Mobile shielding evaluation on the fetal dose during a breast radiotherapy using Monte Carlo simulation

Published:April 04, 2021DOI:https://doi.org/10.1016/j.ejmp.2021.03.023

      Highlights

      • Breast radiotherapy in a pregnant patient was simulated using MCNPX.
      • A third trimester pregnancy was simulated using a voxelized anthropomorphic phantom.
      • Two different types of mobile shieldings, based on AAPM models, were used.
      • The dose on the fetus depends on the shields, type of projection, and collimation.

      Abstract

      Purpose

      Evaluation of the out-of-field dose is an important aspect in radiotherapy. Due to the fetus radiosensitivity, this evaluation becomes even more conclusive when the patient is pregnant. In this work, a linear accelerator Varian Clinac 2100c operating at 6 MV, a pregnant anthropomorphic phantom (Maria), and different shields added above the abdominal region of the phantom were used for the analysis based on MCNPX. Methods: The simulations were performed for the medial and lateral projections, using either an open field collimation (10×16 cm2) or a multileaf collimator. The added shields (M1 and M2) were designed based on models proposed by Stovall et al. [

      Stovall M, Blackwell CR., Cundiff J., Novack DH, Palta JR., Wagner L K, Webster EW, Shalek RJ. Fetal dose from radiotherapy with photon beams: report of aapm radiation therapy committee task group no. 36. Med Phys 2016;39:7662-76. http://dx.doi.org/10.1118/1.597525.

      ], intending to reduce the deposited dose on the fetus and related structures. Results: The presence of the shields showed to be effective in reducing the doses on the fetus, amniotic sac, and placenta, for example. A reduction of about 43% was found in the dose on the fetus when M2 was added, using the open field collimation, in comparison with the situation with no shield, being the lateral projection the main responsible for the dose. The use of MLC significatively reduced the doses in different structures, including on the fetus and amniotic sac, for example, in comparison to the open field situation. A slight increment on the dose in organs such as the eyes, thyroid and brain was found in both collimation systems, due to the presence of the shields. The contribution of the leakage radiation from the tube head of the linear accelerator was found to be in the order of µGy, being reduced by the presence of the M2 shield. Conclusion: Using the shields showed to be an essential feature in order to reduce the dose not only on the fetus, but also in important structures responsible to its development.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Physica Medica: European Journal of Medical Physics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

      1. Stovall M, Blackwell CR., Cundiff J., Novack DH, Palta JR., Wagner L K, Webster EW, Shalek RJ. Fetal dose from radiotherapy with photon beams: report of aapm radiation therapy committee task group no. 36. Med Phys 2016;39:7662-76. http://dx.doi.org/10.1118/1.597525.

      2. WORLD HEALTH ORGANIZATION. Cancer. 2018. Avaliable in: https://www.who.int/health-topics/cancer#tab=tab_1. Access em: 11 jun. 2020.

      3. INSTITUTO NACIONAL DE CÂNCER (National Cancer Institute). Estatísticas de câncer. 2020. Avaliable in:https://www.inca.gov.br/numeros-de-cancer. Access in: 14 set. 2020 [in portuguese].

      4. AMERICAN CANCER SOCIETY. How Common Is Breast Cancer? 2020. Avaliable in: https://www.cancer.org/cancer/breast-cancer/about/how-common-is-breast-cancer.html. Access in: 11 set. 2020.

        • Esposito S.
        • Tenconi R.
        • Preti V.
        • Groppali E.
        • Principi N.
        Chemotherapy against cancer during pregnancy: A systematic review on neonatal outcomes.
        Med (United States). 2016; 95: 1-6https://doi.org/10.1097/MD.0000000000004899
      5. ICRP Publication 90. Biological Effects after Prenatal Irradiation (Embryo and Fetus). Annals of ICRP. 2003;33;1-2.

      6. ICRP publication 103: The 2007 Recommendations of the International Commission on Radiological Protection. Annals of the ICRP. 2007;37:2-4.

        • Otake M.
        • Schull W.J.
        • Lee S.
        Threshold for radiation-related severe mental retardation in prenatally exposed A-bomb survivors: a re-analysis.
        Int J Radiat Biol. 1996; 70: 755-763https://doi.org/10.1080/095530096144644
        • Kal Henk B
        • Struikmans Henk
        Radiotherapy during pregnancy: fact and fiction.
        Lancet Oncol. 2005; 6: 328-333https://doi.org/10.1016/S1470-2045(05)70169-8
        • Antypas Christos
        • Sandilos Panagiotis
        • Kouvaris John
        • Balafouta Ersi
        • Karinou Eleftheria
        • Kollaros Nikos
        • et al.
        Fetal dose evaluation during breast cancer radiotherapy.
        Int J Radiat Oncol Biol Phys. 1998; 40: 995-999https://doi.org/10.1016/S0360-3016(97)00909-7
        • Owrangi Amir M.
        • Roberts Donald A.
        • Covington Elizabeth L.
        • Hayman James A.
        • Masi Kathryn M.
        • Lee Choonik
        • et al.
        Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield.
        J Appl Clin Med Phys. 2016; 17: 34-46https://doi.org/10.1120/jacmp.v17i5.6135
        • Woo Shiao Y.
        • Fuller Lillian M.
        • Cundiff Jackson H.
        • Bondy Melissa L.
        • Hagemeister Fredrick B.
        • McLaughlin Peter
        • et al.
        Radiotherapy during pregnancy for clinical stages IA-IIA Hodgkin’s disease.
        Int J Radiat Oncol Biol Phys. 1992; 23: 407-412https://doi.org/10.1016/0360-3016(92)90761-6
        • Mazonakis Michalis
        • Tzedakis Antonis
        • Varveris Charalambos
        • Damilakis John
        Radiotherapy for supradiaphragmatic Hodgkin’s disease: Determination of the proper fetal shielding conditions using Monte Carlo methodology.
        Phys Med. 2011; 27: 181-187https://doi.org/10.1016/j.ejmp.2010.12.004
        • Geng Changran
        • Moteabbed Maryam
        • Seco Joao
        • Gao Yiming
        • George Xu X
        • Ramos-Méndez José
        • et al.
        Dose assessment for the fetus considering scattered and secondary radiation from photon and proton therapy when treating a brain tumor of the mother.
        Phys Med Biol. 2016; 61: 683-695https://doi.org/10.1088/0031-9155/61/2/683
        • Mazonakis M.
        • Tzedakis A.
        • Damilakis J.
        Monte Carlo simulation of radiotherapy for breast cancer in pregnant patients: how to reduce the radiation dose and risks to fetus?.
        Radiat Prot Dosim. 2017; 175: 10-16https://doi.org/10.1093/rpd/ncw260
        • Bednarz Bryan
        • Xu X. George
        A feasibility study to calculate unshielded fetal doses to pregnant patients in 6-MV photon treatments using Monte Carlo methods and anatomically realistic phantoms.
        Med Phys. 2008; 35: 3054-3061
        • Beierholm A.R.
        • Nygaard D.E.
        • Juhl E.L.
        • Hansen R.
        • Hansen J.
        Evaluating out-of-field doses during radiotherapy of paediatric brain tumours using lead shielding and flattening-filter free beams.
        Phys Med. 2019; 60: 1-6https://doi.org/10.1016/j.ejmp.2019.03.008
        • Kry Stephen F.
        • Bednarz Bryan
        • Howell Rebecca M.
        • Dauer Larry
        • Followill David
        • Klein Eric
        • et al.
        AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy.
        Med Phys. 2017; 44: e391-e429https://doi.org/10.1002/mp.2017.44.issue-1010.1002/mp.12462
        • Bednarz Bryan
        • Xu X George
        Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: Development and validation.
        Phys Med Biol. 2009; 54: N43-N57https://doi.org/10.1088/0031-9155/54/4/N01
      7. Pelowitz, D. B. MCNPX User‘s Manual, version 2.7.0. Report LA-CP-11-00438. Los Alamos National Laboratory, 2011. AAPM Task Group 36, 1994. Fetal Dose from Radiotherapy with Photon Beams.

      8. Sheikh-Bagheri D, Rogers DWO. Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code. Med Phys 2002;29:391–402. https://doi.org/10.1118/1.1445413.

        • Valentin J.
        89: Basic anatomical and physiological data for use in radiological protection: reference values.
        Annals of ICRP. 2002; 32: 1-277
      9. Cabral MOM, Vieira JW, Leal Neto V, Andrade Lima FR de. Development of a pregnant woman phantom using polygonal mesh, for dosimetric evaluations. Brazilian J Radiat Sci 2015;3:1–11. https://doi.org/10.15392/bjrs.v3i1a.127.

        • Al-Bayyari Nahla Subhi
        • Abu-Heija Adel Taha
        Fetal weight normograms for singleton pregnancies in a Jordanian population.
        Ann Saudi Med. 2010; 30: 134-140https://doi.org/10.4103/0256-4947.60519
        • Berris T.
        • Mazonakis M.
        • Stratakis J.
        • Tzedakis A.
        • Fasoulaki A.
        • Damilakis J.
        Calculation of organ doses from breast cancer radiotherapy: a monte carlo study.
        J Appl Clin Med Phys. 2013; 14: 133-146https://doi.org/10.1120/jacmp.v14i1.4029
        • Kry Stephen F.
        • Titt Uwe
        • Pönisch Falk
        • Followill David
        • Vassiliev Oleg N.
        • Allen White R.
        • et al.
        A Monte Carlo model for calculating out-of-field dose from a Varian 6 MV beam.
        Med Phys. 2006; 33: 4405-4413
        • Kry Stephen F.
        • Starkschall George
        • Antolak John A.
        • Salehpour Mohammad
        Evaluation of the accuracy of fetal dose estimates using TG-36 data.
        Med Phys. 2007; 34: 1193-1197https://doi.org/10.1118/1.2710332
        • Cravo Sá Ana
        • Barateiro Andreia
        • Bednarz Bryan
        • Borges Cecília
        • Pereira Joana
        • Baptista Mariana
        • et al.
        Assessment of out-of-field doses in radiotherapy treatments of paediatric patients using Monte Carlo methods and measurements.
        Phys Med. 2020; 71: 53-61https://doi.org/10.1016/j.ejmp.2020.02.008
        • Han B.
        • Bednarz B.
        • Xu X.G.
        A study of the shielding used to reduce leakage and scattered radiation to the fetus in a pregnant patient treated with a 6-MV external x-ray beam.
        Health Phys. 2009; 97: 581-589https://doi.org/10.1097/01.HP.0000363838.70546.4c
      10. ICRP PUBLICATION 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs : threshold doses for tissue reactions in a radiation protection context. Annals of the ICRP.

        • Li J.S.
        • Freedman G.M.
        • Price R.
        • Wang L.
        • Anderson P.
        • Chen L.
        • et al.
        Clinical implementation of intensity-modulated tangential beam irradiation for breast cancer.
        Med Phys. 2004; 31: 1023-1031https://doi.org/10.1118/1.1690195
      11. Majer M, Stolarczyk L, De Saint-Hubert M, Kabat D, Knežević Ž, Miljanić S, et al. Out-of-field dose measurements for 3D conformal and intensity modulated radiotherapy of a paediatric brain tumour. Radiat Prot Dosimetry 2017;176:331–40. https://doi.org/10.1093/rpd/ncx015.

        • Josipović Mirjana
        • Nyström Håkan
        • Kjær-Kristoffersen Flemming
        IMRT in a Pregnant Patient: How to Reduce the Fetal Dose?.
        Med Dosim. 2009; 34: 301-310https://doi.org/10.1016/j.meddos.2008.11.003