The impact of vertical off-centering on image noise and breast dose in chest CT with organ-based tube current modulation: A phantom study

Open AccessPublished:July 16, 2022DOI:https://doi.org/10.1016/j.ejmp.2022.06.014

      Highlights

      • Vertical off-centering impacts chest CT scans using OBTCM.
      • Siemens OBTCM approach increased lateral breast dose at lower table heights.
      • GE OBTCM method never increased organ doses in chest CT scans.
      • Image noise followed roughly the expected inverse power law of the change in dose.
      • Monte Carlo simulations followed MOSFET measurements in a good agreement.

      Abstract

      Purpose

      To determine the effects of patient vertical off-centering when using organ-based tube current modulation (OBTCM) in chest computed tomography (CT) with focus on breast dose.

      Materials and methods

      An anthropomorphic adult female phantom with two different breast attachment sizes was scanned on GE Revolution EVO and Siemens Definition Edge CT systems using clinical chest CT protocols and anterior-to-posterior scouts. Scans with and without OBTCM were performed at different table heights (GE: centered, ±6 cm, and ± 3 cm; Siemens: centered, −6 cm, and ± 3 cm). The dose effects were studied with metal-oxidesemiconductor field-effect transistor dosimeters with complementary Monte Carlo simulations to determine full dose maps. Changes in image noise were studied using standard deviations of subtraction images from repeated acquisitions without dosimeters.

      Results

      Patient off-centering affected both the behavior of the normal tube current modulation as well as the extent of the OBTCM. Generally, both OBTCM techniques provided a substantial decrease in the breast doses (up to 30% local decrease). Lateral breast regions may, however, in some cases receive higher doses when OBTCM is enabled. This effect becomes more prominent when the patient is centered too low in the CT gantry. Changes in noise roughly followed the expected inverse of the change in dose.

      Conclusions

      Patient off-centering was shown to affect the outcome of OBTCM in chest CT examination, and on some occasions, resulting in higher exposure. The use of modern dose optimization tools such as OBTCM emphasizes the importance of proper centering when preparing patients to CT scans.

      Keywords

      1. Introduction

      Chest computed tomography (CT) examinations are among the most common CT studies worldwide. Breast tissue is not the target organ in chest CT examination. However, it is one of the most radiosensitive organs in human body and particularly vulnerable to radiation-induced cancer [

      Valentin J (ed.). The 2007 recommendation of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007;37(2-4):1–332.

      ]. Therefore, it is important to make efforts to decrease exposure to glandular breast tissue. Several CT optimization methods, both technical and user-dependent approaches, are used to minimize the total radiation dose to the patients and to protect radiosensitive tissues. Technical procedures include tube current modulation (TCM), organ-based tube current modulation (OBTCM), automated tube voltage selection, adaptive beam collimation, beam-shaping filters, and iterative reconstruction techniques [
      • Kalender W.A.
      • Wolf H.
      • Suess C.
      Dose reduction in CT by anatomically adapted tube current modulation.
      ,
      • Kalra M.K.
      • Maher M.M.
      • Toth T.L.
      • Schmidt B.
      • Westerman B.L.
      • Morgan H.T.
      • et al.
      Techniques and applications of automatic tube current modulation for CT.
      ,
      • Thibault J.B.
      • Sauer K.D.
      • Bouman C.A.
      • Hsieh J.
      A three-dimensional statistical approach to improve image quality for multislice helical CT.
      ,
      • Kalender W.A.
      • Buchenau S.
      • Deak P.
      • Kellermeier M.
      • Langner O.
      • van Straten M.
      • et al.
      Technical approaches to the optimisation of CT.
      ,
      • Vollmar S.V.
      • Kalender W.A.
      Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.
      ,
      • Deak P.D.
      • Langner O.
      • Lell M.
      • Kalender W.A.
      Effects of adaptive section collimation on patient radiation dose in multisection spiral CT.
      ,
      • Duan X.
      • Wang J.
      • Christner J.
      • Leng S.
      • Grant K.L.
      • McCollough G.H.
      Dose reduction to anterior surfaces with organ-based tube current modulation: evaluation of performance in a phantom study.
      ,
      • Katsura M.
      • Matsuda I.
      • Akahane M.
      • Sato J.
      • Akai H.
      • Yasaka K.
      • et al.
      Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique.
      ,
      • Miéville F.A.
      • Berteloot L.
      • Grandjean A.
      • Ayestaran P.
      • Gudinchet F.
      • Schmidt S.
      • et al.
      Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis.
      ,
      • Schindera S.T.
      • Winklehner A.
      • Goetti R.
      • Fischer M.
      • Gnannt R.
      • Szucs-Farkas Z.
      Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: a phantom study.
      ,
      • Taylor S.
      • Litmanovich D.E.
      • Shahrzad M.
      • Bankier A.A.
      • Genevois P.A.
      • Tack D.
      Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone?.
      ,
      • Samei E.
      • Richard S.
      Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology.
      ,
      • Spearman J.V.
      • Schoepf U.J.
      • Rottenkolber M.
      • Driesser I.
      • Canstein C.
      • Thierfelder K.M.
      • et al.
      Effect of automated attenuation-based tube voltage selection on radiation dose at CT: An observational study on a global scale.
      ,
      • Fu W.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Kapadia A.J.
      • Samei E.
      Breast dose reduction with organ-based, wide-angle tube current modulated CT.
      ]. External bismuth shields have been used to reduce exposure to radiosensitive surface tissues, such as breasts and thyroid, but their use in clinical routine has been discouraged [
      • McLaughlin D.J.
      • Mooney R.B.
      Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the userśs needs?.
      ,
      • Wang J.
      • Duan X.
      • Christner J.A.
      • Leng S.
      • Yu L.
      • McCollough C.H.
      Radiation dose reduction to the breast in thoracic CT: Comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current.
      ,
      • Wang J.
      • Duan X.
      • Christner J.
      • Leng S.
      • Grant K.L.
      • McCollough G.H.
      Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT.
      ,

      AAPM Position statement on the use of bismuth shielding for the purpose of dose reduction in CT scanning. AAPM 2017. PP 26-B (https://www.aapm.org/publicgeneral/BismuthShielding.pdf) [accessed 5 February 2022].

      ]. Despite the technical solutions for CT optimization, the role of radiographers remains important for achieving optimal results. Several studies have shown harmful effects of patient vertical off-centering on patient dose and image quality due to misalignment of beam-shaping filters and impacts of geometric magnification or minification of patient structures in planning radiographs [
      • Li J.
      • Udayasankar U.K.
      • Toth T.L.
      • Seamans J.
      • Small W.C.
      • Kalra M.K.
      Automatic patient centering for MDCT: Effect on radiation dose.
      ,
      • Toth T.
      • Ge Z.
      • Daly M.P.
      The influence of patient centering on CT dose and image noise.
      ,
      • Habibzadeh M.A.
      • Ay M.R.
      • Asl A.R.
      • Gharidi H.
      • Zaidi H.
      Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Lampinen A.
      • Kortesniemi M.
      Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Reijonen V.
      • Kortesniemi M.
      Effect of patient centering on patient dose and image noise in chest CT.
      ,
      • Kataria B.
      • Sandborg M.
      • Althén J.N.
      Implications of patient centring on organ dose in computed tomography.
      ,
      • Filev P.D.
      • Mittal P.K.
      • Tang X.
      • Duong P.A.
      • Wang X.
      • Small W.C.
      • et al.
      Increased computed tomography dose due to miscentering with use of automated tube voltage selection: phantom and patient study.
      ,
      • Saltybaeva N.
      • Alkadhi H.
      Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.
      ,
      • Marsh R.M.
      • Silosky M.S.
      The effects of patient positioning when interpreting CT dose metrics: A phantom study.
      ,
      • Akin-Akintayo O.O.
      • Alexander L.F.
      • Neill R.
      • Krupinksi E.A.
      • Tang X.
      • Mittal P.K.
      • et al.
      Prevalence and severity of off-centering during diagnostic CT: Observations from 57,621 CT scans of the chest, abdomen, and/or pelvis.
      ,
      • Barreto I.
      • Lamoreux R.
      • Olguin C.
      • Quails N.
      • Correa N.
      • Rill L.
      • et al.
      Impact of patient centering in CT on organ dose and the effect of using a positioning compensation system: Evidence from OSLD measurements in postmortem subjects.
      ,
      • Euler A.
      • Saltybaeva N.
      • Alkadhi H.
      How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.
      ,
      • Gang Y.
      • Chen X.
      • Li H.
      • Wang H.
      • Guo Y.
      • Zeng J.
      • et al.
      A comparison between manual and artificial intelligence-based automatic positioning in CT imaging for COVID-19 patients.
      ,
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ]. CT manufacturers have developed technical tools to prevent patient off-centering and to reduce its effects on patient dose and image quality [
      • Barreto I.
      • Lamoreux R.
      • Olguin C.
      • Quails N.
      • Correa N.
      • Rill L.
      • et al.
      Impact of patient centering in CT on organ dose and the effect of using a positioning compensation system: Evidence from OSLD measurements in postmortem subjects.
      ,
      • Gang Y.
      • Chen X.
      • Li H.
      • Wang H.
      • Guo Y.
      • Zeng J.
      • et al.
      A comparison between manual and artificial intelligence-based automatic positioning in CT imaging for COVID-19 patients.
      ,
      • Booij R.
      • Budde R.P.J.
      • Dijkshoorn M.L.
      • van Straten M.
      Accuracy of automated patient positioning in CT using a 3D camera for body contour detection.
      ,
      • Saltybaeva N.
      • Schmidt B.
      • Wimmer A.
      • Flohr T.
      • Alkadhi H.
      Precise and automatic patient positioning in computed tomography: Avatar modeling of the patient surface using a 3-dimensional camera.
      ].
      Recently developed OBTCM methods aim to reduce exposure to radiosensitive surface tissues by reducing photon flux on the anterior projections during the image acquisition. Technical implementations of OBTCM methods vary between the vendors, for example, in terms of angular organ-dose modulation ranges and the reduction rates of the tube current. Moreover, Siemens approach compensates anteriorly reduced tube current by increasing the current during the posterior projections, whereas GE and Canon approaches do not [
      • Vollmar S.V.
      • Kalender W.A.
      Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.
      ,
      • Fu W.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Kapadia A.J.
      • Samei E.
      Breast dose reduction with organ-based, wide-angle tube current modulated CT.
      ,
      • Dixon M.T.
      • Loader R.J.
      • Stevens G.C.
      • Rowles N.P.
      An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.
      ,
      • Fu W.
      • Tian X.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Goodsitt M.M.
      • et al.
      CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.
      ,
      • Franck C.
      • Smeets P.
      • Lapeire L.
      • Achten E.
      • Bacher K.
      Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
      ]. In chest CT, the overall dose-reduction percentages yielded by the OBTCM methods for the breasts have been reported to be between 5% and 48% compared to chest CT scan protocols utilizing normal TCM [
      • Vollmar S.V.
      • Kalender W.A.
      Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.
      ,
      • Fu W.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Kapadia A.J.
      • Samei E.
      Breast dose reduction with organ-based, wide-angle tube current modulated CT.
      ,
      • Dixon M.T.
      • Loader R.J.
      • Stevens G.C.
      • Rowles N.P.
      An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.
      ,
      • Fu W.
      • Tian X.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Goodsitt M.M.
      • et al.
      CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.
      ,
      • Franck C.
      • Smeets P.
      • Lapeire L.
      • Achten E.
      • Bacher K.
      Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
      ,
      • Lungren M.P.
      • Yoshizumi T.T.
      • Brady S.M.
      • Toncheva G.
      • Anderson-Evans C.
      • Lowry C.
      • et al.
      Radiation dose estimations to the thorax using organ-based dose modulation.
      ,
      • Euler A.
      • Szucs-Farkas Z.
      • Falkowski A.L.
      • Kawel-Böhm N.
      • D’Errico L.
      • Kopp S.
      • et al.
      Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
      ,
      • Lopez-Rendon X.
      • Zhang G.
      • Coudyzer W.
      • Develter W.
      • Bosmans H.
      • Zanca F.
      Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT?.
      ,
      • Brand M.
      • Wuest W.
      • May M.
      • Uder M.
      • Sommer M.
      Influence of risk-organ-based tube current modulation on CT-induced DNA double-strand breaks in a biological phantom model.
      ,
      • Mussmann B.R.
      • Mørup S.D.
      • Skov P.M.
      • Brenøe A.S.
      • Eldahl F.
      • Jørgensen G.M.
      • et al.
      Organ-based tube current modulation in chest CT.
      ]. However, it has also been warned that the dose may increase in some parts of the breast tissue outside the reduced dose zone [
      • Taylor S.
      • Litmanovich D.E.
      • Shahrzad M.
      • Bankier A.A.
      • Genevois P.A.
      • Tack D.
      Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone?.
      ,
      • Fu W.
      • Tian X.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Goodsitt M.M.
      • et al.
      CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.
      ,
      • Franck C.
      • Smeets P.
      • Lapeire L.
      • Achten E.
      • Bacher K.
      Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
      ,
      • Euler A.
      • Szucs-Farkas Z.
      • Falkowski A.L.
      • Kawel-Böhm N.
      • D’Errico L.
      • Kopp S.
      • et al.
      Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
      ,
      • Lopez-Rendon X.
      • Zhang G.
      • Coudyzer W.
      • Develter W.
      • Bosmans H.
      • Zanca F.
      Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT?.
      ].
      Previous studies have shown patient off-centering to be a common problem in CT, patients being typically centered too low in the gantry [
      • Kaasalainen T.
      • Palmu K.
      • Reijonen V.
      • Kortesniemi M.
      Effect of patient centering on patient dose and image noise in chest CT.
      ,
      • Akin-Akintayo O.O.
      • Alexander L.F.
      • Neill R.
      • Krupinksi E.A.
      • Tang X.
      • Mittal P.K.
      • et al.
      Prevalence and severity of off-centering during diagnostic CT: Observations from 57,621 CT scans of the chest, abdomen, and/or pelvis.
      ,
      • Saltybaeva N.
      • Schmidt B.
      • Wimmer A.
      • Flohr T.
      • Alkadhi H.
      Precise and automatic patient positioning in computed tomography: Avatar modeling of the patient surface using a 3-dimensional camera.
      ]. Studies considering dose reduction techniques have mainly focused on methods other than OBTCM [
      • Kaasalainen T.
      • Palmu K.
      • Lampinen A.
      • Kortesniemi M.
      Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Reijonen V.
      • Kortesniemi M.
      Effect of patient centering on patient dose and image noise in chest CT.
      ,
      • Kataria B.
      • Sandborg M.
      • Althén J.N.
      Implications of patient centring on organ dose in computed tomography.
      ,
      • Filev P.D.
      • Mittal P.K.
      • Tang X.
      • Duong P.A.
      • Wang X.
      • Small W.C.
      • et al.
      Increased computed tomography dose due to miscentering with use of automated tube voltage selection: phantom and patient study.
      ,
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ]. A previous simulation study based on a single vendor has suggested that off-centering may result in increased exposure to the breast tissue in OBTCM-based chest CT scans [
      • Euler A.
      • Szucs-Farkas Z.
      • Falkowski A.L.
      • Kawel-Böhm N.
      • D’Errico L.
      • Kopp S.
      • et al.
      Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
      ]. The aims of our phantom study were to investigate the impact of patient off-centering on breast dose and image noise in chest CT examinations with and without OBTCM and with different breast sizes, and to report how two vendors’ implementations differ in similar geometries. Furthermore, due to the complexity of combined effects of centering and OBTCM with helical scanning geometry, the absorbed doses were determined via both Monte Carlo (MC) simulations as well as measured with metal-oxidesemiconductor field-effect transistor (MOSFET) detectors. The two dose estimation methods were considered complementary: the MOSFET measurements reflected the actual dose accumulation at a limited number of point locations and the MC simulations provided estimates for more comprehensive volumetric dose distributions. Relative changes in the image noise were reported to aid in reflecting the potential dose savings against changes in the final image appearance.

      2. Material and methods

      2.1 Phantom measurements

      An anthropomorphic adult female phantom (ATOM Model 702-D, CIRS, Norfolk, VA, USA) was scanned in a supine position with 64-slice GE Revolution EVO (GE Healthcare, Milwaukee, Wisconsin, USA) and 128-slice Siemens Definition Edge (Siemens Healthcare, Erlangen, Germany) CT systems. The scans were performed at different table heights using clinically utilized non-contrast helical chest CT scan protocols with and without OBTCM. Each scan was performed with a fixed 120 kVp tube voltage after an anterior-to-posterior (AP) scout image. AP scout was chosen based on Siemens’s recommendation for body region, and the same direction was used with GE for comparability. The large body scan field of view, 64×0.625 mm total collimation, noise index of 36 (at 0.625 mm slice thickness), min–max mA range of 10–540 mA, 0.5 s rotation time, 0.984 pitch, and dose reduction of 20% were used in the GE scanner. Similarly, a 128×0.6 mm total collimation, quality reference mAs of 120, rotation time of 0.33 s, and 0.6 pitch were used in the Siemens CT system. The anatomical characteristics of the adult female phantom were weight of 55 kg, height of 160 cm, and chest dimensions of 20×25 cm without breasts. The chest CT scans were performed with two sizes of breast attachments: smaller 190 cc breasts (702-D-190, CIRS, Norfolk, USA) approximating 50% glandular and 50% adipose equivalent mixed material (-50 HU at 120 kVp), and larger 500 cc custom-made breasts. The custom-made breasts had a 3D-printed polypropylene shell (-150 HU at 120 kVp, 50% of the total volume), which was filled with water (50% of the total volume). The low-density polypropylene shell (thickness approximately 5 mm) acted as a surrogate for adipose tissue and the water mimicked glandular tissue. The phantom did not have any internal structures. The custom design was based on the models 701-BR-02R and 701-BR-02L CIRS attachments that were not suitable for smaller body phantom model 702-D used in this study. The phantom was set to the CT scanner isocenter by lasers and visual landmarks on the phantom. The scans were performed at five different vertical levels of the patient table for the GE system: vertical positions varied from 6 cm below to 6 cm above the reference level in 3 cm steps. The Siemens system had a limited table height adjustment range, and therefore only four different vertical levels of the patient table (from 6 cm below to 3 cm above the reference level in 3 cm steps) were used. The scan range was set to that of a typical chest CT examination (from the apex of the lungs to the lateral phrenic angles). Equal OBTCM ranges (13 cm) were used for both breast sizes with the GE system. The scanner reported volume CT dose index (CTDIvol) values were recorded.
      Before dose measurements, 12 high-sensitivity TN-1002RD MOSFET dosimeters (Best Medical Canada, Ottawa, ON, Canada) with high-bias settings were calibrated in air for 120-kVp tube voltage using clinical CT beam settings. The calibrations were performed separately for both CT systems in axial scanning mode using 100 mAs exposure. The reference air kerma values in the scan isocenter were measured with a 10-cm pencil ionization chamber (RaySafe Xi, Unfors RaySafe AB, Billdal, Sweden). Thereafter, calibration factors for each MOSFET dosimeter were defined separately by repeating measurements five times. The relative standard deviations of the repeated measurements fell in the range from 2% to 6%. A low-dose exposure limit of 1.69 mGy for 25% measurement uncertainty at the 95% confidence level has been reported for this type of MOSFET dosimeter [
      • Koivisto J.
      • Wolf J.
      • Kiljunen T.
      • Schulze D.
      • Kortesniemi M.
      Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms.
      ]. After the calibration, the dosimeters were placed into and onto the adult female phantom and doses to the lungs (one dosimeter in the right central lung), mediastinum (one dosimeter in heart), and right and left breasts (five dosimeters on the large breasts and four dosimeters on the small breasts on both sides) were measured. Thus, altogether 12 dosimeters were used for the large breast setup and 10 dosimeters for the small breast setup. The dosimeter positions in the lungs, heart, and under the breasts (two dosimeters under each breast) were the same for the small and large breasts, whereas the dosimeter positions above the breast attachments (three dosimeters on large breasts and two dosimeters on small breasts on each side) varied between the breasts (Fig. 1). The scans were repeated three times before reading the dosimeters. This was done to improve the reproducibility of the dose measurements and to achieve sufficiently high radiation doses in the dosimeters. Moreover, this procedure was repeated three times (or total of nine scans) per acquisition setup. The individual dosimeter readings were used to report measurement uncertainty (random error) of the mean absorbed point doses.
      Figure thumbnail gr1
      Fig. 1Phantom setups and MOSFET locations (red) for the large (A) and small breast attachments (B) shown as photographic images (top row), frontal 3D projections (middle row) and axial images (bottom row). R1-3 were placed above the large attachment, approximately 26, 14 and 10 cm from the midline, respectively. Similarly, r1-2 were placed laterally and medially on the small attachment. MOSFETs R4-5 and r3-4 were placed laterally and medially between the phantom torso and the breast. MOSFETs placed symmetrically on the right breast are not labeled in the image. Right lung (RL) and mediastinum (M) locations are shown in blue. In the lower images, the MOSFET locations are projected onto the visualized central axial slices. The regions-of-interest shown in different colors were used in reporting the values from Monte Carlo simulations and noise measurements. White cross: the center-of-rotation for the reference (0 cm) vertical centering. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

      2.2 Dose simulations

      Absorbed dose maps were estimated using MC simulations (ImpactMC, Vamp GmbH, Germany). The simulations were performed using 3D voxelized data of the complete anthropomorphic phantom with the same imaging configurations as the MOSFET measurements and using 5×109 photons. For the final maps, simulations from two X-ray tube starting angles (0° and 180°) were averaged, and the raw dose maps were scaled based on the pencil chamber air kerma measurements performed for the MOSFET calibrations. The anthropomorphic phantom datasets used in the simulations were resampled to 1.5 mm in all three directions to lower the memory requirements. As a prerequisite for the MC simulations, the X-ray spectrum was simulated for 120-kVp tube voltage using Spekcalc 2.0 program [
      • Poludniowski G.
      • Landry G.
      • DeBlois F.
      • Evans P.M.
      • Verhaegen F.
      SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes.
      ], and information from the vendor specifications such as tube type, half-value layer, anode material, and filtration were used. Furthermore, beam dose profiles free-in-air were measured for both scanners with the same pencil ionization chamber as used in the MOSFET calibrations to provide source data for beam-shaping filter thicknesses across the axial scan plane. The method used to define the shapes of the beam-shaping filters corresponded to the previously published static X-ray tube method [
      • Alikhani B.
      • Büermann L.
      Non-invasive experimental determination of a CT source model.
      ]. Bowtie thickness data and simulated spectrum were then used as input for the MC simulations together with the anthropomorphic phantom CT voxel data, scan range, pitch, and angle-specific tube current values. The scan range was matched with the phantom chest acquisitions. Adaptive helical collimation was not incorporated in the simulations. For the GE system, tube current values of the scans performed either with or without OBTCM (ODM, GE Healthcare, Milwaukee, Wisconsin, USA) were collected from the mA tables shown by the scanner (four mA values reported per rotation, i.e., anterior, left, posterior, and right). A sinusoidal function was fitted to the tabulated mA values to determine the intermediate angle current values used in the MC simulations. In GE Revolution EVO, the dose reduction angle in OBTCM was 180°, in which the anterior-side projections were scanned with 40% reduced tube current values, while 180° of the posterior-side projections were scanned with the same tube current values as would have been the case also without OBTCM. For the Siemens scanner, mean tube current value in each slice was determined from the DICOM header. Thereafter, the mA values used in the MC simulations were calculated using the information from the CT vendor: in Siemens OBTCM method (X-CARE, Siemens Healthcare, Erlangen, Germany), the real-time mA modulation in xy-direction is turned off, 80° of the anterior projections are scanned with 25% of the mean tube current, 40° of the anterolateral projections (20° on either side of the patient) with 25% to 125% of the mean photon flux (linear ramp, thus the tube current is limited between 25% and 100% of the mean value within the range of 55° on either side of the patient), and 240° of the posterior projections with 125% of the mean tube current values.
      3D absorbed-dose voxel maps were acquired as the output of the simulations. Spherical volumes-of-interest with 15-mm diameters were placed at the corresponding MOSFET dosimeter locations and mean MC doses were recorded. Furthermore, from the simulated 3D dose maps, the average absorbed doses with each scan setup for the breasts, lungs, and mediastinum were determined by subsegments: anterior and lateral for lungs and mediastinum, lateral and medial for large breasts (see for the ROI placement Fig. 1). The breast ROIs covered the whole breasts (with a small margin at the edges). The lung and mediastinum ROIs were limited to the same extent as the large breast attachments. Bones, support structures, and the slices where air between phantom slabs is visible were excluded from the lung and mediastinum ROIs. The values from the corresponding left and right regions (for breasts and lungs) were averaged. Relative changes in the region-of-interest doses were calculated and presented.

      2.3 Image noise

      The chest CT scans were also performed without MOSFET dosimeters on both CT systems. The scans used the same scan settings and table heights as in the dose measurements. The scans were performed twice at each table vertical height, and subtraction images were then calculated. The image noise values, i.e., standard deviations, were calculated from the same phantom regions as were used in the MC simulation analysis (Fig. 1).
      The relative changes in image noise were evaluated from 0.625 mm (GE) and 0.6 mm (Siemens) thick axial reconstructions (512×512 pixels) of the phantom. The scans were reconstructed with clinically used reconstruction kernels: the standard reconstruction filter on the GE and the i31f kernel on the Siemens system. Iterative reconstruction was used in all the reconstructions: 40% ASiR-V/filtered back projection blending for the GE and Admire (level 2) for the Siemens. The axial display field of view was adjusted to 35 cm, resulting in a pixel size of 0.683×0.683 mm2 in all the images. The mean HU values per individual ROIs (as shown in Fig. 1) were recorded to verify contrast consistency between different acquisitions and setups.

      3. Results

      3.1 Dosimeter measurements

      Patient vertical positioning affected the function of TCM and anatomical extent of the OBTCM, and thereby the absorbed organ doses, in the chest CT examinations with clear differences between the CT vendors (Fig. 2). In the Siemens scanner, enabling OBTCM increased the reported CTDIvol values up to 7% compared to normal TCM chest CT scans, whereas in the GE system, OBTCM decreased the CTDIvol up to 17% compared to non-OBTCM scans. The CTDIvol values were the highest with the uppermost table height positions, as expected due to geometric magnification in the AP planning radiographs. The relative changes in CTDIvol with different vertical levels of patient table compared to 0-cm-level were higher with GE (–22% to +38%, −6 cm to + 6 cm) than Siemens scanner (-16% to +12%, −6 cm to +3 cm).
      Figure thumbnail gr2
      Fig. 2MOSFET point dose measurements at different vertical positions and phantom setups for GE (A and B) and Siemens scanner (C and D) with and without OBTCM. Zero (0 cm) position is the reference vertical centering level. Breast doses shown in figures at different breast regions are averages over the right and left breasts.
      The GE OBTCM reduced organ doses by 10% to 50% compared to non-OBTCM scans, depending on the table height. The highest dose savings were measured on the medial sections of breasts (from 42% to 50%) when the phantoms were positioned above the scan isocenter. The dose-reduction percentages were reduced with lower table height positions. Contrary to GE, Siemens OBTCM increased the right central lung dose from 5% to 25% compared to normal TCM chest CT scans with different vertical positions of patient table. Furthermore, absorbed dose to heart was slightly increased in Siemens OBTCM scans compared to non-OBTCM scans when the phantom was off-centered below the scan isocenter. Impacts of Siemens OBTCM on breast dose varied depending on vertical positioning of the phantom and the selected sections on breast tissue. In general, Siemens OBTCM resulted in higher dose-saving percentages for the breast tissue compared to non-OBTCM scans when the phantom was positioned higher in the gantry. Lateral breast sections, especially above the breast attachments, received higher absorbed doses (up to 10%) with OBTCM than without OBTCM. Only the medial breast sections received lower doses in all the vertical table levels with OBTCM. The breast size did not affect the general organ-dose level as much in the Siemens as in the GE CT scanner.

      3.2 Dose simulations

      We compared MC point doses to the corresponding MOSFET dosimeter values (Fig. 3). Due to the approximations used in the protocol modelling (e.g., X-ray spectrum, estimated TCM and OBTCM behavior, bowtie shapes), when compared to the MOSFETs, the MC simulations systematically underestimated the point doses by 20% (the average ratio between the two was 0.80 ± 0.08). However, the relationship was linear (R2 = 0.90), which justifies the following investigations based on the relative changes in dose.
      Figure thumbnail gr3
      Fig. 3MC simulation estimated doses were systematically 20% lower than those measured with MOSFETs. The MC point doses were measured over 15-mm-diameter spherical volumes-of-interest (with air voxels excluded) at the corresponding MOSFET locations.
      According to MC simulations (Table 1), organ doses in chest CT scans with and without OBTCM typically reduced with lower table positions due to geometric minification seen in the AP scout images. This applied to both CT systems. Posterior lung and posterior mediastinum doses varied the most at different vertical table heights. This was mostly because the CT system beam shaping filter assumes cylindrically symmetric target positioned in the isocenter. Vertical off-centering results in angularly varying mismatch between the assumed target attenuation and designed filter thickness. The highest dose-saving percentages with OBTCM methods were observed for the medial breast tissue sections (up to 27% and 30% for GE and Siemens, respectively). The GE OBTCM never increased organ doses compared to non-OBTCM scans, and resulted in organ dose savings of 13% to 27%. However, positioning the phantom below the scan isocenter reduced the dose-saving capability of the method.
      Table 1The relative changes in the average absorbed MC doses measured inside seven regions-of-interest. The reported values are TCM (or OBTCM) at table height × cm compared to TCM (or OBTCM) at 0 cm; and OBTCM wrt TCM at the same table height. The absolute absorbed dose values (mean ± standard deviation over the ROI) and the average tube currents (mA) used in the simulations are reported for the reference height 0 cm (gray background). The mGy values are scaled by factor 1.25 (=1/0.8 in Fig. 3) to better match the more conservative MOSFET measurements. Abbreviations: wrt: with respect to; TCM: tube current modulation (without OBTCM); OBTCM: organ-based tube current modulation enabled (with normal TCM).
      Setup for dose simulationsLateral large breastMedial large/small breastMediastinum anteriorMediastinum posteriorLungs anteriorLungs posteriorAverage tube current
      TCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCM
      wrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCM
      GELarge breast6 cm18%15%−24%4%2%−27%12%11%−24%50%47%−17%17%15%–22%48%45%−18%42%40%–22%
      3 cm9%7%–23%1%0%−26%8%6%–23%25%23%−17%9%8%−21%24%22%−18%18%17%–22%
      0 cm4.6 ± 1.3 mGy3.6 ± 1.0 mGy–22%5.2 ± 1.2 mGy3.9 ± 0.9 mGy−25%5.9 ± 0.6 mGy4.6 ± 0.5 mGy–22%5.8 ± 0.7 mGy4.9 ± 0.6 mGy−16%5.7 ± 1.0 mGy4.6 ± 0.9 mGy−20%6.0 ± 1.0 mGy5.0 ± 0.9 mGy−17%67 mA53 mA−21%
      −3cm−9%−8%−21%−1%0%−24%−10%−9%–22%−21%−21%−15%−11%−10%−20%−21%−21%−16%−13%−13%−21%
      −6cm−18%−17%−20%−4%−3%−24%−21%−20%−21%−39%−38%−14%–23%–23%−19%−39%−38%−15%−24%–23%−20%
      Small breast6 cm−4%−6%−25%5%5%−21%41%39%−16%11%10%−20%41%39%−17%34%33%−19%
      3 cm−4%−4%−25%4%4%−21%20%20%−15%7%6%−19%20%20%−16%15%14%−19%
      0 cm4.1 ± 0.8 mGy3.1 ± 0.6 mGy−24%4.7 ± 0.5 mGy3.7 ± 0.4 mGy−21%4.5 ± 0.5 mGy3.8 ± 0.4 mGy−15%4.7 ± 0.9 mGy3.8 ± 0.7 mGy−19%4.6 ± 0.8 mGy3.9 ± 0.7 mGy−15%48 mA39 mA−19%
      −3cm3%4%–23%−8%−7%−20%−20%−19%−14%−10%−9%−18%−20%−19%−15%−11%−11%−19%
      −6cm1%3%–22%−19%−17%−19%−38%−37%−12%–23%–22%−17%−38%−37%−13%–22%−20%−18%
      SiemensLarge breast3 cm−1%−4%3%−6%−12%−29%−3%−4%−13%13%15%22%0%−1%3%12%14%21%6%9%8%
      0 cm5.1 ± 1.0 mGy5.3 ± 1.2 mGy6%5.6 ± 1.1 mGy4.3 ± 0.8 mGy−24%6.4 ± 0.7 mGy5.7 ± 1.0 mGy−12%6.3 ± 0.8 mGy7.6 ± 1.0 mGy20%6.1 ± 1.1 mGy6.3 ± 1.3 mGy3%6.3 ± 1.2 mGy7.5 ± 1.4 mGy19%93 mA98 mA5%
      −3cm−6%1%13%2%13%−16%−5%0%−7%−18%−16%22%−7%−3%8%−17%−15%22%−8%−7%7%
      −6cm−11%−2%16%3%21%−11%−12%−6%−7%–33%–32%22%−15%−11%8%−31%−30%21%−15%−13%8%
      Small breast3 cm−2%−14%−30%2%−5%−13%19%14%18%6%−2%1%19%13%16%13%8%2%
      0 cm5.4 ± 0.9 mGy4.3 ± 0.8 mGy−20%6.3 ± 0.6 mGy5.9 ± 1.1 mGy−6%6.1 ± 0.7 mGy7.5 ± 0.9 mGy24%6.2 ± 1.1 mGy6.7 ± 1.4 mGy8%6.1 ± 1.1 mGy7.4 ± 1.4 mGy22%87 mA92 mA6%
      −3cm3%15%−11%−5%0%−1%−18%−17%26%−8%−4%13%−18%−16%25%−10%−7%9%
      −6cm6%26%−6%−11%−7%−1%–33%–32%24%−17%−13%13%–32%−31%23%−16%−13%9%
      Fig. 4 shows visually how enabling OBTCM on the GE scanner reduces the breast doses. Likewise, centering patient too low with AP scout direction reduces the overall dose in general and especially on the posterior side of the patient. In relation to normal TCM chest CT scans, the Siemens OBTCM reduced radiation dose to medial breast tissue approximately a quarter in case of large breasts and one fifth with small breasts when the phantom was positioned vertically on the scan isocenter. However, when lowering the patient table by 6 cm from this, the dose-saving percentages for medial breast sections reduced. At the same time, the portion of lateral breast regions receiving higher doses with OBTCM increased (Fig. 5 and Table 1). Fig. 5 also shows that parts of the lateral breast tissues may lay outside the reduced dose zone of Siemens OBTCM, although the patient would be correctly positioned on the scan isocenter. In Siemens, when positioning the phantom with large breast attachments 6 cm below the scan isocenter, a 16% increase of the mean lateral breast dose was observed with OBTCM scans in relation to non-OBTCM scans.
      Figure thumbnail gr4
      Fig. 4MC dose simulations of the GE scanner at the optimal centering (0 cm; A and B) and by lowering the table by 6 cm (C and D) without OBTCM (A and C) and with OBTCM (B and D).
      Figure thumbnail gr5
      Fig. 5MC dose simulations of the Siemens scanner at the optimal centering (0 cm; A and B) and by lowering the table by 6 cm (C and D) without OBTCM (A and C) and with OBTCM (B and D). The competing effects from mispositioning (between A and C: increased dose in the fronto-medial region) and OBTCM (between C and D: decreased dose in the frontal angle) is well visualised.
      Differential dose volume histograms from MC simulations are presented in Fig. 6 with the studied vertical centering levels applied to medial (Fig. 6A) and lateral (Fig. 6B) regions of the large breast model. The dose values in the histogram are normalized to dose in air at isocenter, and thus they exclude the net mAs differences in the scans but represent the combined effect of angular intensity of the rotational exposures and phantom alignment on that exposure geometry.
      Figure thumbnail gr6
      Fig. 6Differential dose-volume histograms at different vertical centering in the A) medial breast and B) lateral breast region of the large breast phantom setup for GE (left hand side) and Siemens (right) scanner. The corresponding left and right breast regions are combined. The values are normalized to air kerma measured at isocenter, and a fixed bin width was used in all the histograms.

      3.3 Image noise

      Changes in the image noise were assessed by measuring HU standard deviations inside the ROIs from the repeated-acquisition subtraction images. The relative changes from TCM-only to TCM-with-OBTCM, and from reference centering to different vertical offsets are shown in Table 2. The noise had generally the expected inverse relation to CTDIvol (approximating the average irradiation over the whole imaging volume). Noise increase after off-centering was emphasized for the regions furthest from the isocenter (e.g. anterior regions at high table height). Vertical offset of ± 3 cm from the reference height resulted in noise changes from –12% to 18%. The largest observed increase in noise (36% to 44%) was for the posterior lung when the table was at –6 cm height for GE scanner. Within ± 3 cm table height range enabling the OBTCM resulted in only minor changes in noise: increase by 5% to 10% for the GE scanner and change from –7% to 1% for the Siemens scanner. These were approximately in line with the CTDIvol changes from –16% to –12% and from 2% to 7% for GE and Siemens, respectively.
      Table 2The relative percentage changes in noise measured inside seven regions-of-interest. The reported values are TCM (or OBTCM) at table height × cm relative to TCM (or OBTCM) at 0 cm; and OBTCM relative to TCM at the same table height. Absolute noise values and CT dose indices are reported for the reference table height 0 cm (gray background). Abbreviations: wrt: with relative to; TCM: tube current modulation (without OBTCM); OBTCM: organ-based tube current modulation enabled (with normal TCM).
      Setup for noise measurementsLateral large breastMedial large/small breastMediastinum anteriorMediastinum posteriorLungs anteriorLungs posteriorCT Dose Index
      TCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCMTCMOBTCMOBTCM
      wrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCMwrt 0 cmwrt 0 cmwrt TCM
      GELarge breast6 cm4%4%6%15%15%7%−1%0%9%−14%−12%7%5%6%10%−16%−16%9%38%36%−17%
      3 cm1%0%5%4%4%7%−3%−2%9%−7%−6%6%0%0%10%−11%−10%10%17%15%−16%
      0 cm16 HU17 HU6%17 HU18 HU7%20 HU21 HU7%20 HU21 HU6%24 HU26 HU9%21 HU23 HU8%2.5 mGy2.1 mGy−15%
      −3cm4%3%5%0%0%6%5%4%7%9%7%5%8%6%7%18%17%7%−12%−12%−15%
      −6cm13%11%4%6%4%5%10%9%6%23%20%3%20%17%7%44%41%6%–22%–22%−15%
      Small breast6 cm11%13%9%−1%1%9%−12%−10%7%7%9%11%−14%−13%10%33%32%−13%
      3 cm4%5%8%−1%−1%7%−5%−5%5%2%1%9%−9%−9%8%14%14%−13%
      0 cm16 HU18 HU7%21 HU22 HU7%21 HU22 HU5%23 HU25 HU10%23 HU25 HU8%2.1 mGy1.8 mGy−12%
      −3cm0%0%7%4%4%7%6%7%7%5%5%9%15%15%7%−12%−11%−12%
      −6cm6%4%5%9%8%6%22%19%3%18%15%7%40%36%5%−21%−20%−12%
      SiemensLarge breast3 cm4%4%−4%8%8%−4%4%4%−3%−5%−4%1%2%4%−4%−3%−2%−1%6%8%7%
      0 cm23 HU22 HU−4%24 HU23 HU−4%27 HU27 HU−3%26 HU26 HU0%22 HU21 HU−5%20 HU19 HU−2%3.5 mGy3.7 mGy6%
      −3cm3%2%−5%−2%−3%−4%2%1%−4%11%10%−1%6%5%−6%11%9%−3%−9%−8%7%
      −6cm11%9%−5%0%0%−4%7%6%−4%25%23%−2%15%14%−6%26%24%−4%−16%−15%7%
      Small breast3 cm4%5%−3%0%2%−2%−7%−4%3%0%3%−4%−6%−2%1%12%9%2%
      0 cm21 HU20 HU−4%25 HU24 HU−5%24 HU24 HU0%19 HU18 HU−7%19 HU18 HU−4%3.4 mGy3.5 mGy5%
      −3cm−1%−1%−5%2%3%−4%12%10%−2%5%4%−7%11%9%−5%−10%−8%7%
      −6cm3%3%−4%7%8%−3%25%22%−3%13%13%−7%26%24%−6%−16%−15%7%
      The mean HU values for different acquisitions, tube modulations and centerings were recorded for each individual ROI. The differences between two identical acquisitions (used for the noise estimations) for individual ROIs ranged from −1.1 to 1.0 HU. When comparing OBTCM against TCM-only, the ROI differences ranged from −1.1 to 1.4 HU. When different table heights were compared against the reference vertical centering (0 cm) the ROI mean differences varied from −4.3 to 2.0 HU. The largest absolute difference for the Siemens scanner was observed for the right medial large breast ROI (−4.3 HU at 3 cm table offset without OBTCM). The largest absolute difference for the GE scanner was seen for the posterior mediastinum ROI (−2.3 HU at 6 cm table offset with the large breast attachements, and without OBTCM).

      4. Discussion

      Breast tissue is one of the most radiosensitive organs in human body [

      Valentin J (ed.). The 2007 recommendation of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007;37(2-4):1–332.

      ]. Therefore, special attention should be given to protect glandular breast tissue from the detrimental effects of ionizing radiation in healthcare. The radiation dose for the breasts and for the entire patient can be reduced in CT scans by using various technical optimization tools. Regardless of technical advances, user-dependent practices (e.g. patient positioning into CT gantry) are currently even more important than before in CT optimization as many dose-reduction tools (e.g. bowtie filters, TCM, and automated tube voltage selection methods) assume patients to be positioned properly to scan isocenter. Several studies have reported remarkable variation in patient positioning, especially in vertical direction, resulting in changes in both exposure to patients and image quality [
      • Li J.
      • Udayasankar U.K.
      • Toth T.L.
      • Seamans J.
      • Small W.C.
      • Kalra M.K.
      Automatic patient centering for MDCT: Effect on radiation dose.
      ,
      • Toth T.
      • Ge Z.
      • Daly M.P.
      The influence of patient centering on CT dose and image noise.
      ,
      • Habibzadeh M.A.
      • Ay M.R.
      • Asl A.R.
      • Gharidi H.
      • Zaidi H.
      Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Lampinen A.
      • Kortesniemi M.
      Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Reijonen V.
      • Kortesniemi M.
      Effect of patient centering on patient dose and image noise in chest CT.
      ,
      • Kataria B.
      • Sandborg M.
      • Althén J.N.
      Implications of patient centring on organ dose in computed tomography.
      ,
      • Filev P.D.
      • Mittal P.K.
      • Tang X.
      • Duong P.A.
      • Wang X.
      • Small W.C.
      • et al.
      Increased computed tomography dose due to miscentering with use of automated tube voltage selection: phantom and patient study.
      ,
      • Saltybaeva N.
      • Alkadhi H.
      Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.
      ,
      • Marsh R.M.
      • Silosky M.S.
      The effects of patient positioning when interpreting CT dose metrics: A phantom study.
      ,
      • Akin-Akintayo O.O.
      • Alexander L.F.
      • Neill R.
      • Krupinksi E.A.
      • Tang X.
      • Mittal P.K.
      • et al.
      Prevalence and severity of off-centering during diagnostic CT: Observations from 57,621 CT scans of the chest, abdomen, and/or pelvis.
      ,
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ]. In this study, the impacts of patient vertical off-centering on the function of recently developed OBTCM techniques, and further on breast dose, were examined by using two CT systems from different vendors and two sizes of breast attachments applied for the anthropomorphic female body phantom.
      Previous studies have shown OBTCM to significantly reduce breast dose in chest CT [
      • Vollmar S.V.
      • Kalender W.A.
      Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.
      ,
      • Fu W.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Kapadia A.J.
      • Samei E.
      Breast dose reduction with organ-based, wide-angle tube current modulated CT.
      ,
      • Dixon M.T.
      • Loader R.J.
      • Stevens G.C.
      • Rowles N.P.
      An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.
      ,
      • Fu W.
      • Tian X.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Goodsitt M.M.
      • et al.
      CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.
      ,
      • Lungren M.P.
      • Yoshizumi T.T.
      • Brady S.M.
      • Toncheva G.
      • Anderson-Evans C.
      • Lowry C.
      • et al.
      Radiation dose estimations to the thorax using organ-based dose modulation.
      ,
      • Brand M.
      • Wuest W.
      • May M.
      • Uder M.
      • Sommer M.
      Influence of risk-organ-based tube current modulation on CT-induced DNA double-strand breaks in a biological phantom model.
      ,
      • Layman R.R.
      • Hardy A.J.
      • Kim H.J.
      • Chou E.N.
      • Bostani M.
      • Cagnon C.
      • et al.
      A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM).
      ]. However, the real anatomy of patients’ breasts in supine position can be very different from what has been seen in the phantom studies or when patient is in a standing or sitting position. Thus, part of the breasts may be drooping aside the chest and become vulnerable for higher exposure in chest CT. In this study, both the dose measurements and MC simulations revealed that breast dose, especially in lateral breast sections, may be increased in chest CT scans if the patient is not properly centered to the scan isocenter. Patient off-centering was shown to affect the function of OBTCM. For Siemens CT, the results showed that parts of the breast tissue may actually be exposed to higher amount of radiation dose when OBTCM is used compared to chest CT scans performed only with conventional TCM method. This results from the fact that Siemens is boosting photon flux from the posterior-side projections to maintain image noise. The tube-current boosting increased the overall scan CTDIvol up to 7%. The GE OBTCM method did not increase the breast dose nor any other organ doses at any vertical positions of a patient table. Moreover, the CTDIvol was decreased up to 17% in GE OBTCM chest CT scans compared to TCM-only scans, resulting in increased image noise to the images. If the image noise needs to be unchanged, the GE noise index should be decreased in OBTCM scans. This would correspondingly lead to an increased dose outside the OBTCM activated scan range. The differences in breast doses between the CT vendors OBTCM solutions can be explained by the corresponding differences in the OBTCM reduced dose angular ranges (projections with reduced photon flux) and distance range limits in z-direction. GE has a user-adjustable z-range coverage for the OBTCM function, whereas Siemens applies OBTCM for the entire scan range. More crucially for this study, a wider dose-reduction angular range was used in the GE system compared to that of Siemens. According to the manufacturer, 180° of the anterior-side projections are scanned with 40% reduced tube current values in GE Revolution EVO system while 180° of the posterior-side projections are scanned with the same tube current values as would be the case without OBTCM. In Siemens OBTCM solution, 80° of the anterior-side projections are scanned with 75% reduced tube currents, 40° of the anterior-side projections with 75% lower to 25% higher tube currents, and 240° of the posterior-side projections with 25% increased tube current values compared to a scan performed without OBTCM. Some differences between the vendors, as well as mismatch between the MOSFET and (idealized) MC doses may also arise from the OBTCM performance dependence on the imaging parameters. For example, increasing tube rotational velocity has been reported to decrease the angular range of the reduced exposure window [
      • Papadakis A.E.
      • Damilakis J.
      Evaluation of an organ-based tube current modulation tool in pediatric CT examinations.
      ]. The dose calculations involved several other sources of uncertainties which were related to the simulation software, the applied x-ray source models (e.g. bowtie filter, spectra and helical scan exposure geometry), material conversion, and the anthropomorphic phantom model. According to a previous study with similar methodology, the maximum difference between dose measurement and simulation results was found to be around 15% as applied to phantom data with ImpactMC program incorporating bowtie filter and spectra models [
      • Sundell V.M.
      • Kortesniemi M.
      • Siiskonen T.
      • Kosunen A.
      • Rosendahl S.
      • Büermann L.
      Patient-specific dose estimates in dynamic computed tomography myocardial perfusion examination.
      ]. Higher uncertainties can be anticipated if simulations are performed for clinical patient data with individual morphology, organ locations and shapes.
      Studies have shown that parts of the glandular breast tissue are usually left outside the dose-reduction zone in Siemens OBTCM solution [
      • Taylor S.
      • Litmanovich D.E.
      • Shahrzad M.
      • Bankier A.A.
      • Genevois P.A.
      • Tack D.
      Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone?.
      ,
      • Dixon M.T.
      • Loader R.J.
      • Stevens G.C.
      • Rowles N.P.
      An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.
      ,
      • Franck C.
      • Smeets P.
      • Lapeire L.
      • Achten E.
      • Bacher K.
      Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
      ]. Our study showed that patient off-centering below the scan isocenter worsened the situation and increased the breast dose further. This is also presented in Fig. 6 showing differential dose-volume histograms of the lateral and medial large breast regions with dose normalized to air kerma in isocenter. Systematic increase in dose-volume histogram peak is shown in both vendors scanners and with both breast regions when lowering the phantom and consequently bringing breast tissue closer to the central beam (in supine position) where the dose output is higher due to the bowtie filter. As anticipated, the effect is accentuated in medial breast region (Fig. 6A) which is closer to the isocenter as compared to lateral region. Further, the histogram peak is wider in medial region (especially with Siemens) on the lowest (-6 cm) vertical position when OBTCM is not used. On the other hand, for the lateral breast region (Fig. 6B) the dispersion of the histogram peak positions is higher across the vertical levels (combined with somewhat wider histogram peaks) with Siemens scanner when OBTCM is used. This reflects the narrower reduced angular dose range which does not extend to the lateral breast part. Consequently, the lateral breast region receives more heterogeneous dose distribution (represented by histogram peak width) and larger variance of histogram peak positions between different vertical levels. Such effect is not observed with lateral breast regions with GE scanner with wider angular OBTCM dose reduction zone or with Siemens when OBTCM is not used. As noted above, information shown in the Fig. 6 offer additional explanatory geometrical aspects of dose which are not obvious from the Table 1 involving relative changes in the mean doses in specified organ regions.
      The study results showed image noise to be maintained or reduced with Siemens OBTCM while increased with GE approach when positioning patient to the scan isocenter. The image noise results were approximately following the Poisson statistics. Image noise altered when positioning the phantom either above or below the scan isocenter. The change in image noise was a combination of changes observed in CTDIvol and functions of beam-shaping filters and OBTCM techniques. Siemens CT systems are known to use flatter beam-shaping filters than GE systems, and therefore the impacts of vertical off-centering on image noise were stronger in GE chest scans. When considering other image quality aspects, Fu et al. [
      • Fu W.
      • Sturgeon G.M.
      • Agasthya G.
      • Segars W.P.
      • Kapadia A.J.
      • Samei E.
      Breast dose reduction with organ-based, wide-angle tube current modulated CT.
      ] discussed that a potential artifact with GE OBTCM method could be introduced by photon starvation especially in the shoulder region or for patients with metal implants in clinical cases. We did not observe photon starvation induced artefacts in our phantom study (which did not involve metal implants). To summarize the contrast value changes: repeating a scan or enabling the OBTCM did not affect the studied contrast values (differences were at the order of 1 HU). However, large vertical off-centering may have a small effect on the HU values at the peripheral regions. These results were consistent with the findings of a previous study [
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ].
      The results of this study concurred with those of previous publications [
      • Li J.
      • Udayasankar U.K.
      • Toth T.L.
      • Seamans J.
      • Small W.C.
      • Kalra M.K.
      Automatic patient centering for MDCT: Effect on radiation dose.
      ,
      • Toth T.
      • Ge Z.
      • Daly M.P.
      The influence of patient centering on CT dose and image noise.
      ,
      • Kaasalainen T.
      • Palmu K.
      • Lampinen A.
      • Kortesniemi M.
      Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
      ,
      • Saltybaeva N.
      • Alkadhi H.
      Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.
      ,
      • Euler A.
      • Szucs-Farkas Z.
      • Falkowski A.L.
      • Kawel-Böhm N.
      • D’Errico L.
      • Kopp S.
      • et al.
      Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
      ]. Toth et al. [
      • Toth T.
      • Ge Z.
      • Daly M.P.
      The influence of patient centering on CT dose and image noise.
      ] showed anterior dose penalties of up to 18% and 41% when a 32 cm CTDI body phantom was vertically off-centered by 3 cm and 6 cm, respectively. Similarly, Li et al. [
      • Li J.
      • Udayasankar U.K.
      • Toth T.L.
      • Seamans J.
      • Small W.C.
      • Kalra M.K.
      Automatic patient centering for MDCT: Effect on radiation dose.
      ] showed surface dose to increase by up to 18% and 49% when a 32 cm cylindrical CTDI phantom was off-centered 3 cm and 6 cm below the isocenter, respectively. Kaasalainen et al. [
      • Kaasalainen T.
      • Palmu K.
      • Lampinen A.
      • Kortesniemi M.
      Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
      ] reported increased breast doses of up to 16% when off-centering a pediatric five-year-old anthropomorphic phantom 6 cm below the scan isocenter in chest CT using a fixed tube current value. Saltybaeva and Alkadhi [
      • Saltybaeva N.
      • Alkadhi H.
      Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.
      ], on the other hand, observed 38% increase in thyroid dose when off-centering an anthropomorphic phantom by 5 cm. Euler et al. [
      • Euler A.
      • Szucs-Farkas Z.
      • Falkowski A.L.
      • Kawel-Böhm N.
      • D’Errico L.
      • Kopp S.
      • et al.
      Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
      ] found in their simulations for a Siemens CT system using OBTCM in chest CT that breast dose increased up to 11% when off-centering a patient 1.8 cm below the scan isocenter. They also reported significantly higher relative doses to breasts for 75 years and older patients compared with younger patients. They assumed this to be related to aging process of breasts, leading to lower elasticity of the subcutaneous tissue, and therefore, resulting in a lateral drooping of the breasts in the supine position. This causes larger amount of glandular tissue to be positioned in the increased dose zone of Siemens OBTCM in chest CT examinations.
      Some of the previous studies have claimed the potential benefits of Siemens OBTCM technique to be overestimated in chest CT. For example, Frank et al. [
      • Franck C.
      • Smeets P.
      • Lapeire L.
      • Achten E.
      • Bacher K.
      Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
      ] stated that dose reduction to breasts may actually be less than 10% due to a limited tube-current reduction zone whereas posterior organs may absorb up to more than 25% more radiation, resulting in no reduction of radiation-induced malignancies. This observation was similar to our study results, as we found a 20% to 25% increase in the posterior mediastinum and posterior lung region doses in our MC simulations. Mussmann et al. [
      • Mussmann B.R.
      • Mørup S.D.
      • Skov P.M.
      • Brenøe A.S.
      • Eldahl F.
      • Jørgensen G.M.
      • et al.
      Organ-based tube current modulation in chest CT.
      ] also reported that Siemens OBTCM may not automatically be advantageous for the patients by observing higher breast doses with OBTCM than without it. They assumed this to be related partly on the lower pitch value used in the OBTCM chest CT protocol. Also, Layman et al. [
      • Layman R.R.
      • Hardy A.J.
      • Kim H.J.
      • Chou E.N.
      • Bostani M.
      • Cagnon C.
      • et al.
      A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM).
      ] observed that some females received higher breast doses with Siemens OBTCM than without it. One of the main reasons for this was lateral sections of glandular breast tissue being outside the dose reduction zone of the OBTCM and instead in the boosted tube current zone.
      This study has certain limitations. First and foremost, the study geometry was limited to only one relatively lean anthropomorphic phantom with two sizes of breasts. Considering the diversity in the anatomic characteristics, breast size, and body mass index among women, considerable variation can be anticipated in dosimetric and image noise estimations in different patient morphologies. The anthropomorphic phantom used in this study represented particularly small female patients with chest dimensions being without breasts attachments 20 cm × 25 cm. In larger patients, a higher portion of glandular breast tissue can be left outside the dose-reduction zone in Siemens OBTCM. Secondly, only two CT scanners from two vendors were examined and the conclusions from this study might not be generalizable to other implementations, different OBTCM angles and dose-reduction percentages. For example, GE uses different dose reduction angles in different CT models. Also, the CT scanners used in this study did not have automatic patient table positioning systems, which could potentially reduce the centering errors resulting in smaller dose and image quality variations between operators and patients. Third major limitation was that only acquisitions with AP scouts were studied. Lateral, posterior-to-anterior (PA) and AP scouts result in different geometric magnifications (especially for the spine), and therefore may have differing TCM dependences on the vertical off-centering magnitude as well as direction. According to a previous study, using the PA instead of AP scout direction resulted in higher breast doses when positioning the patient below the scan isocenter [
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ]. The lateral scout direction resulted in less dose variation when off-centering the patient but suffered from higher image noise than acquisitions with PA or AP scouts [
      • Kaasalainen T.
      • Palmu K.
      • Reijonen V.
      • Kortesniemi M.
      Effect of patient centering on patient dose and image noise in chest CT.
      ,
      • Kaasalainen T.
      • Mäkelä T.
      • Kortesniemi M.
      The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
      ]. Similar differences in behavior can be expected with OBTCM. Fourth, the organ doses were determined as average readings from the limited number of point measurements with MOSFET dosimeters. However, the voxel-based Monte Carlo simulations supported the findings of MOSFET measurements and yielded more comprehensive analysis of breast dose in varying off-centering levels. Also, not all radiosensitive organs and tissues (e.g. thyroid) were included in the dosimetric evaluation as the focus was mainly on the breast doses. Fifth, we did not have access to projection-specific tube current values on the CT systems, but the mA values used for MC simulations were determined using the information received from the CT vendors. Lastly, the noise was investigated by only considering HU variation and only limited contrast changes were considered. In general, image quality could also be affected by varying artefact strengths and changes in the noise textures. Estimating true clinical impact would require clinically relevant diagnostic tasks with realistic ranges of imaging target contrast levels, shapes, and sizes.
      Proper patient centering is an important factor from both dose optimization and image quality perspectives. These effects may be further emphasized when using OBTCM. Potential issues could be mitigated by proper choice of imaging parameters, such as scout direction, paying attention to proper centering in vocational training, or by utilizing automatic patient table positioning systems [
      • Booij R.
      • Budde R.P.J.
      • Dijkshoorn M.L.
      • van Straten M.
      Accuracy of automated patient positioning in CT using a 3D camera for body contour detection.
      ,
      • Saltybaeva N.
      • Schmidt B.
      • Wimmer A.
      • Flohr T.
      • Alkadhi H.
      Precise and automatic patient positioning in computed tomography: Avatar modeling of the patient surface using a 3-dimensional camera.
      ]. Future studies combining clinical images, relevant diagnostic tasks, and individual radiation doses are warranted to estimate the true clinical impact of various technical solutions.

      5. Conclusion

      Despite the CT vendors introducing novel CT optimization tools for dose reduction, users should be aware of their technical implementations and limitations. The results of this study provide a cautionary note for the notable differences between two vendors’ OBTCM implementations with direct consequences on dose and image noise. Specifically, patient off-centering was shown to affect the outcome of OBTCM in chest CT examination, and on some occasions, resulting in higher exposure. The use of modern dose optimization tools, such as OBTCM, emphasizes the importance of proper centering when preparing patients to CT scans.

      Declaration of Competing Interest

      The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

      References

      1. Valentin J (ed.). The 2007 recommendation of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007;37(2-4):1–332.

        • Kalender W.A.
        • Wolf H.
        • Suess C.
        Dose reduction in CT by anatomically adapted tube current modulation.
        II Phantom measurements Med Phys. 1999; 26: 2248-2253
        • Kalra M.K.
        • Maher M.M.
        • Toth T.L.
        • Schmidt B.
        • Westerman B.L.
        • Morgan H.T.
        • et al.
        Techniques and applications of automatic tube current modulation for CT.
        Radiology. 2004; 233: 649-657
        • Thibault J.B.
        • Sauer K.D.
        • Bouman C.A.
        • Hsieh J.
        A three-dimensional statistical approach to improve image quality for multislice helical CT.
        Med Phys. 2007; 34: 4526-4544
        • Kalender W.A.
        • Buchenau S.
        • Deak P.
        • Kellermeier M.
        • Langner O.
        • van Straten M.
        • et al.
        Technical approaches to the optimisation of CT.
        Phys Med. 2008; 24: 71-79
        • Vollmar S.V.
        • Kalender W.A.
        Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.
        Eur Radiol. 2008; 18: 1674-1682
        • Deak P.D.
        • Langner O.
        • Lell M.
        • Kalender W.A.
        Effects of adaptive section collimation on patient radiation dose in multisection spiral CT.
        Radiology. 2009; 252: 140-147
        • Duan X.
        • Wang J.
        • Christner J.
        • Leng S.
        • Grant K.L.
        • McCollough G.H.
        Dose reduction to anterior surfaces with organ-based tube current modulation: evaluation of performance in a phantom study.
        AJR Am J Roentgenol. 2011; 197: 689-695
        • Katsura M.
        • Matsuda I.
        • Akahane M.
        • Sato J.
        • Akai H.
        • Yasaka K.
        • et al.
        Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique.
        Eur Radiol. 2012; 22: 1613-1623
        • Miéville F.A.
        • Berteloot L.
        • Grandjean A.
        • Ayestaran P.
        • Gudinchet F.
        • Schmidt S.
        • et al.
        Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis.
        Pediatr Radiol. 2013; 43: 558-567
        • Schindera S.T.
        • Winklehner A.
        • Goetti R.
        • Fischer M.
        • Gnannt R.
        • Szucs-Farkas Z.
        Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: a phantom study.
        Clin Radiol. 2013; 68: e79-e86
        • Taylor S.
        • Litmanovich D.E.
        • Shahrzad M.
        • Bankier A.A.
        • Genevois P.A.
        • Tack D.
        Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone?.
        Radiology. 2015; 274: 260-266
        • Samei E.
        • Richard S.
        Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology.
        Med Phys. 2015; 42: 314-323
        • Spearman J.V.
        • Schoepf U.J.
        • Rottenkolber M.
        • Driesser I.
        • Canstein C.
        • Thierfelder K.M.
        • et al.
        Effect of automated attenuation-based tube voltage selection on radiation dose at CT: An observational study on a global scale.
        Radiology. 2016; 279: 167-174
        • Fu W.
        • Sturgeon G.M.
        • Agasthya G.
        • Segars W.P.
        • Kapadia A.J.
        • Samei E.
        Breast dose reduction with organ-based, wide-angle tube current modulated CT.
        J Med Imaging. 2017; 4031208
        • McLaughlin D.J.
        • Mooney R.B.
        Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the userśs needs?.
        Clin Radiol. 2004; 59: 446-450
        • Wang J.
        • Duan X.
        • Christner J.A.
        • Leng S.
        • Yu L.
        • McCollough C.H.
        Radiation dose reduction to the breast in thoracic CT: Comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current.
        Med Phys. 2011; 38: 6084-6092
        • Wang J.
        • Duan X.
        • Christner J.
        • Leng S.
        • Grant K.L.
        • McCollough G.H.
        Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT.
        Radiology. 2012; 262: 191-198
      2. AAPM Position statement on the use of bismuth shielding for the purpose of dose reduction in CT scanning. AAPM 2017. PP 26-B (https://www.aapm.org/publicgeneral/BismuthShielding.pdf) [accessed 5 February 2022].

        • Li J.
        • Udayasankar U.K.
        • Toth T.L.
        • Seamans J.
        • Small W.C.
        • Kalra M.K.
        Automatic patient centering for MDCT: Effect on radiation dose.
        AJR Am J Roentgenol. 2007; 188: 547-552
        • Toth T.
        • Ge Z.
        • Daly M.P.
        The influence of patient centering on CT dose and image noise.
        Med Phys. 2007; 34: 3093-3101
        • Habibzadeh M.A.
        • Ay M.R.
        • Asl A.R.
        • Gharidi H.
        • Zaidi H.
        Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies.
        Phys Med. 2012; 28: 191-199
        • Kaasalainen T.
        • Palmu K.
        • Lampinen A.
        • Kortesniemi M.
        Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom study.
        Pediatr Radiol. 2013; 43: 673-678
        • Kaasalainen T.
        • Palmu K.
        • Reijonen V.
        • Kortesniemi M.
        Effect of patient centering on patient dose and image noise in chest CT.
        AJR Am J Roentgenol. 2014; 203: 123-130
        • Kataria B.
        • Sandborg M.
        • Althén J.N.
        Implications of patient centring on organ dose in computed tomography.
        Radiat Prot Dosim. 2016; 169: 130-135
        • Filev P.D.
        • Mittal P.K.
        • Tang X.
        • Duong P.A.
        • Wang X.
        • Small W.C.
        • et al.
        Increased computed tomography dose due to miscentering with use of automated tube voltage selection: phantom and patient study.
        Curr Probl Diagn Radiol. 2016; 45: 265-270
        • Saltybaeva N.
        • Alkadhi H.
        Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.
        Med Phys. 2017; 44: 5697-5704
        • Marsh R.M.
        • Silosky M.S.
        The effects of patient positioning when interpreting CT dose metrics: A phantom study.
        Med Phys. 2017; 44: 1514-1524
        • Akin-Akintayo O.O.
        • Alexander L.F.
        • Neill R.
        • Krupinksi E.A.
        • Tang X.
        • Mittal P.K.
        • et al.
        Prevalence and severity of off-centering during diagnostic CT: Observations from 57,621 CT scans of the chest, abdomen, and/or pelvis.
        Curr Probl Diagn Radiol. 2019; 48: 229-234
        • Barreto I.
        • Lamoreux R.
        • Olguin C.
        • Quails N.
        • Correa N.
        • Rill L.
        • et al.
        Impact of patient centering in CT on organ dose and the effect of using a positioning compensation system: Evidence from OSLD measurements in postmortem subjects.
        J Appl Clin Med Phys. 2019; 20: 141-151
        • Euler A.
        • Saltybaeva N.
        • Alkadhi H.
        How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.
        Eur Radiol. 2019; 29: 6790-6793
        • Gang Y.
        • Chen X.
        • Li H.
        • Wang H.
        • Guo Y.
        • Zeng J.
        • et al.
        A comparison between manual and artificial intelligence-based automatic positioning in CT imaging for COVID-19 patients.
        Eur Radiol. 2021; 31: 6049-6058
        • Kaasalainen T.
        • Mäkelä T.
        • Kortesniemi M.
        The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.
        Eur J Radiol Open. 2018; 6: 24-32
        • Booij R.
        • Budde R.P.J.
        • Dijkshoorn M.L.
        • van Straten M.
        Accuracy of automated patient positioning in CT using a 3D camera for body contour detection.
        Eur Radiol. 2019; 29: 2079-2088
        • Saltybaeva N.
        • Schmidt B.
        • Wimmer A.
        • Flohr T.
        • Alkadhi H.
        Precise and automatic patient positioning in computed tomography: Avatar modeling of the patient surface using a 3-dimensional camera.
        Invest Radiol. 2018; 53: 641-646
        • Dixon M.T.
        • Loader R.J.
        • Stevens G.C.
        • Rowles N.P.
        An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.
        J Appl Clin Med Phys. 2016; 17: 380-391
        • Fu W.
        • Tian X.
        • Sturgeon G.M.
        • Agasthya G.
        • Segars W.P.
        • Goodsitt M.M.
        • et al.
        CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.
        Med Phys. 2017; 44: 665-678
        • Franck C.
        • Smeets P.
        • Lapeire L.
        • Achten E.
        • Bacher K.
        Estimating the patient-specific dose to the thyroid and breasts and overall risk in chest CT when using organ-based tube current modulation.
        Radiology. 2018; 288: 164-169
        • Lungren M.P.
        • Yoshizumi T.T.
        • Brady S.M.
        • Toncheva G.
        • Anderson-Evans C.
        • Lowry C.
        • et al.
        Radiation dose estimations to the thorax using organ-based dose modulation.
        AJR Am J Roentgenol. 2012; 199: W65-W73
        • Euler A.
        • Szucs-Farkas Z.
        • Falkowski A.L.
        • Kawel-Böhm N.
        • D’Errico L.
        • Kopp S.
        • et al.
        Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue.
        Eur Radiol. 2016; 26: 2656-2662
        • Lopez-Rendon X.
        • Zhang G.
        • Coudyzer W.
        • Develter W.
        • Bosmans H.
        • Zanca F.
        Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT?.
        Protocols dose comparisons Eur Radiol. 2017; 27: 4490-4497
        • Brand M.
        • Wuest W.
        • May M.
        • Uder M.
        • Sommer M.
        Influence of risk-organ-based tube current modulation on CT-induced DNA double-strand breaks in a biological phantom model.
        J Radiat Res. 2018; 59: 692-699
        • Mussmann B.R.
        • Mørup S.D.
        • Skov P.M.
        • Brenøe A.S.
        • Eldahl F.
        • Jørgensen G.M.
        • et al.
        Organ-based tube current modulation in chest CT.
        A comparison of three vendors Radiography (Lond). 2021; 27: 1-7
        • Koivisto J.
        • Wolf J.
        • Kiljunen T.
        • Schulze D.
        • Kortesniemi M.
        Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms.
        J Appl Clin Med Phys. 2015; 16: 266-278
        • Poludniowski G.
        • Landry G.
        • DeBlois F.
        • Evans P.M.
        • Verhaegen F.
        SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes.
        Phys Med Biol. 2009; 54 (N438): N433
        • Alikhani B.
        • Büermann L.
        Non-invasive experimental determination of a CT source model.
        Phys Med. 2016; 32: 59-66
        • Layman R.R.
        • Hardy A.J.
        • Kim H.J.
        • Chou E.N.
        • Bostani M.
        • Cagnon C.
        • et al.
        A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM).
        J Appl Clin Med Phys. 2021; 22: 97-109
        • Papadakis A.E.
        • Damilakis J.
        Evaluation of an organ-based tube current modulation tool in pediatric CT examinations.
        Eur Radiol. 2020; 30: 5728-5737
        • Sundell V.M.
        • Kortesniemi M.
        • Siiskonen T.
        • Kosunen A.
        • Rosendahl S.
        • Büermann L.
        Patient-specific dose estimates in dynamic computed tomography myocardial perfusion examination.
        Radiat Prot Dosimetry. 2021; 193: 24-36