer-Krisch Elke
        • Vozenin Marie-Catherine
        X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice.
        Radiother Oncol. 2018; 129: 582-588https://doi.org/10.1016/j.radonc.2018.08.016
        • Bazalova‐Carter Magdalena
        • Esplen Nolan
        On the capabilities of conventional x‐ray tubes to deliver ultra‐high (FLASH) dose rates.
        Med. Phys. 2019; 46: 5690-5695https://doi.org/10.1002/mp.13858
        • Lindfors Karen K.
        • Boone John M.
        • Nelson Thomas R.
        • Yang Kai
        • Kwan Alexander L.C.
        • Miller DeWitt F.
        Dedicated Breast CT: Initial Clinical Experience.
        Radiology. 2008; 246: 725-733https://doi.org/10.1148/radiol.2463070410
        • Prionas Nicolas D.
        • McKenney Sarah E.
        • Stern Robin L.
        • Boone John M.
        Kilovoltage Rotational External Beam Radiotherapy on a Breast Computed Tomography Platform: A Feasibility Study.
        Int J Rad Oncol Biol Phys. 2012; 84: 533-539https://doi.org/10.1016/j.ijrobp.2011.12.042
        • Buonanno F.
        • Sarno A.
        • De Lucia P.A.
        • Di Lillo F.
        • Masi M.
        • Di Franco F.
        • Mettivier G.
        • Russo P.
        Rotational radiotherapy of breast cancer with polyenergetic kilovoltage X-ray beams: An experimental and Monte Carlo phantom study.
        Phys Med. 2019; 62: 63-72https://doi.org/10.1016/j.ejmp.2019.05.002
        • Di Lillo Francesca
        • Mettivier Giovanni
        • Sarno Antonio
        • Castriconi Roberta
        • Russo Paolo
        Towards breast cancer rotational radiotherapy with synchrotron radiation.
        Phys Med. 2017; 41: 20-25https://doi.org/10.1016/j.ejmp.2017.06.009
      9. Di Lillo F, Mettivier G, Castriconi R, Sarno A, Stevenson AW, Hall CJ, et al. Synchrotron radiation external beam rotational radiotherapy of breast cancer: proof of principle. Journal of synchrotron radiation. 2018;25:857-68.

        • Bazalova-Carter Magdalena
        • Weil Michael D.
        • Breitkreutz Dylan Yamabe
        • Wilfley Brian P.
        • Graves Edward E.
        Feasibility of external beam radiation therapy to deep-seated targets with kilovoltage x-rays.
        Med Phys. 2017; 44: 597-607https://doi.org/10.1002/mp.12047
        • Breitkreutz Dylan Y.
        • Weil Michael D.
        • Zavgorodni Sergei
        • Bazalova-Carter Magdalena
        Monte Carlo simulations of a kilovoltage external beam radiotherapy system on phantoms and breast patients.
        Med Phys. 2017; 44: 6548-6559https://doi.org/10.1002/mp.12619
        • Breitkreutz Dylan Y.
        • Renaud Marc‐André
        • Seuntjens Jan
        • Weil Michael D.
        • Zavgorodni Sergei
        • Bazalova‐Carter Magdalena
        Inverse optimization of low‐cost kilovoltage x‐ray arc therapy plans.
        Med Phys. 2018; 45: 5161-5171https://doi.org/10.1002/mp.13153
      10. Breitkreutz DY, Renaud M-A, Weil MD, Zavgorodni S, Han J, Baxter H, et al. Monte Carlo calculated kilovoltage x-ray arc therapy plans for three lung cancer patients. Biomedical Physics & Engineering Express. 2019;5:065022.

      11. Stalbaum T, Boyd D, Weil M, Chen H, Plies M, Song S, et al. 200 kV x-ray source for radiotherapy and imaging: preliminary results and discussion. Medical Imaging 2020: Physics of Medical Imaging: International Society for Optics and Photonics; 2020. p. 113123I.

      12. Zankowski C, Adelsheim C, Pyyry J, Kuusela E. Radiotherapy treatment planning using artificial intelligence (ai) engines. Google Patents; 2019.

      13. Sarno A, Mettivier G, Russo P, Drebot I, Petrillo V, Bacci A, et al. Kilovoltage rotational radiotherapy of breast cancer with the BriXS source. Journal of Instrumentation. 2020;15:C05012.

      14. Loughery B, Halford R, Dess K, Rakowski J, Cox M, Koh J, et al. A proof of concept kilovoltage intensity modulated radiotherapy platform for the treatment of glioblastoma multiforme. Biomedical Physics & Engineering Express. 2019;5:055013.

        • Bentzen Søren M.
        • Constine Louis S.
        • Deasy Joseph O.
        • Eisbruch Avi
        • Jackson Andrew
        • Marks Lawrence B.
        • Ten Haken Randall K.
        • Yorke Ellen D.
        Quantitative Analyses of Normal Tissue Effects in the Clinic (QUANTEC): An Introduction to the Scientific Issues.
        Int J Rad Oncol Biol Phys. 2010; 76: S3-S9https://doi.org/10.1016/j.ijrobp.2009.09.040
        • Icrp I.
        Publication 61: Annual limits on intake of radionuclides for workers based on the 1990 recommendations.
        Pergamon Press, New York1990
        • Amols H.I.
        • Lagueux B.
        • Cagna D.
        Radiobiological Effectiveness (RBE) of Megavoltage X-Ray and Electron Beams in Radiotherapy.
        Radiat Res. 1986; 105: 58https://doi.org/10.2307/3576725
        • Noone A.
        • Howlader N.
        • Krapcho M.
        • Miller D.
        • Brest A.
        • Yu M.
        • et al.
        SEER cancer statistics review, 1975–2015.
        National Cancer Institute, Bethesda, MD2018: 4
        • Sethi Anil
        • Emami Bahman
        • Small Jr., William
        • Thomas Tarita
        Intraoperative Radiotherapy With INTRABEAM: Technical and Dosimetric Considerations.
        Front Oncol. 2018; 8https://doi.org/10.3389/fonc.2018.00074
        • Esplen Nolan
        • Mendonca Marc
        • Bazalova-Carter Magdalena
        Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review.
        Phys Med Biol. 2020; (In press)