Advertisement

Simplified method for determining dose to a non-water phantom through the use of ND,w and IAEA TRS 483 for the GammaPod

      Highlights

      • Dosimetry using TRS 483 for GammaPod Breast SRS device.
      • Dosimetry for non-standard phantom, PMMA.
      • Simple method, ready to use for clinic.
      • Improved uncertainty.

      Abstract

      Purpose

      GammaPod, a breast stereotactic radiosurgery device, utilizes 25 rotating Co-60 sources to deliver highly conformal dose distributions. The GammaPod system requires that reference dosimetry be performed in a specific vendor-supplied poly-methylmethacrylate (PMMA) phantom. The nonstandard nature of GammaPod dosimetry, in both the phantom material and machine-specific reference (msr), prohibits use of the American Association of Physicists in Medicine Task Group 51 (TG-51) protocol. This study proposes a practical method using TRS 483 to make the reference dosimetry procedure simpler and to reduce overall uncertainties.

      Methods

      The dose to PMMA (DPMMA) is determined under msr conditions using TRS 483 with an Exradin A1SL chamber placed in a PMMA phantom. The conversion factor, which converts from the dose-to-water (Dw) in broad-beam Co-60 reference geometry to DPMMA in the msr small field Co-60 (Qmsr) geometry, is derived using the Monte Carlo simulations and procedure described in TRS 483.

      Results

      The new conversion factor value for an Exradin A1SL chamber is 0.974. When combined with ND,w, DPMMA differs by 0.5% from the TG-21/Nx method and 0.2% from the IROC values. Uncertainty decreased from 2.2% to 1.6%.

      Conclusion

      We successfully implemented TRS 483 reference dosimetry protocols utilizing ND,w for the GammaPod in the PMMA phantom. These results show not only agreement between measurements performed with the previously published method and independent thermoluminescent dosimetry measurements but also reductions in uncertainty. This also provides readers with a pathway to develop their own IAEA TRS 483 factor for any new small field machine that may be developed.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Physica Medica: European Journal of Medical Physics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Becker S.
        • Sabouri P.
        • Niu Y.
        • Prado K.
        • Chen S.
        • Nichols E.
        • et al.
        Commissioning and acceptance guide for the GammaPod.
        Phys Med Biol. 2019; 64: 205021
        • Mutaf Y.D.
        • Zhang J.
        • Yu C.X.
        • Yong Yi B.
        • Prado K.
        • D'Souza W.D.
        • et al.
        Dosimetric and geometric evaluation of a novel stereotactic radiotherapy device for breast cancer: the GammaPod.
        Med Phys. 2013; 40: 041722
        • Yu C.X.
        • Shao X.
        • Zhang J.
        • Regine W.
        • Zheng M.
        • Yu Y.S.
        • et al.
        GammaPod-a new device dedicated for stereotactic radiotherapy of breast cancer [published online ahead of print 2013/05/03].
        Med Phys. 2013; 40: 051703
        • Becker S.J.
        • Niu Y.
        • Mutaf Y.
        • Chen S.
        • Poirier Y.
        • Nichols E.M.
        • et al.
        Development and validation of a comprehensive patient-specific quality assurance program for a novel stereotactic radiation delivery system for breast lesions.
        J Appl Clin Med Phys. 2019; 20: 138-148
        • Huq M.S.
        • Hwang M.S.
        • Teo T.P.
        • Jang S.Y.
        • Heron D.E.
        • Lalonde R.J.
        A dosimetric evaluation of the IAEA-AAPM TRS483 code of practice for dosimetry of small static fields used in conventional linac beams and comparison with IAEA TRS-398, AAPM TG51, and TG51 Addendum protocols.
        Med Phys. 2018; (Online ahead of print)
        • Meltsner S.G.
        • DeWerd L.A.
        Air kerma based dosimetry calibration for the Leksell Gamma Knife.
        Med Phys. 2009; 36: 339-350
        • Moutsatsos A.
        • Petrokokkinos L.
        • Karaiskos P.
        • Papagiannis P.
        • Georgiou E.
        • Dardoufas K.
        • et al.
        Gamma knife output factor measurements using VIP polymer gel dosimetry.
        Med Phys. 2009; 36: 4277-4287
      1. American Association of Physicists in Medicine, Task Group 21, Radiation Therapy Committee. A protocol for the determination of absorbed dose from high-energy photon and electron beams. Med Phys. 1983;10(6):741-771.

        • Palmans H.
        • Andreo P.
        • Huq M.S.
        • Seuntjens J.
        • Christaki K.E.
        • Meghzifene A.
        Dosimetry of small static fields used in external photon beam radiotherapy: summary of TRS-483, the IAEA-AAPM international Code of Practice for reference and relative dose determination.
        Med Phys. 2018; 45: e1123-e1145
        • Becker S.J.
        • Culberson W.S.
        • Poirier Y.
        • Mutaf Y.
        • Niu Y.
        • Nichols E.M.
        • et al.
        Dosimetry evaluation of the GammaPod stereotactic radiosurgery device based on established AAPM and IAEA protocols.
        Med Phys. 2020; 47: 3614-3620
        • Almond P.R.
        • Biggs P.J.
        • Coursey B.M.
        • Hanson W.F.
        • Huq M.S.
        • Nath R.
        • et al.
        AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams.
        Med Phys. 1999; 26: 1847-1870