Numerical modeling of air-vented parallel plate ionization chambers for ultra-high dose rate applications

  • Jose Paz-Martín
    Affiliations
    Departamento de Física de Partículas, Universidad de Santiago, Praza do Obradoiro, Santiago de Compostela, 15782, A Coruña, Spain
    Search for articles by this author
  • Andreas Schüller
    Affiliations
    Physikalisch-Technische Bundesanstalt, Bundesallee 100, Braunschweig, 38116, Germany
    Search for articles by this author
  • Alexandra Bourgouin
    Affiliations
    Physikalisch-Technische Bundesanstalt, Bundesallee 100, Braunschweig, 38116, Germany
    Search for articles by this author
  • Diego M. González-Castaño
    Affiliations
    Laboratorio de Radiofísica, Universidad de Santiago, Estrada de San Lourenzo, Santiago de Compostela, 15782, A Coruña, Spain
    Search for articles by this author
  • Nicolás Gómez-Fernández
    Affiliations
    Laboratorio de Radiofísica, Universidad de Santiago, Estrada de San Lourenzo, Santiago de Compostela, 15782, A Coruña, Spain
    Search for articles by this author
  • Juan Pardo-Montero
    Affiliations
    Group of Medical Physics and Biomathematics, Instituto de Investigacion Sanitaria de Santiago (IDIS), Travesía da Choupana, Santiago de Compostela, A Coruña, 15706, Spain

    Department of Medical Physics, Complexo Hospitalario Universitario de Santiago de Compostela, Travesía da Choupana, Santiago de Compostela, A Coruña, 15706, Spain
    Search for articles by this author
  • Faustino Gómez
    Correspondence
    Corresponding author at: Departamento de Física de Partículas, Universidad de Santiago, Praza do Obradoiro, Santiago de Compostela, 15782, A Coruña, Spain.
    Affiliations
    Departamento de Física de Partículas, Universidad de Santiago, Praza do Obradoiro, Santiago de Compostela, 15782, A Coruña, Spain

    Laboratorio de Radiofísica, Universidad de Santiago, Estrada de San Lourenzo, Santiago de Compostela, 15782, A Coruña, Spain
    Search for articles by this author
Open AccessPublished:October 28, 2022DOI:https://doi.org/10.1016/j.ejmp.2022.10.006

      Highlights

      • Numerical method to describe air ionization chambers at UHDR.
      • Free electron fraction as a function of dose per pulse and electrode distance.
      • Instantaneous current, ion collection time, free electron fraction for Roos chamber.
      • Comparative study of numerical and analytical models.
      • Effect of the time pulse structure on charge collection efficiency.

      Abstract

      Purpose:

      Air-vented ionization chambers have been the secondary standard for radiation dosimetry since the origins of radiation metrology. However, the feasibility of their use in ultra-high dose rate pulsed beams has been a matter of discussion, as large losses are caused by ion recombinations and no suitable theoretical model is available for their correction. The theories developed by Boag and his contemporaries since the 1950s, which have provided the standard ion recombination correction factor in clinical dosimetry, do not provide an accurate description when used under the limit conditions of ultra-high dose rates (UHDRs). Moreover, the high-ion recombination effects of ionization chambers under extreme dose-rate applications are an obstacle to the development of adequate dosimetry standards.

      Methods:

      In this article, the charge carrier transport equations within a parallel plate ionization chamber (PPIC) have been solved numerically with a double aim. First, this numerical model provides a more accurate tool that can be used to evaluate ion recombination correction for established PPICs in pulsed ultra-high dose rate regimes. Second, studying the chamber behavior in detail allow as to explore the limits of new chamber designs in order to improve their performance under UHDRs. The model presented here has been tested by measuring the instantaneous current of one unit of a Roos chamber (i.e., the time-resolved current during and after the irradiation pulse under UHDR conditions) and comparing these results with the absolute value of the simulated current.

      Results:

      The experimental data show consistent agreement with the results obtained using the numerical model. The experimental instantaneous current reveals effects such as the variation of the free electron fraction with the dose per pulse that are supported by the numerical model but cannot be explained in the framework of Boag’s theory.

      Conclusions:

      Numerical solutions of the charge carrier released and transport in ionization chambers are able to estimate the effects observed when PPICs are irradiated with ultra-high dose rate beams and to provide new insight into processes related to recombination losses at UHDRs. These models can be reliably extended to include regions where current analytical solutions are not valid. An agreement of better than 5 % between the experimental and simulated effective free electron fraction is found. We were able to reproduce the instantaneous current from a Roos chamber. The discrepancies observed between the experimental data and the numerical simulations can be attributed to the uncertainty about the transport parameters involved in the calculation.

      Keywords

      1. Introduction

      FLASH radiotherapy has arisen as a new form of therapy based on the seminal work of researchers including Fauvaudon et al. [
      • Favaudon V.
      • Caplier L.
      • Monceau V.
      • Pouzoulet F.
      • Sayarath M.
      • Fouillade C.
      • et al.
      Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice.
      ] who have examined the radiobiological advantages of ultra-short radiation delivery times. Existing scientific evidence only confirms that research is needed to clarify the limits, benefits and potential risks of this novel form of therapy [
      • Kim M.M.
      • Darafsheh A.
      • Schuemann J.
      • Dokic I.
      • Lundh O.
      • Zhao T.
      • et al.
      Development of ultra-high dose-rate (FLASH) particle therapy,vol. 6, no.3.
      ,
      • Lin B.
      • Gao F.
      • Yang Y.
      • Wu D.
      • Zhang Y.
      • Feng G.
      • et al.
      FLASH radiotherapy: History and future.
      ,
      • Esplen N.
      • Mendonca M.S.
      • Bazalova-Carter M.
      Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review.
      ,
      • Wilson J.D.
      • Hammond E.M.
      • Higgins G.S.
      • Petersson K.
      Ultra-high dose rate (FLASH) radiotherapy: Silver Bullet or fool’s gold?.
      ,
      • Jolly S.
      • Owen H.
      • Schippers M.
      • Welsch C.
      Technical challenges for FLASH proton therapy.
      ]. However, in the context of this new ultra-high dose rate application [
      • Schüler E.
      • Acharya M.
      • Montay-Gruel P.
      • Loo Jr. B.W.
      • Vozenin M.-C.
      • Maxim P.G.
      Ultra-high dose rate electron beams and the FLASH effect: From preclinical evidence to a new radiotherapy paradigm.
      ], many established ionization chambers currently used as a secondary standard for absorbed dose to water exhibit severe saturation problems [
      • Petersson K.
      • Jaccard M.
      • Germond J.-F.
      • Buchillier T.
      • Bochud F.
      • Bourhis J.
      • et al.
      High dose-per-pulse electron beam dosimetry — A model to correct for the ion recombination in the advanced Markus ionization chamber.
      ]. Thus, in order to define the parameters in which the FLASH effect is observed and to investigate its clinical feasibility, we require reliable, accurate and convenient dosimetry standards.
      The European UHDpulse project has the fundamental objective of developing and improving dosimetry standards for FLASH radiotherapy, very high energy electron (VHEE) radiotherapy and laser-driven medical accelerators. Such a challenge cannot be addressed in a unidirectional way. In fact, there is a collection of possible paths to perform metrology in the ultra-high dose rate range. These include dosimetry like Fricke, alanine, calorimetry, ionization chambers and solid state detectors [
      • Schüller A.
      • Heinrich S.
      • Fouillade C.
      • Subiel A.
      • De Marzi L.
      • Romano F.
      • et al.
      The European joint research project UHDpulse – metrology for advanced radiotherapy using particle beams with ultra-high pulse dose rates.
      ,
      • Gómez F.
      • Gonzalez-Castaño D.M.
      • Fernández N.G.
      • Pardo-Montero J.
      • Schüller A.
      • Gasparini A.
      • et al.
      Development of an ultra-thin parallel plate ionization chamber for dosimetry in FLASH radiotherapy.
      ,
      • Marinelli M.
      • Felici G.
      • Galante F.
      • Gasparini A.
      • Giuliano L.
      • Heinrich S.
      • et al.
      Design, realization, and characterization of a novel diamond detector prototype for FLASH radiotherapy dosimetry.
      ].
      In this paper, we will focus on the use of ionization chambers in the ultra-high dose rate (UHDR) regime for pulsed beams. At low doses per pulse, the performance and, more specifically, the charge collection efficiency (CCE) of plane-parallel chambers is well known. The incomplete charge collection is mainly due to the so-called volume recombination, which occurs when two ions with opposite charges neutralize themselves during the drift of the charge. Other contributions to the incomplete charge collection such as initial recombination and back-diffusion recombination are independent of the dose per pulse. The ion recombination correction factor is usually determined using the two-voltage method based on Boag’s formalism [
      • Boag J.W.
      Ionization measurements at very high intensities—Part I.
      ]. However, in the Intra-Operative Radiotherapy (IORT) and UHDR regimes, the corrections derived from the theories developed by Boag fail. When an ionization chamber is irradiated with a UHDR beam that has a large dose per pulse, the so-called space charge effect becomes more relevant [
      • Kranzer R.
      • Poppinga D.
      • Weidner J.
      • Schüller A.
      • Hackel T.
      • Looe H.K.
      • et al.
      Ion collection efficiency of ionization chambers in ultra-high dose-per-pulse electron beams.
      ]. Thus, a theory capable of reproducing the charge transport and collection in an ionization chamber must take into account this effects not considered in Boag’s theory in order to be accurate.
      Below, we present a computational model capable of handling these effects in the UHDR regime. By solving the transport equations, our aim is to obtain a more complete picture of the behavior of a parallel plate ionization chamber (PPIC). As we will describe in the following, accurate knowledge of transport parameters such as drift velocities, attachment to neutral molecules and volume recombination parameter is necessary for a precise description of the behavior of the ionization chambers. The results of the model presented are satisfying, bringing us closer to a deeper understanding of the processes taking place in PPICs at UHDR and to an accurate description of ion recombination effects in air ionization chambers under UHDR and conventional dose rates.

      2. Methods

      2.1 Mathematical description of charge carrier transport

      The charge carrier transport inside an air ionization chamber can be described via a system of coupled partial differential equations. In these equations, it is possible to allocate the different fundamental processes that take place in an ionization chamber during irradiation. The set of equations can be made more complex according to the level of detail that we want to assign to this description, since the actual number of species and processes is quite large [
      • Kossyi I.A.
      • Kostinsky A.Y.
      • Matveyev A.A.
      • Silakov V.P.
      Kinetic scheme of the non-equilibrium discharge in nitrogen- oxygen mixtures.
      ]. Nevertheless, it can be assumed that a summarized approach of the physics could be sufficient for most of the dosimetry properties of interest. The following simplifications have been adopted for the building of the model:
      • The electric field is considered homogeneous in the ionization chamber and no edge effects are considered.
      • The ionization chamber is symmetric along the transverse dimensions y and z. This allows us to write the system of differential equations as a one-dimensional effective problem along the coordinate x perpendicular to the electrode planes. Parallel-plate chamber geometry is considered ideal (i.e. electrodes are considered parallel).
      • The release of the electron–ions pairs in the chamber occurs homogeneously in space.
      • The simulated signal produced in the ionization chamber is caused by the drift of the charges towards the electrodes (i.e., primary electrons contributing to the measured signal are not considered).
      • The only contribution to the recombination considered is the volume recombination. Initial recombination, back-diffusion recombination and electron–ion recombination are neglected.
      • The drift of ionic species is summarized in only one specie for positive ions and another for negative ions. Drift velocity is considered equal to the average velocity in a given electric field.
      • Ion and electron drift velocity are function of the local electric field.
      n+(x,t)t=I(x,t)αn+(x,t)n(x,t)θn+(x,t)ne(x,t)++g(x,t)ve(x,t)ne(x,t)++xD+(x,t)n+(x,t)xxE(x,t)μ+n+(x,t),n(x,t)t=γne(x,t)αn+(x,t)n(x,t)++xD(x,t)xn(x,t)+xE(x,t)μn(x,t),ne(x,t)t=I(x,t)γne(x,t)θn+(x,t)ne(x,t)++g(x,t)ve(x,t)ne(x,t)++xDe(x,t)xne(x,t)+xve(x,t)ne(x,t),


      where ni (i=+,,e) are the positive ion, negative ion and electron densities in m−3. We denote E as the electric field across the chamber. In this work, we have not considered charge carrier multiplication in the chamber, taking the contribution from g(x,t) term as negligible. The equations take all the carrier species as being charged with the elementary charge denoted as e. The source term I(x,t) represents the number of charge carriers of either sign escaping initial recombination produced in the sensitive medium per unit volume and time in m−3 s−1. The parameter α in m3 s−1 accounts for the volume recombination effect between positive and negative ions in the same way that θ in m3 s−1 accounts for the positive ion and electron recombination. For this work, the parameters α and θ have been considered constant, although it could be argued that they may exhibit some dependence on the external electric field applied.
      The electron attachment process is governed by γ in s−1. The attachment coefficient (inverse of the attachment time τ) has been taken as a function of the electric field. The coefficients Di (i=+,,e) in m2 s−1 model the diffusion of charge carriers inside the chamber. Finally, ve is the drift velocity of electrons in m s−1 and μ+ and μ in m2 V−1 s−1 denote the mobility of positive and negative ions, respectively. In these equations, we have not included the de-attachment process and consider this effect to be negligible for the scope of this work.
      The electric field dependence across the chamber can be computed by solving the following one-dimensional Poisson equation:
      E(x,t)x=eϵn+(x,t)n(x,t)ne(x,t),
      (2)


      where ϵ is the dielectric constant of the medium. The bias voltage provided to the chamber sets an additional boundary condition for the electric field. Assuming an ideal behavior of the high voltage supply and neglecting other additional electrical effects (i.e. cable, connector, electrometer and chamber impedance), the electric field between electrodes separated by a distance d should fulfill the following boundary condition
      0dE(x,t)dx=Vt,
      (3)


      where V is the high voltage applied to the ionization chamber. The instantaneous current density in A m−2 produced at any time may be calculated using the Shockley–Ramo [
      • Ramo S.
      Currents induced by electron motion.
      ] theorem in the following way:
      IIC(t)=edΔx×i=0Nμ+E(xi,t)n+(xi,t)+μE(xi,t)n(xi,t)+ve(xi,t)ne(xi,t).
      (4)


      The numerical solution of this system of equations was realized via a Cython (version 0.29.28) code in which the equations were discretized in position and time. Similar approaches were proposed in order to reproduce computationally the behavior of the ionization chambers [
      • Aguiar P.
      • González-Castaño D.
      • Gómez F.
      • Pardo-Montero J.
      Recombination in liquid filled ionisation chambers with multiple charge carrier species: Theoretical and numerical results.
      ,
      • Gotz M.
      • Karsch L.
      • Pawelke J.
      A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
      ]. The discretization parameters were optimized to keep the error of the CCE below 0.1 % typically, with dN=Δx and ΔtΔxvmax usually with N= 3000 (see Fig. 1). This means that Δt is dynamically set in each simulation iteration according to vmax, which denotes the maximum speed of all carriers present in the area between the chamber electrodes at a given time.
      For the purpose of analyzing the convergence of the numerical solution of the system of partial differential equations, a simulation were carried out by introducing the charge release in the medium instantaneously, turning off the electric field perturbation and considering only two carriers (positive ions and negative ions). In this situation, we can compare the charge collection efficiency via the numerical model against the evaluation using Boag’s formula. Because this analysis does not involve the system as a whole, it allows us to obtain an absolute evaluation of the numerical error. Based on Fig. 1, 3000 spatial steps are sufficient to obtain a marginal numerical error compared to the uncertainties on the transport parameters.
      The model can be used to estimate any characteristic of a PPIC such as the CCE or the free electron fraction for a pulsed or continuous beam.
      Figure thumbnail gr1
      Fig. 1Numerical error obtained by comparing Boag’s formula with the results of the simulation. A 2 mm gap ionization chamber with a dose per pulse to air of 1 Gy and a 300 V bias voltage was used. The transport parameters employed for this comparison are discussed in detail in the following section.

      2.2 Model parameters

      One of the major problems we encounter when working with a computational model designed to obtain a satisfactory prediction of the experimental data is that the values of the physical variables involved must be known. Thus, in the model described above, we have diffusion, recombination and attachment coefficients together with mobility parameters for each charge carrier. The feasibility of the results of this model will be crucially dependent on the parametrization of these coefficients.
      We used results from the work of Boissonnat et al. [
      • Boissonnat G.
      • Fontbonne J.-M.
      • Colin J.
      • Remadi A.
      • Salvador S.
      Measurement of ion and electron drift velocity and electronic attachment in air for ionization chambers.
      ] to set the attachment parameter of the model. Average ion mobilities in air were taken from the work of Zhang et al. [
      • Zhang B.
      • He J.
      • Ji Y.
      Prediction of average mobility of ions from corona discharge in air with respect to pressure, humidity and temperature.
      ], in which full dependence on pressure, temperature and air humidity was considered. The Boissonnat mobilities were taken as an input parameter of the model to compare the simulation results against the experimental data. The electron drift velocity ve was simulated using the Magboltz code [
      • Biagi S.
      Monte Carlo simulation of electron drift and diffusion in counting gases under the influence of electric and magnetic fields.
      ] (version 11.14).
      Considering the conventional Nernst–Townsend formula, the diffusion coefficients can be written as D=k(273.2+T)μe, where k is the Boltzmann constant and μ is the mobility. Although the diffusion coefficients of ions are three orders of magnitude smaller than those of electrons, the standard deviation, due to the space diffusion after traveling a distance d, would not be dependent on the mobility in a first order approximation and would amount to
      σdphys=2k(273.2+T)deE
      (5)


      By contrast, the analogous numerical diffusion contribution [
      • Gotz M.
      • Karsch L.
      • Pawelke J.
      A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
      ] for a gap discretization in N steps of the distance d between electrodes has an approximate expression of
      σdnumdN.
      (6)


      Taking σdphys=σdnum, this leads to the relationship for the step distance
      dN=2k(273.2+T)eE0.170μm,
      (7)


      where we have considered a reference numerical figure corresponding to a voltage bias of 300 V evaluated at 20 °C. Using a distance between electrodes d of 1 mm, the number of steps is equal to
      N=eV2k(273.2+T)6000.
      (8)


      As a result of the limit with equal numerical and physical diffusion, the number of steps N needed to discretize a chamber does not depend on the distance d between electrodes but only on its voltage bias V. Normally, this discretization limit would yield long computing times for the simulation. For this reason, larger spatial step sizes with N3000 are applied, thus provoking numerical diffusion to dominate the overall diffusion contributions. In this scenario, we have neglected to first order the physical diffusion contribution in Eq. (1).
      It should be noted that Nernst–Townsend formula is only valid when the charge carriers are in thermal equilibrium with the surrounding molecules. Considering the electric field regime in which established ionization chambers usually operate, this is true for positive and negative ions but not for electrons [
      • Sauli F.
      ]. It is to be expected that the diffusion due to electrons is considerably greater than that predicted by the Nernst–Townsend relation and, therefore, that the number of steps necessary to equalize both physical and numerical contributions in the case of electrons is lower.
      The volume recombination coefficient α in air has values in the literature [
      • Gotz M.
      • Karsch L.
      • Pawelke J.
      A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
      ] from 0.9×1012 m3 s−1 up to 1.3×1012 m3 s−1; however, this coefficient was set to a value of 1.25×1012 m3 s−1 for theoretical examples. The volume recombination coefficient α may also exhibit some dependence on the electric field, temperature and humidity [
      • Franchin A.
      • Ehrhart S.
      • Leppä J.
      • Nieminen T.
      • Gagné S.
      • Schobesberger S.
      • et al.
      Experimental investigation of ion–ion recombination under atmospheric conditions.
      ]; this dependence has not been included in this model. The recombination between ions and electrons is not considered here, as its contribution is estimated to be within the uncertainties of the transport parameters.
      In the following simulations, unless not explicitly mentioned, standard temperature, pressure and humidity (20 °C, 101.325 kPa and 50 % relative humidity) were used.

      2.3 Free electron fraction

      As shown above, the free electron fraction (FEF) parameter p plays a major role in describing the behavior of PPICs. In addition, Boag derived a simple expression that allows the calculation of the FEF, p, if the characteristic electron properties are known:
      p=QeQ0=veτd1edveτ,
      (9)


      where Qe is the total free electron charge arriving to the positive electrode and Q0 is the total charge of either sign escaping initial recombination produced by the ionizing radiation per pulse. However, due to the fact that, in the UHDR regime, the electric field disturbance is very important, the FEF is no longer constant. In fact, by varying the dose per pulse to which the PPIC is exposed, it is possible to observe significant changes in the FEF.
      The FEF p is defined with respect to the total charge released by ionization that escapes initial recombination. We can also define a fraction in relation to the charge actually collected, which we denote as p. The relationship between these two quantities is given by:
      p=pksat;p=QeQcol;p=QeQ0,
      (10)


      where p denotes the ratio of the free electron charge Qe that reaches the positive electrode to the total collected charge Qcol, with p being the standard definition (see Eq. (9)) in Boag et al. formalism. The ksat parameter corresponds to the inverse of the CCE. The behavior of these two fractions for different PPICs with different electrode distances (as PPC05: 0.6 mm, Advanced Markus: 1 mm, Roos: 2 mm) can be seen in Fig. 3A and B. Even for the smallest gap chamber of 0.6 mm, the simulated p starts to decrease significantly when the dose per pulse exceeds 0.1 Gy. The Boag FEF p monotonically reduces its value with the dose per pulse in contrast to the original Boag model, where p is a constant (dashed lines).
      If the free electron fraction p is considered, this fraction tends to increase at very high doses per pulse. This is due to the fact that, when large volume recombination is present, the ion collected charge decreases more dramatically than the corresponding free electron charge. For chambers with small gaps or large electron fractions, p reaches its minimum at a dose where ion recombination losses begin to be dominant (Fig. 3B).
      Figure thumbnail gr2
      Fig. 2Drift velocity of electrons in dry air and humid air simulated using the Magboltz code (lines). The symbol xH2O represents the molar fraction of water in air. Experimental data of Davies [
      • Davies D.
      Measurements of swarm parameters in dry air.
      ,
      • Davies D.
      • Chantry P.
      Air chemistry measurements II.
      ] and Reid  and the parametrization of Laitano  are shown for comparison.
      Figure thumbnail gr3
      Fig. 3Simulation of different free electron fraction definitions as a function of the dose per pulse for chamber gaps between 0.6 mm and 2 mm at 300 V bias voltage and 2.5μs pulse duration. The straight dotted line illustrates the prediction expected by the analytical formulas.
      Figure thumbnail gr4
      Fig. 4On the left, the effect on CCE of structure (δ from 0 to 1) and pulse duration (Δt from 2.5μs to 100μs) is shown for a 2 mm chamber at 300 V bias voltage with a 0.5 Gy dose per pulse to air. On the right, a scheme explains the dose rate profile used to study the pulse structure effect. The non-dimensional parameter δ is equal to Δt1/Δt and Ḋ1=(1δ)D0Δt1.
      The free electron fraction used in Boag’s formalism does not correspond to the contribution to the total charge actually collected due to the presence of free electrons in the ionization chamber. The measured charge is the result of the electrically induced signal of the charge carriers that can be calculated using the above-mentioned Shockley–Ramo theorem. In this case, the ratio of the signal induced by the electrons to the total charge released in the medium that escapes the initial recombination is given by:
      pind=veτd1p
      (11)


      As in the previous case, it is convenient to define the alternative quantity pind based on the total collected charge:
      pind=pindksat
      (12)


      The dependence of these two quantities, pind and pind, on the dose per pulse can be seen in Fig. 3C and D. Solid lines represent the full simulation of the induced charge due to free electrons in the ionization chamber. While Boag’s free electron fraction tends to unity (Fig. 3A) for a very small distance between electrodes, the signal induced due to electrons would contribute to half the chamber readout value for uniform irradiation (Fig. 3C). The quantity pind represents the actual fractional contribution of the drifting free electrons to the electrometer readout.

      2.4 Pulse time structure

      In many cases, the ion recombination correction factor is considered as an overall function of the dose per pulse delivered by the accelerator. One of the underlying questions in evaluating recombination effects is the extent to which the recombination is affected by the pulse-time structure. The dose rate variation during the pulse delivery may have a significant effect on the actual collected charge. Due to the possibility large variations of the dose rate within the pulse duration, we have chosen a simple test template. We considered the fact that the pulse of duration Δt is divided into two subintervals with different dose rate as (see right part of Fig. 4):
      Ḋ1=1δδD0Δt;Ḋ2=δ1δD0Δt0<δ<1
      (13)


      Δt1=δΔt;Δt2=(1δ)Δt
      (14)


      The first subinterval delivers a dose rate Ḋ1 in a time interval Δt1 and the second Ḋ2 in Δt2. This allows us to quantify the impact of the pulse structure in a simple way and show its impact in the final collection efficiency. Thus, the non-dimensional delta parameter controls the shape of the pulse. When δ = 0.5 the dose rate is constant over the entire pulse duration while having a short and ultra high dose rate when δ0 and δ1 at the beginning or end of the pulse, respectively. The results were also analyzed in terms of different pulse durations Δt.
      The results for a 2 mm electrode distance operated at 300 V with a dose per pulse of 0.5 Gy can be seen in Fig. 4. For the pulse duration of conventional medical accelerators, which is typically less than 5μs, the variation of the charge collection efficiency is below 3 % when the pulse structure is changed from an instantaneous dose delivery to a constant dose rate along the pulse. However, when using pulses with longer durations, relative variation of the ion recombination correction factor of the chamber can be dramatic (i.e., around 20 % for 100μs).
      It is important to realize that Boag’s models do not take into account either pulse duration or pulse structure. Thus, it is important to verify the variation in charge collection efficiency of a chamber when the pulse duration/structure is altered in a specific irradiation facility. Due to the relatively low ion mobilities, variations within the pulse duration over periods of time shorter than the ion drift time do not significantly affect the chamber performance. However, when the duration of the pulse approaches the ion drift time dμE, the recombination effect can exhibit a large variation.

      3. Experimental validation

      Some of the existing experimental data on the recombination effect concern only ion recombination correction factors or CCE [
      • Petersson K.
      • Jaccard M.
      • Germond J.-F.
      • Buchillier T.
      • Bochud F.
      • Bourhis J.
      • et al.
      High dose-per-pulse electron beam dosimetry — A model to correct for the ion recombination in the advanced Markus ionization chamber.
      ,
      • McManus M.
      • Romano F.
      • Lee N.D.
      • Farabolini W.
      • Gilardi A.
      • Royle G.
      • et al.
      The challenge of ionisation chamber dosimetry in ultra-short pulsed high dose-rate very high energy electron beams.
      ]. A detailed numerical model is also able to provide a dynamic solution for the electrically induced current in air ionization chambers. Our study of the chamber electrical current description is much more demanding than a study of the overall balance of charge. The instantaneous current (i.e., the time resolved current during and shortly after the irradiation pulse), depends not only on the density of electrons, negative and positive ions but also on their mobility and on the accurate description of the electric field during the carrier drift. For this reason, the experimental measurement of the instantaneous current constitutes a benchmark for numerical models.
      For our experiments, we used one unit of a Roos chamber (Serial No. 2981) irradiated using PTB’s ultra-high pulse dose rate reference electron beam [
      • Bourgouin A.
      • Knyziak A.
      • Marinelli M.
      • Kranzer R.
      • Schüller A.
      • Kapsch R.-P.
      Characterization of the PTB ultra-high pulse dose rate reference electron beam.
      ,
      • Schuller A.
      • Makowski C.
      • Kapsch R.-P.
      • Nolte R.
      • Beck P.
      MELAF -a 50 MeV electron accelerator facility for research in radiation effects.
      ] with a nominal electron energy of 20 MeV, doses per pulse ranging from 0.2 Gy to 5.8 Gy and a fixed pulse duration of 2.5μs. The chamber was operated at either a negative and a positive polarity from 100 V up to 450 V. The time-resolved beam current during each electron beam pulse was measured by means of an integrating current transformer (ICT from Bergoz instrumentation) connected to a transient recorder (Spectrum M3i.4142). The instantaneous current from the Roos chamber was recorded via a time-resolved measurement of the voltage drop across a 50 Ω resistor connected directly at the chamber output in combination with a preamplifier (FEMTO HVA-200M-40-F) connected to the second input channel of the same transient recorder as used for beam current measurement. The dose rate (Ḋ(t)) at the position of the ionization chamber was obtained from the instantaneous beam current (Ibeam(t)). This was achieved via a previous cross calibration of the charge per beam pulse (0ΔtIbeam(t)dt) against the dose per pulse measured by means of alanine pellets at the position of the ionization chamber [
      • Bourgouin A.
      • Schüller A.
      • Hackel T.
      • Kranzer R.
      • Poppinga D.
      • Kapsch R.-P.
      • et al.
      Calorimeter for real-time dosimetry of pulsed ultra-high dose rate electron beams.
      ]:
      Ḋ(t)=Dw0ΔtIbeam(t)dtIbeam(t)0tΔt
      (15)


      where Dw denotes the dose per pulse to water at the reference depth of the ionization chamber and Δt denotes the pulse duration. This dose rate profile was introduced in the numerical simulation in order to predict the chamber current via the Shockley–Ramo theorem. The nominal collecting electrode area of the chamber, whose dimension is 1.91 ×104 m2, was used to obtain the absolute value of the instantaneous current in order to compare it to the measured time-resolved current. For the conversion of the dose rate into a liberated charge inside the medium per unit of time, the absorbed dose-to-water calibration coefficient ND,w was measured using PTB’s calibrated Co-60 source [
      • Krauss A.
      The PTB water calorimeter for the absolute determination of absorbed dose to water in 60co radiation.
      ] and the corresponding beam quality correction factors were calculated using a Monte Carlo model of the beam used [
      • Bourgouin A.
      • Knyziak A.
      • Marinelli M.
      • Kranzer R.
      • Schüller A.
      • Kapsch R.-P.
      Characterization of the PTB ultra-high pulse dose rate reference electron beam.
      ].
      Fig. 5 shows examples of the experimental data. Graph A shows an example of the beam current measured by means of the ICT and the simultaneously measured instantaneous current, which is produced in the ionization chamber. Graph B shows the instantaneous currents at different doses per pulse. Graph C shows the effect of the operating voltage on the ion collection time.
      The instantaneous current can be divided into two distinct parts. The fast current is almost able to follow the intra pulse beam current variations, aside from the fact that the fast current always higher than the beam current at the beginning of the pulse. The existence of the fast current part is due to the high mobility of the free electrons in the medium. Subsequently, after the end of the beam pulse, a small and long-lasting tail is observed. This is due to the slow ions, whose mobility is several orders of magnitude lower than the mobility of the electrons.
      Figure thumbnail gr5
      Fig. 5(A) Example of the time-resolved beam current and the simultaneously measured current produced in the ionization chamber. (B) Time-resolved chamber current at different DPPs. (C) Integrated chamber current normalized to the total charge for different operating voltages and thus different ion collection times.
      Thus, by taking the area under the fast current curve over the integral of the total signal, it is possible to estimate an effective free electron fraction (EFEF), peff as
      peff=0ΔtIIC(t)dt0IIC(t)dt=0ΔtIe(t)dt+0ΔtIion(t)dt0IIC(t)dt=pind+0ΔtIion(t)dt0IIC(t)dt
      (16)


      where Ie denotes the induced current due to electrons, Iion denotes the current induced by ions (positive and negative) and IIC the total induced current. The EFEF is the integral of the instantaneous current up to the end of the 2.5μs beam pulse over the integral of the total induced current. The same procedure was followed for the simulated current curves. The relative contribution of the instantaneous positive and negative ion current to peff is always lower than 10 %. For the simulation using the Zhang et al. ion mobilities, a volume recombination parameter of 1.15 × 10−12 m3 s−1 was chosen as it showed good agreement between the simulated and experimental data. Fig. 6 compares the experimental instantaneous current and the simulated current. The contribution of each charge carrier to the signal is also plotted. On the right, the integral of the simulated and experimental signals is shown. The dashed area shows the part of the integral used for the calculation of the EFEF.
      The comparison between the experimental and simulated EFEFs can be seen in Fig. 7. The definition of the effective free electron fraction represents a figure of merit for the electron current during the pulse delivery. Experimental results indicate that the free electron fraction is independent of the bias voltage for the dose per pulse over 2 Gy. Fig. 7 includes experimental results that can be compared to Fig. 3D for the 2 mm electrode distance that corresponds to a Roos chamber. The results for peff in Fig. 7 differ from pind shown in Fig. 3D by less than 10 % due to the integration of the ion-induced current during the radiation pulse.
      Figure thumbnail gr6
      Fig. 6The graphs on the left compare experimental and simulated PPIC instantaneous electrical currents for different doses per pulse and operational voltages. For the simulation, the contributions of electrons, positive ions and negative ions are plotted separately. The vertical dashed line in the graphs on the left represents the pulse end time. In the graphs on the right, the integral of the experimental and simulated signal is shown. The dashed area represents the time interval considered for the calculation of the EFEF. The simulation presented here was performed using the mobilities of Zhang et al. for (A) −400 V and 4.82 Gy, (B) −401.9 V and 0.46 Gy and (C) −100.6 V and 0.49 Gy.
      These results show an unequivocal variation of the EFEF with the dose per pulse. This is also supported by a numerical simulation that shows qualitatively good agreement with the experimental data. The differences may be attributed to the uncertainties of the transport parameters used.
      Figure thumbnail gr7
      Fig. 7Comparison between the experimental and the simulated EFEF (according to definition of Eq. ). Zhang et al. ion mobilities were used with a volume recombination coefficient of 1.15 ×1012 m3 s−1. Simulated data is presented as a gray exclusion zone to avoid overloading the figure.
      The ion collection time was defined here as the time required in order to collect 99.95 % of the total charge (dashed line in Fig. 5C). The results of the simulation using the Boissonnat and the Zhang et al. ion mobility measurements are shown in Fig. 8 compared to the experimental values.
      It can be observed that the collection times obtained from the simulations differ by up to 20 % from those measured experimentally, suggesting that possibly a value between both used parametrizations may be more adequate than the effective positive ion mobility (i.e., the ions with lower mobility). A particularly remarkable finding is that the ion collection times show no significant variation with the dose per pulse (each point in Fig. 8 displays the average and uncertainty bar corresponding to the dose per pulse from 0.2 Gy up to 5.8 Gy), although they are affected differently by the dose-dependent electric field perturbation inside the ionization chamber and entangled with the volume recombination parameter.
      Many of the numerical models considered previously have only considered the recombination effect [
      • Kranzer R.
      • Poppinga D.
      • Weidner J.
      • Schüller A.
      • Hackel T.
      • Looe H.K.
      • et al.
      Ion collection efficiency of ionization chambers in ultra-high dose-per-pulse electron beams.
      ,
      • Aguiar P.
      • González-Castaño D.
      • Gómez F.
      • Pardo-Montero J.
      Recombination in liquid filled ionisation chambers with multiple charge carrier species: Theoretical and numerical results.
      ,
      • Gotz M.
      • Karsch L.
      • Pawelke J.
      A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
      ]. In the previous paragraphs, we have seen that the correct prediction of the instantaneous current is a much more demanding task and requires the proper calculation of the charge carrier densities, the electric field perturbation and drift speed. Alternatively, the results comparing the modeled and experimental CCEs can be seen as a function of the dose per pulse in Fig. 9 for several positive operational voltages. The CCE was calculated using the following expression:
      CCE=0IIC(t)dtkQND,w60CokfieldDw,
      (17)


      where kQ is the beam quality correction factor obtained from [
      • Muir B.R.
      • Rogers D.W.O.
      Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types.
      ], ND,w60Co is the absorbed dose to water calibration coefficient in 60Co and kfield is the field size correction factor for the non uniformity of the beam profile. The agreement shown in Fig. 9 is, in general, better than 2 %. Likewise, in Fig. 10, the simulated and measured CCEs are shown as a function of the polarization voltage (including positive and negative polarity) for three different doses per pulse (0.27, 0.72 and 3.6 Gy). This suggest that the ion recombination correction ksat may be calculated for a Roos chamber using the numerical approach described in this work.
      Figure thumbnail gr8
      Fig. 8Comparison between the experimental ion collection times and the simulated ion collection times using the ion mobilities reported by Boissonnat and Zhang et al. The volume recombination coefficient used with the Zhang et al. mobilities was 1.15 ×1012 m3 s−1 and with Boissonnat ones 1.8 ×1012 m3 s−1. The uncertainty bars show the standard deviation of the values due to variation in ambient conditions (pressure, temperature and humidity) and for dose per pulse varying from 0.2 Gy up to 5.8 Gy using a coverage factor of 2.
      Figure thumbnail gr9
      Fig. 9Comparison of experimental determined and simulated charge collection efficiency for a Roos chamber as function of dose per pulse for different operational voltages.
      Figure thumbnail gr10
      Fig. 10Comparison of experimental determined and simulated charge collection efficiency for a Roos chamber as function of operational voltage for different dose per pulse.

      4. Discussion

      The results obtained via the numerical model showed good agreement with the experimental data from a Roos ionization chamber in UHDR; the numerical solution of the transport equations is able to qualitatively reproduce the effects observed on the instantaneous current as seen in Fig. 6. We have defined an alternative effective free electron fraction based on the contribution of free electron induction to the total charge collected in the ionization chamber. This quantity differs from the original Boag definition, which amounts to the fraction of free electrons produced in the gas reaching the positive electrode. To evaluate the effective free electron fraction, we integrated the instantaneous current during the pulse delivery time of 2.5μs both experimentally and in the simulation. The model shows good agreement with the defined effective free electron fraction of about ± 5 %. The experimental and simulated charge collection efficiency also show accordance of about ± 2 %. We have to underline that the use of an incorrect value of electron drift velocity (see humid versus dry air in Fig. 2) in an intermediate region of the electric field such as 50 V mm−1 may yield errors greater than 50 % in the electron induced instantaneous current. In fact, the need to review and upgrade the electron drift velocity was a consequence of the comparison of the first version of the numerical model with the experimental data.
      When evaluating the ion collection times, we were able to benchmark the average ion mobilities used; there were discrepancies between the experimental and calculated ion times of around 20 %. In general, the average ion mobilities measured by Zhang et al. provided slightly better agreement with the experimental determined ion collection times and with the instantaneous current of the ionization chamber than the Boissonnat parametrization as shown in Fig. 8. Of particular interest is the fact that the ion collection times follow a linear relationship with the inverse of the operational voltage, even when the electric field is strongly perturbed. For large doses per pulse greater than 2 Gy, ion collection times tend to be independent of the dose per pulse. In general, the discrepancies observed between the simulation and the experimental data can be attributed to the uncertainty of the charge transport parameters used in the model such as the average ion mobilities in air, the uncertainties of the nominal value of the area and distance between electrodes and the simplifications assumed in the model.
      The polarity effect has a well know contribution to the charge readout of parallel plate ionization chambers [
      • Abdel-Rahman W.
      • Seuntjens J.P.
      • Verhaegen F.
      • Podgorsak E.B.
      Radiation induced currents in parallel plate ionization chambers: Measurement and Monte Carlo simulation for megavoltage photon and electron beams.
      ] either as a Compton current or voltage dependent contribution. Unfortunately the version of the numerical model used in this work does not accommodate this effect and the data was compared with the simulation separately for each polarity without further correction. Small polarity differences can be seen in the comparison with experimental data both for the EFEF (Fig. 7) and CCE (Fig. 10). In general terms, the average experimental relative discrepancy between chamber readout at opposite polarity for similar dose per pulse is around 1.7 % for our data set. Considering the uncertainties of the experimental data it is not possible to establish a significant voltage dependence of such polarity effect. Nevertheless, due to the large dose per pulse delivered, we cannot discard that, depending on the high voltage supply and electronic readout chain, some perturbation of the drift electrode, collecting electrode and guard ring voltages may appear.
      The evaluation of the contribution of ion and free electrons to the ionization chamber readout via the numerical model has been performed using the Shockley–Ramo theorem. To this end, we have defined the contribution of free electrons to the chamber signal pind with respect to the saturation charge and also pind with respect to the collected charge. Predictions for these quantities are shown in this work as a function of the dose per pulse and the distance between electrodes for a PPIC. Our simulation could potentially be used to study the effect of the pulse time structure on the chamber response. To this end, we have taken a simple approach dividing a given dose per pulse into in two time subintervals with different instantaneous dose rates. Whenever the delivery time is much smaller than the ion collection time (i.e., below 3μs for a Roos chamber), the pulse time structure effect in the collected charge is negligible. Alternatively, if the duration of the radiation pulse is comparable with the charge carrier drift time the effect of the pulse time structure can be significant.

      5. Conclusions

      In this paper, the design and results of a computational model for evaluating the performance of a plane-parallel plate ionization chamber under pulsed beams have been presented. We have used an electron UHDR beam for the experimental validation of this numerical model. Instead of only testing the charge collection efficiency as a summarized way to benchmark the model [
      • Gotz M.
      • Karsch L.
      • Pawelke J.
      A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
      ], the electrical instantaneous current due to the charge carrier drift in a Roos chamber was also measured up to full the ion collection time. The instantaneous dose rate in the chamber, which was needed for the numerical simulation, was taken from the non-destructive time-resolved beam current measurement along with the alanine calibration of the dose to water. This methodology allowed us to obtain the expected time-dependent instantaneous current from numerical simulations and to compare it with the experimental data. This method proved to be a more demanding test for this type of models being sensitive not only to the produced charge carrier densities and recombination coefficient, but also to the drift velocities of ions and electrons in the local electric field inside the chamber. The total ion collection time has small variation with the total dose per pulse (from 0.2 Gy up to 5.8 Gy) and exhibits a linear dependence with the inverse of the chamber bias voltage in spite of the large perturbation of the electric field inside the chamber. Moreover, this numerical approach can be used to study the effect of the time structure of the radiation delivery on the chamber readout.
      We have also reviewed the Boag definition of the free electron fraction p which does not correspond to the fraction of the total collected charge due to the free electron drift inside the chamber that we have denominated pind. To verify experimentally this contribution we have measured the integrated charge during a pulse of 2.5μs, dominated by the free electron component, produced at the PTB electron accelerator facility MELAF and compared it to the total charge collected in a Roos chamber. This ratio, that we denominate the effective free electron fraction, depends on the dose per pulse and the agreement between the numerical model and measured data is in general better than 5 %.
      This work demonstrates the great advantage of numerical tools for updating the current understanding of ionization chambers and describing their behavior in exhaustive detail, even in extreme regimes such as those of UHDRs. Specifically, this work supports the possibility of calculating the ion recombination correction factors in regimes where Boag’s theories are not valid using the numerical approximation described. Unfortunately, the poor knowledge of the transport coefficient currently limits the accuracy of these models. Thus, further investigation of such quantities is required in order to produce a high precision tool for the medical physics community.

      Acknowledgments

      This work has received funding from the EMPIR programme project 18HLT04 UHDpulse co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme .

      References

        • Favaudon V.
        • Caplier L.
        • Monceau V.
        • Pouzoulet F.
        • Sayarath M.
        • Fouillade C.
        • et al.
        Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice.
        Sci Transl Med. 2014; 6: 245ra93https://doi.org/10.1126/scitranslmed.3008973
        • Kim M.M.
        • Darafsheh A.
        • Schuemann J.
        • Dokic I.
        • Lundh O.
        • Zhao T.
        • et al.
        Development of ultra-high dose-rate (FLASH) particle therapy,vol. 6, no.3.
        2022: 252-262https://doi.org/10.1109/TRPMS.2021.3091406
        • Lin B.
        • Gao F.
        • Yang Y.
        • Wu D.
        • Zhang Y.
        • Feng G.
        • et al.
        FLASH radiotherapy: History and future.
        Front Oncol. 2021; 11https://doi.org/10.3389/fonc.2021.644400
        • Esplen N.
        • Mendonca M.S.
        • Bazalova-Carter M.
        Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review.
        Phys Med Biol. 2020; 65https://doi.org/10.1088/1361-6560/abaa28
        • Wilson J.D.
        • Hammond E.M.
        • Higgins G.S.
        • Petersson K.
        Ultra-high dose rate (FLASH) radiotherapy: Silver Bullet or fool’s gold?.
        Front Oncol. 2020; 9https://doi.org/10.3389/fonc.2019.01563
        • Jolly S.
        • Owen H.
        • Schippers M.
        • Welsch C.
        Technical challenges for FLASH proton therapy.
        Physica Med. 2020; 78: 71-82https://doi.org/10.1016/j.ejmp.2020.08.005
        • Schüler E.
        • Acharya M.
        • Montay-Gruel P.
        • Loo Jr. B.W.
        • Vozenin M.-C.
        • Maxim P.G.
        Ultra-high dose rate electron beams and the FLASH effect: From preclinical evidence to a new radiotherapy paradigm.
        Med Phys. 2022; 49: 2082-2095https://doi.org/10.1002/mp.15442
        • Petersson K.
        • Jaccard M.
        • Germond J.-F.
        • Buchillier T.
        • Bochud F.
        • Bourhis J.
        • et al.
        High dose-per-pulse electron beam dosimetry — A model to correct for the ion recombination in the advanced Markus ionization chamber.
        Med Phys. 2017; 44: 1157-1167https://doi.org/10.1002/mp.12111
        • Schüller A.
        • Heinrich S.
        • Fouillade C.
        • Subiel A.
        • De Marzi L.
        • Romano F.
        • et al.
        The European joint research project UHDpulse – metrology for advanced radiotherapy using particle beams with ultra-high pulse dose rates.
        Physica Med. 2020; 80: 134-150https://doi.org/10.1016/j.ejmp.2020.09.020
        • Gómez F.
        • Gonzalez-Castaño D.M.
        • Fernández N.G.
        • Pardo-Montero J.
        • Schüller A.
        • Gasparini A.
        • et al.
        Development of an ultra-thin parallel plate ionization chamber for dosimetry in FLASH radiotherapy.
        Med Phys. 2022; 49: 4705-4714https://doi.org/10.1002/mp.15668
        • Marinelli M.
        • Felici G.
        • Galante F.
        • Gasparini A.
        • Giuliano L.
        • Heinrich S.
        • et al.
        Design, realization, and characterization of a novel diamond detector prototype for FLASH radiotherapy dosimetry.
        Med Phys. 2022; 49: 1902-1910https://doi.org/10.1002/mp.15473
        • Boag J.W.
        Ionization measurements at very high intensities—Part I.
        Br J Radiol. 1950; 23: 601-611https://doi.org/10.1259/0007-1285-23-274-601
        • Kranzer R.
        • Poppinga D.
        • Weidner J.
        • Schüller A.
        • Hackel T.
        • Looe H.K.
        • et al.
        Ion collection efficiency of ionization chambers in ultra-high dose-per-pulse electron beams.
        Med Phys. 2021; 48: 819-830https://doi.org/10.1002/mp.14620
        • Kossyi I.A.
        • Kostinsky A.Y.
        • Matveyev A.A.
        • Silakov V.P.
        Kinetic scheme of the non-equilibrium discharge in nitrogen- oxygen mixtures.
        Plasma Sources Sci Technol. 1992; 1: 207-220https://doi.org/10.1088/0963-0252/1/3/011
        • Ramo S.
        Currents induced by electron motion.
        Proc IRE. 1939; 27: 584-585https://doi.org/10.1109/JRPROC.1939.228757
        • Aguiar P.
        • González-Castaño D.
        • Gómez F.
        • Pardo-Montero J.
        Recombination in liquid filled ionisation chambers with multiple charge carrier species: Theoretical and numerical results.
        Radiat Phys Chem. 2014; 103: 172-177https://doi.org/10.1016/j.radphyschem.2014.05.058
        • Gotz M.
        • Karsch L.
        • Pawelke J.
        A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse.
        Phys Med Biol. 2017; 62: 8634-8654https://doi.org/10.1088/1361-6560/aa8985
        • Boissonnat G.
        • Fontbonne J.-M.
        • Colin J.
        • Remadi A.
        • Salvador S.
        Measurement of ion and electron drift velocity and electronic attachment in air for ionization chambers.
        2016https://doi.org/10.48550/arXiv.1609.03740
        • Zhang B.
        • He J.
        • Ji Y.
        Prediction of average mobility of ions from corona discharge in air with respect to pressure, humidity and temperature.
        IEEE Trans Dielectrics Electr Insul. 2019; 26: 1403-1410https://doi.org/10.1109/TDEI.2019.008001
        • Biagi S.
        Monte Carlo simulation of electron drift and diffusion in counting gases under the influence of electric and magnetic fields.
        Nucl Instrum Methods Phys Res A. 1999; 421: 234-240https://doi.org/10.1016/S0168-9002(98)01233-9
        • Sauli F.
        Gaseous radiation detectors: fundamentals and applications. Cambridge monographs on particle physics, nuclear physics and cosmology. Cambridge University Press, 2014https://doi.org/10.1017/CBO9781107337701
        • Franchin A.
        • Ehrhart S.
        • Leppä J.
        • Nieminen T.
        • Gagné S.
        • Schobesberger S.
        • et al.
        Experimental investigation of ion–ion recombination under atmospheric conditions.
        Atmos Chem Phys. 2015; 15: 7203-7216https://doi.org/10.5194/acp-15-7203-2015
        • Davies D.
        Measurements of swarm parameters in dry air.
        in: Technical note 346. 1983
        • Davies D.
        • Chantry P.
        Air chemistry measurements II.
        Air Force Weapons Laboratory, 1985
        • Hegerberg R.
        • Reid I.D.
        Electron drift velocities in air.
        Aust J Phys. 1980; 33: 227-238https://doi.org/10.1071/ph800227a
        • Laitano R.F.
        • Guerra A.S.
        • Pimpinella M.
        • Caporali C.
        • Petrucci A.
        Charge collection efficiency in ionization chambers exposed to electron beams with high dose per pulse.
        Phys Med Biol. 2006; 51: 6419-6436https://doi.org/10.1088/0031-9155/51/24/009
        • McManus M.
        • Romano F.
        • Lee N.D.
        • Farabolini W.
        • Gilardi A.
        • Royle G.
        • et al.
        The challenge of ionisation chamber dosimetry in ultra-short pulsed high dose-rate very high energy electron beams.
        Sci Rep. 2020; 10: 9089https://doi.org/10.1038/s41598-020-65819-y
        • Bourgouin A.
        • Knyziak A.
        • Marinelli M.
        • Kranzer R.
        • Schüller A.
        • Kapsch R.-P.
        Characterization of the PTB ultra-high pulse dose rate reference electron beam.
        Phys Med Biol. 2022; 67https://doi.org/10.1088/1361-6560/ac5de8
        • Schuller A.
        • Makowski C.
        • Kapsch R.-P.
        • Nolte R.
        • Beck P.
        MELAF -a 50 MeV electron accelerator facility for research in radiation effects.
        in: 2017 17th european conference on radiation and its effects on components and systems. 2017: 1-4https://doi.org/10.1109/RADECS.2017.8696226
        • Bourgouin A.
        • Schüller A.
        • Hackel T.
        • Kranzer R.
        • Poppinga D.
        • Kapsch R.-P.
        • et al.
        Calorimeter for real-time dosimetry of pulsed ultra-high dose rate electron beams.
        Front Phys. 2020; : 400https://doi.org/10.3389/FPHY.2020.567340
        • Krauss A.
        The PTB water calorimeter for the absolute determination of absorbed dose to water in 60co radiation.
        Metrologia. 2006; 43: 259-272https://doi.org/10.1088/0026-1394/43/3/008
        • Muir B.R.
        • Rogers D.W.O.
        Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types.
        Med Phys. 2014; 41111701https://doi.org/10.1118/1.4893915
        • Abdel-Rahman W.
        • Seuntjens J.P.
        • Verhaegen F.
        • Podgorsak E.B.
        Radiation induced currents in parallel plate ionization chambers: Measurement and Monte Carlo simulation for megavoltage photon and electron beams.
        Med Phys. 2006; 33: 3094-3104https://doi.org/10.1118/1.2208917